Siemens Healthcare ADVIA Centaur Digitoxin (DGTN) 250 Test 09046419 50 Test 08861968 Intended Use For in vitro diagnostic use in the quantitative determination of digitoxin in serum or plasma using the ADVIA Centaur and ADVIA Centaur XP systems.

Reported: December 19, 2012 Initiated: November 14, 2012 #Z-0510-2013

Product Description

Siemens Healthcare ADVIA Centaur Digitoxin (DGTN) 250 Test 09046419 50 Test 08861968 Intended Use For in vitro diagnostic use in the quantitative determination of digitoxin in serum or plasma using the ADVIA Centaur and ADVIA Centaur XP systems.

Reason for Recall

Operating range of the ADVIA Centaur systems is 18¿C to 30¿C (64.4¿F to 86.0¿F), a change in room temperature may increase or decrease assay results, depending on the specific assay and ADVIA Centaur system used.

Details

Units Affected
50 test=792 and 250 test=525
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to the following countries: Argentina Australia Austria Azerbaijan Belgium Bosnia Herzeg. Brazil Canada Chile COLOMBIA Croatia Czech Republic Denmark Egypt Estonia Finland France Germany Greece Hong Kong Hungary Iceland India Indonesia Iran Ireland Italy Japan Kazakhstan Latvia Lithuania Malaysia Mexico Netherlands New Caledonia Norway Oman P.R. China Philippines Poland Portugal Qatar Rep. of Yemen Republic Korea Reunion Romania Russian Fed. Saudi Arabia Serbia Slovakia Slovenia South Africa Spain Sweden Switzerland Taiwan Thailand Turkey U.A.E. United Kingdom Uruguay Vietnam
Location
East Walpole, MA

Frequently Asked Questions

What product was recalled?
Siemens Healthcare ADVIA Centaur Digitoxin (DGTN) 250 Test 09046419 50 Test 08861968 Intended Use For in vitro diagnostic use in the quantitative determination of digitoxin in serum or plasma using the ADVIA Centaur and ADVIA Centaur XP systems.. Recalled by Siemens Healthcare Diagnostics, Inc. Units affected: 50 test=792 and 250 test=525.
Why was this product recalled?
Operating range of the ADVIA Centaur systems is 18¿C to 30¿C (64.4¿F to 86.0¿F), a change in room temperature may increase or decrease assay results, depending on the specific assay and ADVIA Centaur system used.
Which agency issued this recall?
This recall was issued by the FDA Devices on December 19, 2012. Severity: Moderate. Recall number: Z-0510-2013.