1202 Flex Focus Ultrasound System, cart models UA1214 and UA1814, used with: Flex Focus 200, Flex Focus 300, Flex Focus 400, Flex Focus 500, Flex Focus 700, Flex Focus 800; System, Imaging, Pulsed Doppler, Ultrasonic

Reported: November 12, 2025 Initiated: October 9, 2025 #Z-0435-2026 4426 units units

B-K Medical A/S issued this FDA Devices recall on November 12, 2025. Classified as Moderate severity (Class II). Approximately 4426 units units are affected. The recall was issued because: GE HealthCare has become aware that the Instructions for Use within the user guide and service manual do not adequately…. This recall notice is sourced from official FDA Devices enforcement records. Below you will find the complete product description, hazard information, remedy instructions, and related recalls from the same manufacturer or product category.

Recall Insight

This FDA Devices action (record #Z-0435-2026) was formally reported on November 12, 2025, with the manufacturer initiating the action on October 9, 2025. It is classified under Moderate severity (Class II), with a current status of Ongoing. B-K Medical A/S is listed as the recalling firm, operating out of Herlev, N/A. Federal records indicate 4426 units units are affected.

The documented reason for this recall is: GE HealthCare has become aware that the Instructions for Use within the user guide and service manual do not adequately state the required battery replacement intervals, end-of-life handling procedures, or general batte… Distribution data in the federal record shows the product reached: Worldwide distribution: US (nationwide) and OUS (foreign) to countries of: Algeria, Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark,…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

To put this record in context, PlainRecalls indexes 83,949 recalls across the FDA, CPSC, NHTSA and USDA FSIS going back to 1995. Within the same product category, the database holds 6 closely related recalls, of which 6 were also issued by FDA Devices. That clustering is a signal — repeated actions in a narrow category often indicate a systemic quality-control issue, a supplier-wide contamination, or a design defect that has propagated across product lines. This recall is roughly 1 year old; older recalls can remain relevant because many units enter resale, rental, and secondary-market channels where the original warning never reaches the end user. Always cross-check the recall number against the official agency page before relying on any summary.

Recall Distribution by Severity Class

Severity1Class I (Critical)Class II (Moderate)Class III (Low)
Recall Distribution by Severity Class

Severity

Moderate

Units Affected

4426 units

Related Recalls

6

6 from same agency

Product Description

1202 Flex Focus Ultrasound System, cart models UA1214 and UA1814, used with: Flex Focus 200, Flex Focus 300, Flex Focus 400, Flex Focus 500, Flex Focus 700, Flex Focus 800; System, Imaging, Pulsed Doppler, Ultrasonic

Reason for Recall

GE HealthCare has become aware that the Instructions for Use within the user guide and service manual do not adequately state the required battery replacement intervals, end-of-life handling procedures, or general battery safety precautions.

Details

Recalling Firm
B-K Medical A/S
Units Affected
4426 units
Distribution
Worldwide distribution: US (nationwide) and OUS (foreign) to countries of: Algeria, Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Rep, Ecuador, Estonia, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Korea (Republic Of), Kuwait, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Mongolia, Morocco, Nepal, Netherlands, New Zealand, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkiye, Ukraine, United Arab Emirates, and United Kingdom.
Location
Herlev, N/A

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Ongoing
Recall number Z-0435-2026
Date reported November 12, 2025
Date initiated October 9, 2025
Recalling firm B-K Medical A/S
Units affected 4426 units
Distribution Worldwide distribution: US (nationwide) and OUS (foreign) to countries of: Algeria, Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Rep, Ecua…

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

4426 units units affected — limited or regional distribution scale.

Regional (<10K units) ✓ This recall
Multi-state (10K – 100K units)
Large-scale (100K – 1M units)
Massive (≥1M units)

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

Frequently Asked Questions

What product was recalled?
1202 Flex Focus Ultrasound System, cart models UA1214 and UA1814, used with: Flex Focus 200, Flex Focus 300, Flex Focus 400, Flex Focus 500, Flex Focus 700, Flex Focus 800; System, Imaging, Pulsed Doppler, Ultrasonic. Recalled by B-K Medical A/S. Units affected: 4426 units.
Why was this product recalled?
GE HealthCare has become aware that the Instructions for Use within the user guide and service manual do not adequately state the required battery replacement intervals, end-of-life handling procedures, or general battery safety precautions.
Which agency issued this recall?
This recall was issued by the FDA Devices on November 12, 2025. Severity: Moderate. Recall number: Z-0435-2026.
Where was the recalled product distributed?
Distribution: Worldwide distribution: US (nationwide) and OUS (foreign) to countries of: Algeria, Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Rep, Ecuador, Estonia, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Korea (Republic Of), Kuwait, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Mongolia, Morocco, Nepal, Netherlands, New Zealand, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkiye, Ukraine, United Arab Emirates, and United Kingdom..
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-0435-2026) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Recall Context

Product recalls are issued when a manufacturer, distributor, or federal agency determines that a product poses a safety risk to consumers. This recall is classified as moderate severity, indicating the product may cause temporary or medically reversible health consequences. Across PlainRecalls, we track 83,000+ recalls from FDA, CPSC, and NHTSA to help consumers stay informed and act quickly when safety issues arise.

Nearby Recalls in This Category

Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Endo-Model Replacement Plateau; Item Number: 15-0027/11; →

Data Sources

Data as of 2025. Source: FDA, CPSC, NHTSA, USDA FSIS federal recall databases.

  • Source: FDA — Food and Drug Administration, openFDA Enforcement API (food, drug, and medical device recalls)
  • Source: CPSC — Consumer Product Safety Commission Recalls API (consumer product recalls and hazards)
  • Source: NHTSA — National Highway Traffic Safety Administration Recalls API (vehicle safety recalls)
  • Source: USDA FSIS — Food Safety and Inspection Service (meat, poultry, and egg product recalls)

Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.

All federal data sources used on this page

Source: Federal recall agencies (FDA, CPSC, NHTSA, USDA FSIS) Aggregated multi-agency recall feeds · 2024 Recall data normalized across federal agency feeds; severity classifications follow each agency's own taxonomy (FDA Class I/II/III; CPSC, NHTSA, USDA FSIS).