NovoTHOR Gen 2.0 whole body red light therapy bed, model S1176
Reported: December 13, 2023 Initiated: September 13, 2022 #Z-0453-2024
Product Description
NovoTHOR Gen 2.0 whole body red light therapy bed, model S1176
Reason for Recall
Failure of ball stud components causing the gas strut on one side to detach from the canopy, making the canopy difficult to lift.
Details
- Recalling Firm
- Thor Photomedicine Ltd
- Units Affected
- 19
- Distribution
- Worldwide Distribution: US (Nationwide): Arizona, California, Colorado, Connecticut, Florida, Hawaii, Idaho, Illinois, Iowa, Maryland, Michigan, Minnesota, Missouri, Nevada, New Hampshire, North Carolina, Ohio, Oregon, Texas, Utah, Virginia, West Virginia. OUS (foreign) to countries of: Australia, Canada, Netherlands, New Zealand, Singapore, United Kingdom.
- Location
- Chesham
Frequently Asked Questions
What product was recalled? ▼
NovoTHOR Gen 2.0 whole body red light therapy bed, model S1176. Recalled by Thor Photomedicine Ltd. Units affected: 19.
Why was this product recalled? ▼
Failure of ball stud components causing the gas strut on one side to detach from the canopy, making the canopy difficult to lift.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on December 13, 2023. Severity: Moderate. Recall number: Z-0453-2024.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11