Severity
Moderate
FDA Devices recall · Reported November 21, 2018
The firm is issuing this voluntary removal following receipt of reports of parastomal mesh failure identified several years following parastomal hernia repair using the modified S…
Covidien LLC recalled Parietex Composite Parastomal Mesh (Product Number PCOPM15 / PCOPM20) Product Usage: … - a moderate-severity action.
Parietex Composite Parastomal Mesh (Product Number PCOPM15 / PCOPM20) Product Usage: … was recalled by Covidien LLC in November 21, 2018. Reason: The firm is issuing this voluntary removal following receipt of reports of parastomal mesh failure identified…. Check the official notice for the remedy. Verify recall #Z-0461-2019 with the FDA Devices before acting.
The recall
Covidien LLC issued this moderate-severity FDA Devices recall-The firm is issuing this voluntary removal following receipt of reports of parastomal mesh failure identified….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0461-2019 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0461-2019) was formally reported on November 21, 2018, with the manufacturer initiating the action on October 5, 2018. It is classified under Moderate severity (Class II), with a current status of Terminated. Covidien LLC is listed as the recalling firm, operating out of North Haven, CT. Federal records list the affected scope as 7333.
The documented reason for this recall is: The firm is issuing this voluntary removal following receipt of reports of parastomal mesh failure identified several years following parastomal hernia repair using the modified Sugarbaker repair technique. In these rep… Distribution data in the federal record shows the product reached: Worldwide distribution--US Nationwide. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
7333
Related Recalls
6
3 from same agency
Parietex Composite Parastomal Mesh (Product Number PCOPM15 / PCOPM20) Product Usage: The PCOPM15 and PCOPM20 reorder codes are circular in shape, made from a three-dimensional monofilament polyester fabric with a two-dimensional monofilament polyester central band. One side of the design is completely covered with the hydrophilic film. On the opposite side, only the two-dimensional central band is coated with the absorbable hydrophilic film. The film extends 5 mm over the external edge of the reinforcement. The Parietex composite parastomal mesh is indicated for the reinforcement of soft tissues during surgical repair, and specifically for the repair of parastomal hernias. The non-absorbable polyester mesh provides long term reinforcement of soft tissues. The absorbable hydrophilic film minimizes tissue attachment to the mesh when in direct contact with the viscera.
The firm is issuing this voluntary removal following receipt of reports of parastomal mesh failure identified several years following parastomal hernia repair using the modified Sugarbaker repair technique. In these reports, product failure led to hernia recurrence requiring additional surgical treatment. Symptoms of hernia recurrence may include discomfort, localized pain-free or painful bulging, and possible changes in the overlying skin. Medtronic has received, worldwide, a total of ten reports of mesh failure following use of the product in the last five years. Patients who have received a Parietex" composite parastomal mesh for the treatment of a parastomal hernia should continue to receive ongoing monitoring by their healthcare providers for the recurrence of a parastomal hernia.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0461-2019 |
| Date reported | November 21, 2018 |
| Date initiated | October 5, 2018 |
| Recalling firm | Covidien LLC |
| Firm location | North Haven, CT |
| Affected scope | 7333 |
| Distribution | Worldwide distribution--US Nationwide |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported November 21, 2018. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.