FDA Devices recall · Reported November 14, 2018

Alaris Infusion Pump administration sets labeled as: -Infusion Set Alaris Pump Module/Medley Pump Module/Gemini Infusion Pump, REF 10011301, CareFusion, WARNING: TO PREVENT FREE-FLOW, CLOSE SET CLAMP WHEN SAFETY CLAMP FITMENT ON PUMP SEGMENT IS OPEN, Fluid path is STERILE and NONPYROGENIC. 12276678 Rev 00 -VersaSafe Infusion Set Alaris Pump Module/Medley Pump Module/Gemini Infusion Pump, REF 2120-0500, WARNING: TO PREVENT FREE-FLOW, CLOSE SET CLAMP WHEN SAFETY CLAMP ON PUMP SEGMENT IS OPE

Complaints where users were unable to prime the administration set.

Recall #
Z-0385-2019
Affected scope
118,254,342
Initiated
April 20, 2018
Verify with FDA Devices →
View my saved recalls

CareFusion 303, Inc. recalled Alaris Infusion Pump administration sets labeled as: -Infusion Set Alaris Pump Module/M… - a moderate-severity action.

Alaris Infusion Pump administration sets labeled as: -Infusion Set Alaris Pump Module/M… was recalled by CareFusion 303, Inc. in November 14, 2018. Reason: Complaints where users were unable to prime the administration set.. Check the official notice for the remedy. Verify recall #Z-0385-2019 with the FDA Devices before acting.

The recall

CareFusion 303, Inc. issued this moderate-severity FDA Devices recall-Complaints where users were unable to prime the administration set..

Moderate
severity level
118.3M units
affected scope
Class II
classification
November 14, 2018
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-0385-2019 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This recall's affected scope (see the Recall Profile below) places it in the million-unit bracket that typically triggers nationwide consumer alerts and retailer sweeps. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 41,021 medical devices recalls on record

01,0002,0003,0004,000 20052008201120142017202020232026 1,947
Severity1906864652High severity (most serious)Moderate severityLow severity
Where this agency-classified recall sits among FDA Class I/II/III notices

Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Product description

Alaris Infusion Pump administration sets labeled as: -Infusion Set Alaris Pump Module/Medley Pump Module/Gemini Infusion Pump, REF 10011301, CareFusion, WARNING: TO PREVENT FREE-FLOW, CLOSE SET CLAMP WHEN SAFETY CLAMP FITMENT ON PUMP SEGMENT IS OPEN, Fluid path is STERILE and NONPYROGENIC. 12276678 Rev 00 -VersaSafe Infusion Set Alaris Pump Module/Medley Pump Module/Gemini Infusion Pump, REF 2120-0500, WARNING: TO PREVENT FREE-FLOW, CLOSE SET CLAMP WHEN SAFETY CLAMP ON PUMP SEGMENT IS OPEN. FLUID PATH IS STERILE and NON-PYROGENIC. PRODUCT is sterilized by GAMMA IRRADIATION. CareFusion, 630-01104 Rev 00 -SmartSite Infusion Set Alaris Pump Module/Medley Pump Module/Gemini Infusion Pump REF 2420-0500,WARNING: TO PREVENT FREE-FLOW, CLOSE SET CLAMP WHEN SAFETY CLAMP ON PUMP SEGMENT IS OPEN. FLUID PATH IS STERILE and NON-PYROGENIC. PRODUCT is sterilized by GAMMA IRRADIATION. CareFusion, 630-00952 Rev 01 -Infusion Set Alaris Pump Module/Medley Pump Module REF 24009-0007T, WARNING: TO PREVENT FREE-FLOW, CLOSE SET CLAMP WHEN SAFETY CLAMP ON PUMP SEGMENT IS OPEN, FLUID PATH IS STERILE and NON-PYROGENIC. PRODUCT is sterilized by GAMMA IRRADIATION. CareFusion 630-00650 Rev 00 Production Usage: Administration sets allow for the intravenous administration of fluids, medications, nutritional support and transfusion therapy through a needle or catheter inserted into the patient s artery or vein. The sets are comprised with components commonly found on intravascular administration sets and extension sets. The components include an Alaris Pump module silicone segment and fitments, drip chamber, check valve, one or more SmartSite needleless connectors, slide clamp, male luer with spin lock and tubing with different length of tubing.

