WECK Hem-O-Lok AutoEndo5, 5mm Automatic Endoscopie Hem-o-lok Applier, REF 543965 Automatic Endoscopic Surgical Ligating Clip Appliers are delivery devices for Hem-o-lok ML non-absorbable polymer ligating clips
Reported: November 27, 2019 Initiated: October 4, 2019 #Z-0497-2020
Product Description
WECK Hem-O-Lok AutoEndo5, 5mm Automatic Endoscopie Hem-o-lok Applier, REF 543965 Automatic Endoscopic Surgical Ligating Clip Appliers are delivery devices for Hem-o-lok ML non-absorbable polymer ligating clips
Reason for Recall
Certain clip appliers have an increase in misloading and/or jamming related complaints that may be due to a discrepancy in the specification a component. Misloading and/or jamming may lead to clips falling out of the applier, clip breakage, or the jaws becoming locked in a partially closed position (jammed).
Details
- Recalling Firm
- Teleflex Medical
- Units Affected
- 4099 devices
- Distribution
- Worldwide distribution, including US nationwide (including Puerto Rico).
- Location
- Morrisville, NC
Frequently Asked Questions
What product was recalled? ▼
WECK Hem-O-Lok AutoEndo5, 5mm Automatic Endoscopie Hem-o-lok Applier, REF 543965 Automatic Endoscopic Surgical Ligating Clip Appliers are delivery devices for Hem-o-lok ML non-absorbable polymer ligating clips. Recalled by Teleflex Medical. Units affected: 4099 devices.
Why was this product recalled? ▼
Certain clip appliers have an increase in misloading and/or jamming related complaints that may be due to a discrepancy in the specification a component. Misloading and/or jamming may lead to clips falling out of the applier, clip breakage, or the jaws becoming locked in a partially closed position (jammed).
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on November 27, 2019. Severity: Moderate. Recall number: Z-0497-2020.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11