Severity
Moderate
FDA Devices recall · Reported December 29, 2021
Set screws securing the two gliding rods of the Extra Monitors (called 7th and 8th monitor) mounted on the Monitor Ceiling Suspension (MCS) can come loose, the Extra monitors migh…
Philips North America Llc recalled Philips Allura Xper FD series with Software Version Number: 2.1.x- intended for Vascular,… - a moderate-severity action.
Philips Allura Xper FD series with Software Version Number: 2.1.x- intended for Vascular,… was recalled by Philips North America Llc in December 29, 2021. Reason: Set screws securing the two gliding rods of the Extra Monitors (called 7th and 8th monitor) mounted on the Mo…. Check the official notice for the remedy. Verify recall #Z-0416-2022 with the FDA Devices before acting.
The recall
Philips North America Llc issued this moderate-severity FDA Devices recall-Set screws securing the two gliding rods of the Extra Monitors (called 7th and 8th monitor) mounted on the Mo….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0416-2022 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0416-2022) was formally reported on December 29, 2021, with the manufacturer initiating the action on November 19, 2021. It is classified under Moderate severity (Class II), with a current status of Terminated. Philips North America Llc is listed as the recalling firm, operating out of Cambridge, MA. Federal records list the affected scope as USA: 5 systems R.O.W.: 15 systems.
The documented reason for this recall is: Set screws securing the two gliding rods of the Extra Monitors (called 7th and 8th monitor) mounted on the Monitor Ceiling Suspension (MCS) can come loose, the Extra monitors might disengage from the Mounting Ceiling Su… Distribution data in the federal record shows the product reached: Worldwide distribution - US Nationwide distribution in the states of DC, GA, MA, TX, VA and the countries of Dominican Republic, INDIA, Japan, Poland, Spain, Taiwan.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
USA: 5 systems R.O.W.: 15 systems
Related Recalls
6
3 from same agency
Philips Allura Xper FD series with Software Version Number: 2.1.x- intended for Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. Model Number: 722005, 722006, 722008, 722010, 722012, 722013, 722025, 722026, 722027, 722028.
Set screws securing the two gliding rods of the Extra Monitors (called 7th and 8th monitor) mounted on the Monitor Ceiling Suspension (MCS) can come loose, the Extra monitors might disengage from the Mounting Ceiling Suspension structure and fall down, may lead to patient, user or service engineer harm that may require medical intervention.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0416-2022 |
| Date reported | December 29, 2021 |
| Date initiated | November 19, 2021 |
| Recalling firm | Philips North America Llc |
| Firm location | Cambridge, MA |
| Affected scope | USA: 5 systems R.O.W.: 15 systems |
| Distribution | Worldwide distribution - US Nationwide distribution in the states of DC, GA, MA, TX, VA and the countries of Dominican Republic, INDIA, Japan, Poland, Spain, Taiwan. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported December 29, 2021. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.