Severity
Moderate
FDA Devices recall · Reported December 4, 2013
All lots of the Synthes Tensioning Device were recalled due to the potential for the reaction fork to come loose from the main body of the device and not allow the wire to release…
Synthes USA Hq, Inc. recalled Synthes Tensioning Device The Synthes Osteogenesis System, MR Conditional is intended… - a moderate-severity action.
Synthes Tensioning Device The Synthes Osteogenesis System, MR Conditional is intended… was recalled by Synthes USA Hq, Inc. in December 4, 2013. Reason: All lots of the Synthes Tensioning Device were recalled due to the potential for the reaction fork to come lo…. Check the official notice for the remedy. Verify recall #Z-0388-2014 with the FDA Devices before acting.
The recall
Synthes USA Hq, Inc. issued this moderate-severity FDA Devices recall-All lots of the Synthes Tensioning Device were recalled due to the potential for the reaction fork to come lo….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0388-2014 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0388-2014) was formally reported on December 4, 2013, with the manufacturer initiating the action on July 31, 2013. It is classified under Moderate severity (Class II), with a current status of Terminated. Synthes USA Hq, Inc. is listed as the recalling firm, operating out of West Chester, PA. Federal records list the affected scope as 369.
The documented reason for this recall is: All lots of the Synthes Tensioning Device were recalled due to the potential for the reaction fork to come loose from the main body of the device and not allow the wire to release, nor the device to apply tension. Distribution data in the federal record shows the product reached: Nationwide Distribution.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
369
Related Recalls
6
3 from same agency
Synthes Tensioning Device The Synthes Osteogenesis System, MR Conditional is intended for fracture fixation (open and closed), pseudoarthrosis or non-unions of long bones, limb lengthening by epiphyseal or metaphyseal distraction, correction of bony or soft tissue deformities and correction of segmental bony or soft tissue defects in adult and pediatric patients.
All lots of the Synthes Tensioning Device were recalled due to the potential for the reaction fork to come loose from the main body of the device and not allow the wire to release, nor the device to apply tension.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0388-2014 |
| Date reported | December 4, 2013 |
| Date initiated | July 31, 2013 |
| Recalling firm | Synthes USA Hq, Inc. |
| Firm location | West Chester, PA |
| Affected scope | 369 |
| Distribution | Nationwide Distribution. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported December 4, 2013. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.