Beckman Coulter UniCel DxI 600, Part Numbers: A71460, A30260; Photometric Chemistry Analyzer
Reported: November 12, 2025 Initiated: October 1, 2025 #Z-0428-2026
Product Description
Beckman Coulter UniCel DxI 600, Part Numbers: A71460, A30260; Photometric Chemistry Analyzer
Reason for Recall
An internal investigation has determined that due to defective sub-components within the gantry assembly, a z-axis CAD movement motion error may occur and cause an analyzer to enter a Not Ready or Paused state, leading to a delay in obtaining results.
Details
- Recalling Firm
- Beckman Coulter, Inc.
- Units Affected
- 80 units
- Distribution
- Worldwide
- Location
- Chaska, MN
Frequently Asked Questions
What product was recalled? ▼
Beckman Coulter UniCel DxI 600, Part Numbers: A71460, A30260; Photometric Chemistry Analyzer. Recalled by Beckman Coulter, Inc.. Units affected: 80 units.
Why was this product recalled? ▼
An internal investigation has determined that due to defective sub-components within the gantry assembly, a z-axis CAD movement motion error may occur and cause an analyzer to enter a Not Ready or Paused state, leading to a delay in obtaining results.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on November 12, 2025. Severity: Moderate. Recall number: Z-0428-2026.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11