PlainRecalls
CriticalClass IOngoing

FDA Devices recall · Reported November 19, 2025

Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; Product description/Product numbers: ADULT 120 EXPANDABLE W/MASK DYNJAA10591; ANES CIRC 120IN EXP-LF DYNJAA6555G; ANES CIRCUIT ADULT 120"-LF DYNJAA10817; ANESTHESIA 120" ADLT-LF DYNJAA10775; ANESTHESIA CIRC ADULT 120-LF DYNJAA10601; ANESTHESIA CIRCUIT - LF DYNJAA10261B; ANESTHESIA CIRCUIT 120EXP-LF DYNJAA10741A; ANESTHESIA CIRCUIT ADULT-LF DYNJAA0380A DYNJAA10641 DYNJAA10706B DYNJAA10851 DYN

Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and ina…

Recall #
Z-0442-2026
Affected scope
96,406 total units
Initiated
October 10, 2025
Verify with FDA Devices →
View my saved recalls

Medline Industries, LP recalled Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; Produ… - a critical-severity action.

Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; Produ… was recalled by Medline Industries, LP in November 19, 2025. Reason: Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cr…. Check the official notice for the remedy. Verify recall #Z-0442-2026 with the FDA Devices before acting.

The recall

Medline Industries, LP issued this critical-severity FDA Devices recall-Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cr….

Critical
severity level
Class I
classification
November 19, 2025
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-0442-2026 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices and 1 from Medline Industries, LP - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 41,021 medical devices recalls on record

01,0002,0003,0004,000 20052008201120142017202020232026 1,947
Severity1906864652High severity (most serious)Moderate severityLow severity
Where this agency-classified recall sits among FDA Class I/II/III notices

Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified high severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Product description

Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; Product description/Product numbers: ADULT 120 EXPANDABLE W/MASK DYNJAA10591; ANES CIRC 120IN EXP-LF DYNJAA6555G; ANES CIRCUIT ADULT 120"-LF DYNJAA10817; ANESTHESIA 120" ADLT-LF DYNJAA10775; ANESTHESIA CIRC ADULT 120-LF DYNJAA10601; ANESTHESIA CIRCUIT - LF DYNJAA10261B; ANESTHESIA CIRCUIT 120EXP-LF DYNJAA10741A; ANESTHESIA CIRCUIT ADULT-LF DYNJAA0380A DYNJAA10641 DYNJAA10706B DYNJAA10851 DYNJAA10922B DYNJAA10971A DYNJAA11030A DYNJAA11087 DYNJAA11126; ANESTHESIA CIRCUIT-LF DYNJAA0309A DYNJAA0368A DYNJAA10368A DYNJAA10392 DYNJAA10430A DYNJAA10547A DYNJAA10639A DYNJAA10656A DYNJAA10685A DYNJAA10685B; CIRC 120" EXP 1 BV HMEF 3L GSL DYNJAA10920; EXPANDABLE CIRCUIT-LF DYNJAA11095; NM ANESTHESIA TURNOVER KIT DYNJAA0433; PROV HEALTH HEPA CIRC #2-LF DYNJAA0360; SPTMPX01IA CIRCUIT ADULT-LF DYNJAA10973A;

Reason for recall

Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Critical (Class I)
Status Ongoing
Recall number Z-0442-2026
Date reported November 19, 2025
Date initiated October 10, 2025
Recalling firm Medline Industries, LP
Firm location Northfield, IL
Affected scope 96,406 total units
Distribution International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-0442-2026) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • This is a Class I (high-risk) recall, stop using the product immediately and follow the disposal or return instructions. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; Product description/Product numbers: ADULT 120 EXPANDABLE W/MASK DYNJAA10591; ANES CIRC 120IN EXP-LF DYNJAA6555G; ANES CIRCUIT ADULT 120"-LF DYNJAA10817; ANESTHESIA 120" ADLT-LF DYNJAA10775; ANESTHESIA CIRC ADULT 120-LF DYNJAA10601; ANESTHESIA CIRCUIT - LF DYNJAA10261B; ANESTHESIA CIRCUIT 120EXP-LF DYNJAA10741A; ANESTHESIA CIRCUIT ADULT-LF DYNJAA0380A DYNJAA10641 DYNJAA10706B DYNJAA10851 DYNJAA10922B DYNJAA10971A DYNJAA11030A DYNJAA11087 DYNJAA11126; ANESTHESIA CIRCUIT-LF DYNJAA0309A DYNJAA0368A DYNJAA10368A DYNJAA10392 DYNJAA10430A DYNJAA10547A DYNJAA10639A DYNJAA10656A DYNJAA10685A DYNJAA10685B; CIRC 120" EXP 1 BV HMEF 3L GSL DYNJAA10920; EXPANDABLE CIRCUIT-LF DYNJAA11095; NM ANESTHESIA TURNOVER KIT DYNJAA0433; PROV HEALTH HEPA CIRC #2-LF DYNJAA0360; SPTMPX01IA CIRCUIT ADULT-LF DYNJAA10973A;. Recalled by Medline Industries, LP. Units affected: 96,406 total units.
Why was this product recalled?
Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
Which agency issued this recall?
This recall was issued by the FDA Devices on November 19, 2025. Severity: Critical. Recall number: Z-0442-2026.
Where was the recalled product distributed?
Distribution: International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nationwide recalls with similar severity or category timing

Drug and device notices outside this firm and state, matched by severity timing and product category.

Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls

This recall: FDA Devices, reported November 19, 2025. Severity classes follow each agency's own taxonomy; related-recall context is computed across the full archive.

Methodology · Corrections · Archive as of August 2026