Severity
Moderate
FDA Devices recall · Reported November 27, 2019
Certain clip appliers have an increase in misloading and/or jamming related complaints that may be due to a discrepancy in the specification a component. Misloading and/or jamming…
Teleflex Medical recalled WECK Hem-O-Lok AutoEndo5, 5mm Automatic Endoscopie Hem-o-lok Applier, REF 543965 Au… - a moderate-severity action.
WECK Hem-O-Lok AutoEndo5, 5mm Automatic Endoscopie Hem-o-lok Applier, REF 543965 Au… was recalled by Teleflex Medical in November 27, 2019. Reason: Certain clip appliers have an increase in misloading and/or jamming related complaints that may be due to a d…. Check the official notice for the remedy. Verify recall #Z-0497-2020 with the FDA Devices before acting.
The recall
Teleflex Medical issued this moderate-severity FDA Devices recall-Certain clip appliers have an increase in misloading and/or jamming related complaints that may be due to a d….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0497-2020 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0497-2020) was formally reported on November 27, 2019, with the manufacturer initiating the action on October 4, 2019. It is classified under Moderate severity (Class II), with a current status of Ongoing. Teleflex Medical is listed as the recalling firm, operating out of Morrisville, NC. Federal records list the affected scope as 4099 devices.
The documented reason for this recall is: Certain clip appliers have an increase in misloading and/or jamming related complaints that may be due to a discrepancy in the specification a component. Misloading and/or jamming may lead to clips falling out of the ap… Distribution data in the federal record shows the product reached: Worldwide distribution, including US nationwide (including Puerto Rico).. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
4099 devices
Related Recalls
6
3 from same agency
WECK Hem-O-Lok AutoEndo5, 5mm Automatic Endoscopie Hem-o-lok Applier, REF 543965 Automatic Endoscopic Surgical Ligating Clip Appliers are delivery devices for Hem-o-lok ML non-absorbable polymer ligating clips
Certain clip appliers have an increase in misloading and/or jamming related complaints that may be due to a discrepancy in the specification a component. Misloading and/or jamming may lead to clips falling out of the applier, clip breakage, or the jaws becoming locked in a partially closed position (jammed).
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-0497-2020 |
| Date reported | November 27, 2019 |
| Date initiated | October 4, 2019 |
| Recalling firm | Teleflex Medical |
| Firm location | Morrisville, NC |
| Affected scope | 4099 devices |
| Distribution | Worldwide distribution, including US nationwide (including Puerto Rico). |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported November 27, 2019. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.