Reason for recall

Complaints where users were unable to prime the administration set.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Terminated
Recall number Z-0385-2019
Date reported November 14, 2018
Date initiated April 20, 2018
Recalling firm CareFusion 303, Inc.
Firm location San Diego, CA
Affected scope 118,254,342
Distribution Worldwide Distribution - US Nationwide including Guam International: Canada, Australia, New Zealand, EU, Middel East, South Africa, Okinawa, Saipan, Buenos Aires, Lima, Selangor, Hong Kong, Taiwan, and Budapest

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

118,254,342 units affected - million-unit bracket.

Regional (<10K units) -
Multi-state (10K – 100K units) -
Large-scale (100K – 1M units) -
Massive (≥1M units) ✓ This recall

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-0385-2019) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
Alaris Infusion Pump administration sets labeled as: -Infusion Set Alaris Pump Module/Medley Pump Module/Gemini Infusion Pump, REF 10011301, CareFusion, WARNING: TO PREVENT FREE-FLOW, CLOSE SET CLAMP WHEN SAFETY CLAMP FITMENT ON PUMP SEGMENT IS OPEN, Fluid path is STERILE and NONPYROGENIC. 12276678 Rev 00 -VersaSafe Infusion Set Alaris Pump Module/Medley Pump Module/Gemini Infusion Pump, REF 2120-0500, WARNING: TO PREVENT FREE-FLOW, CLOSE SET CLAMP WHEN SAFETY CLAMP ON PUMP SEGMENT IS OPEN. FLUID PATH IS STERILE and NON-PYROGENIC. PRODUCT is sterilized by GAMMA IRRADIATION. CareFusion, 630-01104 Rev 00 -SmartSite Infusion Set Alaris Pump Module/Medley Pump Module/Gemini Infusion Pump REF 2420-0500,WARNING: TO PREVENT FREE-FLOW, CLOSE SET CLAMP WHEN SAFETY CLAMP ON PUMP SEGMENT IS OPEN. FLUID PATH IS STERILE and NON-PYROGENIC. PRODUCT is sterilized by GAMMA IRRADIATION. CareFusion, 630-00952 Rev 01 -Infusion Set Alaris Pump Module/Medley Pump Module REF 24009-0007T, WARNING: TO PREVENT FREE-FLOW, CLOSE SET CLAMP WHEN SAFETY CLAMP ON PUMP SEGMENT IS OPEN, FLUID PATH IS STERILE and NON-PYROGENIC. PRODUCT is sterilized by GAMMA IRRADIATION. CareFusion 630-00650 Rev 00 Production Usage: Administration sets allow for the intravenous administration of fluids, medications, nutritional support and transfusion therapy through a needle or catheter inserted into the patient s artery or vein. The sets are comprised with components commonly found on intravascular administration sets and extension sets. The components include an Alaris Pump module silicone segment and fitments, drip chamber, check valve, one or more SmartSite needleless connectors, slide clamp, male luer with spin lock and tubing with different length of tubing.. Recalled by CareFusion 303, Inc.. Units affected: 118,254,342.
Why was this product recalled?
Complaints where users were unable to prime the administration set.
Which agency issued this recall?
This recall was issued by the FDA Devices on November 14, 2018. Severity: Moderate. Recall number: Z-0385-2019.
Where was the recalled product distributed?
Distribution: Worldwide Distribution - US Nationwide including Guam International: Canada, Australia, New Zealand, EU, Middel East, South Africa, Okinawa, Saipan, Buenos Aires, Lima, Selangor, Hong Kong, Taiwan, and Budapest.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nationwide recalls with similar severity or category timing

Drug and device notices outside this firm and state, matched by severity timing and product category.

Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls

This recall: FDA Devices, reported November 14, 2018. Severity classes follow each agency's own taxonomy; related-recall context is computed across the full archive.

Methodology · Corrections · Archive as of August 2026