Severity
Moderate
FDA Devices recall · Reported November 28, 2012
MAKO Surgical Group recalled their RIO System software, version 2.4 and is implementing software version 2.5 to address a software functional issue with the existing version of th…
Mako Surgical Corporation recalled The RIO (TGS 2.), Model # MAKO TGS 2.0 (Part No. 204000). The RIO is intended to assis… - a moderate-severity action.
The RIO (TGS 2.), Model # MAKO TGS 2.0 (Part No. 204000). The RIO is intended to assis… was recalled by Mako Surgical Corporation in November 28, 2012. Reason: MAKO Surgical Group recalled their RIO System software, version 2.4 and is implementing software version 2.5 …. Check the official notice for the remedy. Verify recall #Z-0385-2013 with the FDA Devices before acting.
The recall
Mako Surgical Corporation issued this moderate-severity FDA Devices recall-MAKO Surgical Group recalled their RIO System software, version 2.4 and is implementing software version 2.5 ….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0385-2013 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0385-2013) was formally reported on November 28, 2012, with the manufacturer initiating the action on September 12, 2012. It is classified under Moderate severity (Class II), with a current status of Terminated. Mako Surgical Corporation is listed as the recalling firm, operating out of Plantation, FL. Federal records list the affected scope as 137 devices.
The documented reason for this recall is: MAKO Surgical Group recalled their RIO System software, version 2.4 and is implementing software version 2.5 to address a software functional issue with the existing version of the system's software. Loss of tactile fee… Distribution data in the federal record shows the product reached: Worldwide Distribution - USA including IL, OR, IA, AZ, CA, CO, FL, MS, OK, PA, WV, NC, TX, NY, GA, NJ, OH, WA, VA, MA, MD, WI, MI, NV, MO, AL, UT, RI, AR, LA, ID, IN, and TN and Internationally to Japan, Scotland, Ital…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
137 devices
Related Recalls
6
3 from same agency
The RIO (TGS 2.), Model # MAKO TGS 2.0 (Part No. 204000). The RIO is intended to assist the surgeon in providing software defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures. The RIO is indicated for use in surgical knee and hip procedures, in which the use of stereotactic surgery may be appropriate, and where reference to rigid anatomical bony structures can be identified relative to a CT based model of the anatomy. These procedures include: Unicondylar knee replacement (UKA) and/or Patellofemoral knee replacement (PKA) and Total Hip Arthroplasty (THA).
MAKO Surgical Group recalled their RIO System software, version 2.4 and is implementing software version 2.5 to address a software functional issue with the existing version of the system's software. Loss of tactile feedback constraining the cutting burr has been reported.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0385-2013 |
| Date reported | November 28, 2012 |
| Date initiated | September 12, 2012 |
| Recalling firm | Mako Surgical Corporation |
| Firm location | Plantation, FL |
| Affected scope | 137 devices |
| Distribution | Worldwide Distribution - USA including IL, OR, IA, AZ, CA, CO, FL, MS, OK, PA, WV, NC, TX, NY, GA, NJ, OH, WA, VA, MA, MD, WI, MI, NV, MO, AL, UT, RI, AR, LA, ID, IN, and TN and Internationally to Japan, Scotland, Italy, Singapore, Turkey… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported November 28, 2012. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.