Severity
Moderate
FDA Devices recall · Reported December 4, 2013
There is a software synchronization issue with the ACCU-CHEK Combo system. In rare cases, when the Manual Pump option is chosen on the ACCU-CHEK Aviva Combo meter, there is a poss…
Roche Diagnostics Operations, Inc. recalled ACCU-CHEK Aviva Combo Blood Glucose Meter (as a part of the ACCU-CHEK Aviva Combo System)… - a moderate-severity action.
ACCU-CHEK Aviva Combo Blood Glucose Meter (as a part of the ACCU-CHEK Aviva Combo System)… was recalled by Roche Diagnostics Operations, Inc. in December 4, 2013. Reason: There is a software synchronization issue with the ACCU-CHEK Combo system. In rare cases, when the Manual Pum…. Check the official notice for the remedy. Verify recall #Z-0385-2014 with the FDA Devices before acting.
The recall
Roche Diagnostics Operations, Inc. issued this moderate-severity FDA Devices recall-There is a software synchronization issue with the ACCU-CHEK Combo system. In rare cases, when the Manual Pum….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0385-2014 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0385-2014) was formally reported on December 4, 2013, with the manufacturer initiating the action on October 25, 2013. It is classified under Moderate severity (Class II), with a current status of Terminated. Roche Diagnostics Operations, Inc. is listed as the recalling firm, operating out of Indianapolis, IN. Federal records list the affected scope as 6,595 units.
The documented reason for this recall is: There is a software synchronization issue with the ACCU-CHEK Combo system. In rare cases, when the Manual Pump option is chosen on the ACCU-CHEK Aviva Combo meter, there is a possibility of receiving an incorrect bolus … Distribution data in the federal record shows the product reached: Nationwide Distribution.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
6,595 units
Related Recalls
6
3 from same agency
ACCU-CHEK Aviva Combo Blood Glucose Meter (as a part of the ACCU-CHEK Aviva Combo System)\ The ACCU-CHEK Combo System is indicated for the treatment of insulin-requiring diabetes and for the quantitative measurement of glucose in fresh capillary whole blood from the finger.
There is a software synchronization issue with the ACCU-CHEK Combo system. In rare cases, when the Manual Pump option is chosen on the ACCU-CHEK Aviva Combo meter, there is a possibility of receiving an incorrect bolus advice recommendation that may cause a temporary under delivery of insulin.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0385-2014 |
| Date reported | December 4, 2013 |
| Date initiated | October 25, 2013 |
| Recalling firm | Roche Diagnostics Operations, Inc. |
| Firm location | Indianapolis, IN |
| Affected scope | 6,595 units |
| Distribution | Nationwide Distribution. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported December 4, 2013. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.