Dimension Vista Lipase Flex Reagent Cartridge, Model Number 10461745 Product Usage: Dimension Vista - The LIPL method is an in vitro diagnostic test for the quantitative measurement of lipase in human serum and plasma on the Dimension Vista System.
Reported: November 21, 2018 Initiated: September 7, 2018 #Z-0466-2019
Product Description
Dimension Vista Lipase Flex Reagent Cartridge, Model Number 10461745 Product Usage: Dimension Vista - The LIPL method is an in vitro diagnostic test for the quantitative measurement of lipase in human serum and plasma on the Dimension Vista System.
Reason for Recall
There is lot to lot variability for QC and patient results when using the identified reagent cartridge lots.
Details
- Recalling Firm
- Siemens Healthcare Diagnostics, Inc.
- Units Affected
- 3956
- Distribution
- Worldwide - US Nationwide Distributions The products were distributed to the following foreign countries: Argentina, Australia, Bangladesh, Brazil, Canada, Chile, China, Colombia, Germany, Ecuador, Honduras, Indonesia, India, Japan, Republic of Korea, Mexico, Malaysia, Nicaragua, New Zealand, Peru, Philippines, Paraguay, El Salvador, Thailand, Taiwan, Uruguay and South Africa, Austria, Bahamas, Belgium, Czech Republic, Denmark, , France, , Italy, Netherland, Norway , Portugal, Kuwait, Slovenia, Spain, Switzerland, United Kingdom, Qatar, Republic of Korea, Saudi Arabia, Slovakia, Algeria, Bahrain, Bulgaria, Egypt, , Hungary, Latvia, Lebanon, Lithuania, Macedonia, Nepal, Poland, Romania, Russia Fed, Serbia, Turkey, U.A.E, United Kingdom.
- Location
- Newark, DE
Frequently Asked Questions
What product was recalled? ▼
Dimension Vista Lipase Flex Reagent Cartridge, Model Number 10461745 Product Usage: Dimension Vista - The LIPL method is an in vitro diagnostic test for the quantitative measurement of lipase in human serum and plasma on the Dimension Vista System.. Recalled by Siemens Healthcare Diagnostics, Inc.. Units affected: 3956.
Why was this product recalled? ▼
There is lot to lot variability for QC and patient results when using the identified reagent cartridge lots.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on November 21, 2018. Severity: Moderate. Recall number: Z-0466-2019.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11