Severity
Moderate
FDA Devices recall · Reported November 25, 2020
Stryker has discovered a characterization issue associated with the Mako Integrated Cutting System (MICS) Handpiece. The MICS Handpiece is attached to the distal end of the robot…
Mako Surgical Corporation recalled Mako Integrated Cutting System (MICS) Handpiece, Catalog Number 209063, Software Version#… - a moderate-severity action.
Mako Integrated Cutting System (MICS) Handpiece, Catalog Number 209063, Software Version#… was recalled by Mako Surgical Corporation in November 25, 2020. Reason: Stryker has discovered a characterization issue associated with the Mako Integrated Cutting System (MICS) Han…. Check the official notice for the remedy. Verify recall #Z-0472-2021 with the FDA Devices before acting.
The recall
Mako Surgical Corporation issued this moderate-severity FDA Devices recall-Stryker has discovered a characterization issue associated with the Mako Integrated Cutting System (MICS) Han….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0472-2021 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0472-2021) was formally reported on November 25, 2020, with the manufacturer initiating the action on October 23, 2020. It is classified under Moderate severity (Class II), with a current status of Completed. Mako Surgical Corporation is listed as the recalling firm, operating out of Davie, FL. Federal records list the affected scope as 367 units.
The documented reason for this recall is: Stryker has discovered a characterization issue associated with the Mako Integrated Cutting System (MICS) Handpiece. The MICS Handpiece is attached to the distal end of the robotic arm and is used to register the RIO a… Distribution data in the federal record shows the product reached: Worldwide distribution - US Nationwide distribution and the countries of Australia, Japan, Korea, Netherlands, and Chile, and France, Germany Luxembourg, and UK. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
367 units
Related Recalls
6
3 from same agency
Mako Integrated Cutting System (MICS) Handpiece, Catalog Number 209063, Software Version#: THA 3.0 and later, PKA 3.0 and later, and TKA 1.0 - Product Usage: The MICS Handpiece is attached to the distal end of the robotic arm and is used to register the RIO (Mako robot) and to execute bone preparation during a Mako Total Knee or Mako Partial Knee procedure (PKA Software Application 3.0 or later).
Stryker has discovered a characterization issue associated with the Mako Integrated Cutting System (MICS) Handpiece. The MICS Handpiece is attached to the distal end of the robotic arm and is used to register the RIO and to execute bone preparation during a Mako Total Knee or Mako Partial Knee procedure (PKA Software Application 3.0 or later). This issue may result in unsuccessful checkpoint verification when the Mako System attempts to verify the location of the saw blade prior to bone preparation. This issue may also cause a discrepancy during bone preparation in a Mako Total Knee or a Mako Partial Knee procedure in which the surgeon uses a straight or angled saw attachment with the MICS Handpiece. The surgeon may make cuts during bone preparation that differ from the surgeon s preoperative plan.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Completed |
| Recall number | Z-0472-2021 |
| Date reported | November 25, 2020 |
| Date initiated | October 23, 2020 |
| Recalling firm | Mako Surgical Corporation |
| Firm location | Davie, FL |
| Affected scope | 367 units |
| Distribution | Worldwide distribution - US Nationwide distribution and the countries of Australia, Japan, Korea, Netherlands, and Chile, and France, Germany Luxembourg, and UK |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported November 25, 2020. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.