PlainRecalls
FDA Devices Moderate Class II Ongoing

1. Sensis, material # 10764561; UDI: 04056869010137 2. Sensis Vibe Hemo, material # 11007641; UDI: 04056869010199 3. Sensis Vibe Combo, material # 11007642; UDI: 04056869010205 The Sensis recording system is a diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies, for cardiac, interventional Radiology, and surgical studies.

Reported: January 26, 2022 Initiated: November 23, 2021 #Z-0499-2022

Product Description

1. Sensis, material # 10764561; UDI: 04056869010137 2. Sensis Vibe Hemo, material # 11007641; UDI: 04056869010199 3. Sensis Vibe Combo, material # 11007642; UDI: 04056869010205 The Sensis recording system is a diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies, for cardiac, interventional Radiology, and surgical studies.

Reason for Recall

Software error which affects Sensis Vibe Hemo, Sensis and Sensis Vibe Combo systems with software version VD12A, which could cause possible hazard to patients, operators, or other persons and equipment. Under certain sporadic circumstances, the CO (Cardiac Output) measurement using the Thermodilution method will temporarily no longer be possible.

Details

Units Affected
1047
Distribution
US Nationwide distribution in the states of AK, AL, AR, AZ, CA, FL, GU, HI, IA, IL, IN, KS, KY, MA, MI, MO, MS, NC, NE, NJ, NY, OH, OK,PA, PR, SD, TN, TX, WA, WV.
Location
Malvern, PA

Frequently Asked Questions

What product was recalled?
1. Sensis, material # 10764561; UDI: 04056869010137 2. Sensis Vibe Hemo, material # 11007641; UDI: 04056869010199 3. Sensis Vibe Combo, material # 11007642; UDI: 04056869010205 The Sensis recording system is a diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies, for cardiac, interventional Radiology, and surgical studies.. Recalled by Siemens Medical Solutions USA, Inc. Units affected: 1047.
Why was this product recalled?
Software error which affects Sensis Vibe Hemo, Sensis and Sensis Vibe Combo systems with software version VD12A, which could cause possible hazard to patients, operators, or other persons and equipment. Under certain sporadic circumstances, the CO (Cardiac Output) measurement using the Thermodilution method will temporarily no longer be possible.
Which agency issued this recall?
This recall was issued by the FDA Devices on January 26, 2022. Severity: Moderate. Recall number: Z-0499-2022.