PlainRecalls
FDA Devices Moderate Class II Terminated

MOSAIQ Oncology Information System. Used to manage workflows for treatment planning and delivery.

Reported: December 23, 2015 Initiated: November 24, 2015 #Z-0447-2016

Product Description

MOSAIQ Oncology Information System. Used to manage workflows for treatment planning and delivery.

Reason for Recall

First Databank (FDB) is a third party product incorporated into MOSAIQ. FDB routinely modifies drugs in their database to reflect market and formulary changes. In FDB Framework v2.0, those drugs are not being included in drug-drug, drug-allergy and duplicate drug therapy interaction checking.

Details

Recalling Firm
Elekta, Inc.
Units Affected
211
Distribution
Nationwide Distribution -- AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NJ, NM, NY, NC, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WY, & Puerto Rico.
Location
Atlanta, GA

Frequently Asked Questions

What product was recalled?
MOSAIQ Oncology Information System. Used to manage workflows for treatment planning and delivery.. Recalled by Elekta, Inc.. Units affected: 211.
Why was this product recalled?
First Databank (FDB) is a third party product incorporated into MOSAIQ. FDB routinely modifies drugs in their database to reflect market and formulary changes. In FDB Framework v2.0, those drugs are not being included in drug-drug, drug-allergy and duplicate drug therapy interaction checking.
Which agency issued this recall?
This recall was issued by the FDA Devices on December 23, 2015. Severity: Moderate. Recall number: Z-0447-2016.