Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; Product description/Product numbers: ADULT 120 EXPANDABLE W/MASK DYNJAA10591; ANES CIRC 120IN EXP-LF DYNJAA6555G; ANES CIRCUIT ADULT 120"-LF DYNJAA10817; ANESTHESIA 120" ADLT-LF DYNJAA10775; ANESTHESIA CIRC ADULT 120-LF DYNJAA10601; ANESTHESIA CIRCUIT - LF DYNJAA10261B; ANESTHESIA CIRCUIT 120EXP-LF DYNJAA10741A; ANESTHESIA CIRCUIT ADULT-LF DYNJAA0380A DYNJAA10641 DYNJAA10706B DYNJAA10851 DYN
Product Description
Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; Product description/Product numbers: ADULT 120 EXPANDABLE W/MASK DYNJAA10591; ANES CIRC 120IN EXP-LF DYNJAA6555G; ANES CIRCUIT ADULT 120"-LF DYNJAA10817; ANESTHESIA 120" ADLT-LF DYNJAA10775; ANESTHESIA CIRC ADULT 120-LF DYNJAA10601; ANESTHESIA CIRCUIT - LF DYNJAA10261B; ANESTHESIA CIRCUIT 120EXP-LF DYNJAA10741A; ANESTHESIA CIRCUIT ADULT-LF DYNJAA0380A DYNJAA10641 DYNJAA10706B DYNJAA10851 DYNJAA10922B DYNJAA10971A DYNJAA11030A DYNJAA11087 DYNJAA11126; ANESTHESIA CIRCUIT-LF DYNJAA0309A DYNJAA0368A DYNJAA10368A DYNJAA10392 DYNJAA10430A DYNJAA10547A DYNJAA10639A DYNJAA10656A DYNJAA10685A DYNJAA10685B; CIRC 120" EXP 1 BV HMEF 3L GSL DYNJAA10920; EXPANDABLE CIRCUIT-LF DYNJAA11095; NM ANESTHESIA TURNOVER KIT DYNJAA0433; PROV HEALTH HEPA CIRC #2-LF DYNJAA0360; SPTMPX01IA CIRCUIT ADULT-LF DYNJAA10973A;
Reason for Recall
Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
Details
- Recalling Firm
- Medline Industries, LP
- Units Affected
- 96,406 total units
- Distribution
- International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;
- Location
- Northfield, IL
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11