SYNCHRON Systems Cholesterol (CHOL) Reagent, REF: 467825

Reported: December 29, 2021 Initiated: November 10, 2021 #Z-0399-2022

Product Description

SYNCHRON Systems Cholesterol (CHOL) Reagent, REF: 467825

Reason for Recall

Cholesterol Reagent may fail with Out of Calibration Range Low (OCR Low) and/or Back to Back (B/B) errors. Low quality control recovery failures may be observed after passing calibration. Brown particulates in compartment C of the reagent cartridge may be observed. Continued use may lead to, 1) failed QC and/or calibrations, 2) Erroneous low Cholesterol patient results.

Details

Recalling Firm
Beckman Coulter Inc.
Units Affected
1433
Distribution
US Nationwide distribution in the states of Louisiana, Ohio, North Carolina, California, South Carolina, Oklahoma, Pennsylvania, New York, Florida, New Jersey, Wisconsin, Alabama, Texas, Georgia, Maine, North Dakota, Virginia, Kansas, Washington, New Mexico, Iowa ,Nebraska, Idaho, Michigan, Maryland, Arizona, Hawaii, Missouri, Oregon, Mississippi, Rhode Island, Alaska, Tennessee, Illinois, Delaware, Kentucky, Arkansas, West Virginia, Massachusetts. OUS: India, South Africa, Saudi Arabia, Canada, Lebanon, Singapore, Philippines, Thailand
Location
Brea, CA

Frequently Asked Questions

What product was recalled?
SYNCHRON Systems Cholesterol (CHOL) Reagent, REF: 467825. Recalled by Beckman Coulter Inc.. Units affected: 1433.
Why was this product recalled?
Cholesterol Reagent may fail with Out of Calibration Range Low (OCR Low) and/or Back to Back (B/B) errors. Low quality control recovery failures may be observed after passing calibration. Brown particulates in compartment C of the reagent cartridge may be observed. Continued use may lead to, 1) failed QC and/or calibrations, 2) Erroneous low Cholesterol patient results.