FDA Devices Moderate Class II Terminated

LineSider Spinal System 5.5/6.0mm, Non-mod MIS 7.1mm - Dual Lead, Cannulated. Thoracolumbosacral pedicle screw system; Spinal interlaminal fixation orthosis. Product sold as non-sterile and shipped in protective sterilization trays. Screws can also be shipped individually, as tray replenishments. Labeled as the following sizes: a. 5.5mm x 35mm; b. 5.5mm x 40mm; c. 5.5mm x 45mm; d. 5.5mm x 50mm; e. 6.5mm x 35mm; f. 6.5mm x 40mm; g. 6.5mm x 45mm; h. 6.5m

Reported: December 14, 2022 Initiated: January 27, 2022 #Z-0420-2023

Product Description

LineSider Spinal System 5.5/6.0mm, Non-mod MIS 7.1mm - Dual Lead, Cannulated. Thoracolumbosacral pedicle screw system; Spinal interlaminal fixation orthosis. Product sold as non-sterile and shipped in protective sterilization trays. Screws can also be shipped individually, as tray replenishments. Labeled as the following sizes: a. 5.5mm x 35mm; b. 5.5mm x 40mm; c. 5.5mm x 45mm; d. 5.5mm x 50mm; e. 6.5mm x 35mm; f. 6.5mm x 40mm; g. 6.5mm x 45mm; h. 6.5mm x 50mm; i. 6.5mm x 55mm; j. 7.5mm x 35mm; k. 7.5mm x 40mm; l. 7.5mm x 45mm; m. 7.5mm x 50mm; n. 7.5mm x 55mm; o. 8.5mm x 35mm; p. 8.5mm x 40mm; q. 8.5mm x 45mm; r. 8.5mm x 50mm; s.8.5mm x 55mm; t. 9.5mm x 35mm; u. 9.5mm x 40mm; v. 9.5mm x 45mm; w. 9.5mm x 50mm; x. 9.5mm x 55mm; Used in spinal fixation surgery.

Reason for Recall

Potential for weld separation of the Upper Tulip Head component from the Lower Tulip Head component, which would result in the separation of the Tulip Head from the Screw Shank.

Details

Recalling Firm
Integrity Implants Inc.
Units Affected
1,369 screws
Distribution
Nationwide distribution to AZ, CA, CT, DC, FL, IL, IN, LA, MA, MN, NJ, OK, PR, TN, TX, VA.
Location
Palm Beach Gardens, FL

Frequently Asked Questions

What product was recalled?
LineSider Spinal System 5.5/6.0mm, Non-mod MIS 7.1mm - Dual Lead, Cannulated. Thoracolumbosacral pedicle screw system; Spinal interlaminal fixation orthosis. Product sold as non-sterile and shipped in protective sterilization trays. Screws can also be shipped individually, as tray replenishments. Labeled as the following sizes: a. 5.5mm x 35mm; b. 5.5mm x 40mm; c. 5.5mm x 45mm; d. 5.5mm x 50mm; e. 6.5mm x 35mm; f. 6.5mm x 40mm; g. 6.5mm x 45mm; h. 6.5mm x 50mm; i. 6.5mm x 55mm; j. 7.5mm x 35mm; k. 7.5mm x 40mm; l. 7.5mm x 45mm; m. 7.5mm x 50mm; n. 7.5mm x 55mm; o. 8.5mm x 35mm; p. 8.5mm x 40mm; q. 8.5mm x 45mm; r. 8.5mm x 50mm; s.8.5mm x 55mm; t. 9.5mm x 35mm; u. 9.5mm x 40mm; v. 9.5mm x 45mm; w. 9.5mm x 50mm; x. 9.5mm x 55mm; Used in spinal fixation surgery.. Recalled by Integrity Implants Inc.. Units affected: 1,369 screws.
Why was this product recalled?
Potential for weld separation of the Upper Tulip Head component from the Lower Tulip Head component, which would result in the separation of the Tulip Head from the Screw Shank.
Which agency issued this recall?
This recall was issued by the FDA Devices on December 14, 2022. Severity: Moderate. Recall number: Z-0420-2023.