Severity
Moderate
FDA Devices recall · Reported December 5, 2012
Zimmer is initiating a lot specific recall of the NexGen¿ Stemmed Nonaugmentable Tibial Component due to the devices being processed through a manufacturing cleaning operation tha…
Zimmer, Inc. recalled NexGen CompleteKnee Solution Stemmed Nonaugmentable Tibial component The NexGen¿¿ syst… - a moderate-severity action.
NexGen CompleteKnee Solution Stemmed Nonaugmentable Tibial component The NexGen¿¿ syst… was recalled by Zimmer, Inc. in December 5, 2012. Reason: Zimmer is initiating a lot specific recall of the NexGen¿ Stemmed Nonaugmentable Tibial Component due to the …. Check the official notice for the remedy. Verify recall #Z-0437-2013 with the FDA Devices before acting.
The recall
Zimmer, Inc. issued this moderate-severity FDA Devices recall-Zimmer is initiating a lot specific recall of the NexGen¿ Stemmed Nonaugmentable Tibial Component due to the ….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0437-2013 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0437-2013) was formally reported on December 5, 2012, with the manufacturer initiating the action on September 21, 2012. It is classified under Moderate severity (Class II), with a current status of Terminated. Zimmer, Inc. is listed as the recalling firm, operating out of Warsaw, IN. Federal records list the affected scope as 123 worldwide (33 USA).
The documented reason for this recall is: Zimmer is initiating a lot specific recall of the NexGen¿ Stemmed Nonaugmentable Tibial Component due to the devices being processed through a manufacturing cleaning operation that was operating outside of the validated… Distribution data in the federal record shows the product reached: Worldwide Distribution - USA including AZ,CA, FL, IA, IL, IN, MO, NE, NJ, NY, OH, PA, WA. Internationally to Thailand, Canada, Germany, France, UK, Hungary, Netherlands, Russian Fed, and Sweden,. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
123 worldwide (33 USA)
Related Recalls
6
3 from same agency
NexGen CompleteKnee Solution Stemmed Nonaugmentable Tibial component The NexGen¿¿ system is a semiconstrained, nonlinked condylar knee prostheses. The device is indicated for patients with severe knee pain and disability due to rheumatoid arthritis; osteoarthritis; traumatic arthritis; polyarthritis; collagen disorders and/or avascular necrosis of the femoral condyle; post-traumatic loss of joint configuration (particularly when there is patellofemoral erosion, dysfunction or prior patellectomy); moderate valgus, varus or flexion deformities; salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery.
Zimmer is initiating a lot specific recall of the NexGen¿ Stemmed Nonaugmentable Tibial Component due to the devices being processed through a manufacturing cleaning operation that was operating outside of the validated parameters. As a result, the devices may contain residual particulate from the manufacturing process. Risks include: Allergic reaction due to the foreign material or late onset
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0437-2013 |
| Date reported | December 5, 2012 |
| Date initiated | September 21, 2012 |
| Recalling firm | Zimmer, Inc. |
| Firm location | Warsaw, IN |
| Affected scope | 123 worldwide (33 USA) |
| Distribution | Worldwide Distribution - USA including AZ,CA, FL, IA, IL, IN, MO, NE, NJ, NY, OH, PA, WA. Internationally to Thailand, Canada, Germany, France, UK, Hungary, Netherlands, Russian Fed, and Sweden, |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported December 5, 2012. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.