PlainRecalls
FDA Devices Moderate Class II Ongoing

B105P, B125P, B105M, B125M, B155M Patient Monitors; Model nos. 6160000-002 6160000-003 6160000-004 6160000-005 6160000-101 6160000-102 6160000-103 6160000-104 6160000-105 6160001-063 6160001-064 6160001-065 Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms

Reported: November 19, 2025 Initiated: September 26, 2025 #Z-0455-2026

Product Description

B105P, B125P, B105M, B125M, B155M Patient Monitors; Model nos. 6160000-002 6160000-003 6160000-004 6160000-005 6160000-101 6160000-102 6160000-103 6160000-104 6160000-105 6160001-063 6160001-064 6160001-065 Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms

Reason for Recall

Potential loss of monitoring on affected monitors when powered by certain batteries. Affected batteries can lose capability to hold charge and provide back up power. If this situation occurs, it will cause loss of patient monitoring and potentially delay recognizing patient status changes requiring treatment.

Details

Recalling Firm
GE Healthcare Finland Oy
Units Affected
7408
Distribution
Worldwide - US Nationwide distribution.
Location
Helsinki

Frequently Asked Questions

What product was recalled?
B105P, B125P, B105M, B125M, B155M Patient Monitors; Model nos. 6160000-002 6160000-003 6160000-004 6160000-005 6160000-101 6160000-102 6160000-103 6160000-104 6160000-105 6160001-063 6160001-064 6160001-065 Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms. Recalled by GE Healthcare Finland Oy. Units affected: 7408.
Why was this product recalled?
Potential loss of monitoring on affected monitors when powered by certain batteries. Affected batteries can lose capability to hold charge and provide back up power. If this situation occurs, it will cause loss of patient monitoring and potentially delay recognizing patient status changes requiring treatment.
Which agency issued this recall?
This recall was issued by the FDA Devices on November 19, 2025. Severity: Moderate. Recall number: Z-0455-2026.