PlainRecalls
FDA Devices Moderate Class II Terminated

MEDICOM SAFEMASK PREMIER PLUS ASTM LEVEL 2 MASKS, Item Code 2040

Reported: November 27, 2019 Initiated: October 11, 2019 #Z-0494-2020

Product Description

MEDICOM SAFEMASK PREMIER PLUS ASTM LEVEL 2 MASKS, Item Code 2040

Reason for Recall

Due to an error on the packaging line, some level 2 masks may have been packed in some level 3 mask boxes. The two masks are visually indistinguishable, causing the risk of using a mask with a lower protection level than intended.

Details

Recalling Firm
AMD Medicom Inc.
Units Affected
756 cases (500 units/case)
Distribution
The products were distributed tot he following US states: CA, FL, IA, IN, MD, MI, NV, NY, PA, SC, TN, TX, and WA. The products were distributed to the following foreign countries: Canada.
Location
Lachine, N/A

Frequently Asked Questions

What product was recalled?
MEDICOM SAFEMASK PREMIER PLUS ASTM LEVEL 2 MASKS, Item Code 2040. Recalled by AMD Medicom Inc.. Units affected: 756 cases (500 units/case).
Why was this product recalled?
Due to an error on the packaging line, some level 2 masks may have been packed in some level 3 mask boxes. The two masks are visually indistinguishable, causing the risk of using a mask with a lower protection level than intended.
Which agency issued this recall?
This recall was issued by the FDA Devices on November 27, 2019. Severity: Moderate. Recall number: Z-0494-2020.