FDA Devices Critical Class I Terminated

BrightMatter Guide, Product Number SYN-0026, Versions: 1.5.0, 1.5.1 and 1.5.2 Product Usage: BrightMatter Guide can navigate a third-party port (NICO BrainPath device) by attaching Synaptive s custom Tracking Array to the port directly. As the BrainPath is navigable in two configurations (sheath only / sheath and obturator), the tip of the obturator is represented in the software with a tip extension beyond the bottom of the sheath. Note sheath and cannula are used interchangeably in this do

Reported: December 5, 2018 Initiated: October 9, 2018 #Z-0470-2019

Product Description

BrightMatter Guide, Product Number SYN-0026, Versions: 1.5.0, 1.5.1 and 1.5.2 Product Usage: BrightMatter Guide can navigate a third-party port (NICO BrainPath device) by attaching Synaptive s custom Tracking Array to the port directly. As the BrainPath is navigable in two configurations (sheath only / sheath and obturator), the tip of the obturator is represented in the software with a tip extension beyond the bottom of the sheath. Note sheath and cannula are used interchangeably in this document.

Reason for Recall

This recall has been initiated due to a software defect found in the Guide System software when used with NICO BrainPath ports under certain circumstances. When one trajectory is set with a blue port and another trajectory is set with a gold port, the software defect is triggered when the user switches between these trajectories during a surgical procedure.

Details

Recalling Firm
Synaptive Medical Inc
Units Affected
37 units
Distribution
Worldwide - US Nationwide Distribution and the country of Canada
Location
Toronto, N/A

Frequently Asked Questions

What product was recalled?
BrightMatter Guide, Product Number SYN-0026, Versions: 1.5.0, 1.5.1 and 1.5.2 Product Usage: BrightMatter Guide can navigate a third-party port (NICO BrainPath device) by attaching Synaptive s custom Tracking Array to the port directly. As the BrainPath is navigable in two configurations (sheath only / sheath and obturator), the tip of the obturator is represented in the software with a tip extension beyond the bottom of the sheath. Note sheath and cannula are used interchangeably in this document.. Recalled by Synaptive Medical Inc. Units affected: 37 units.
Why was this product recalled?
This recall has been initiated due to a software defect found in the Guide System software when used with NICO BrainPath ports under certain circumstances. When one trajectory is set with a blue port and another trajectory is set with a gold port, the software defect is triggered when the user switches between these trajectories during a surgical procedure.
Which agency issued this recall?
This recall was issued by the FDA Devices on December 5, 2018. Severity: Critical. Recall number: Z-0470-2019.