Siemens Healthcare ADVIA Centaur Digitoxin (DGTN) 250 Test 09046419 50 Test 08861968 Intended Use For in vitro diagnostic use in the quantitative determination of digitoxin in serum or plasma using the ADVIA Centaur and ADVIA Centaur XP systems.
Reported: December 19, 2012 Initiated: November 14, 2012 #Z-0510-2013
Product Description
Siemens Healthcare ADVIA Centaur Digitoxin (DGTN) 250 Test 09046419 50 Test 08861968 Intended Use For in vitro diagnostic use in the quantitative determination of digitoxin in serum or plasma using the ADVIA Centaur and ADVIA Centaur XP systems.
Reason for Recall
Operating range of the ADVIA Centaur systems is 18¿C to 30¿C (64.4¿F to 86.0¿F), a change in room temperature may increase or decrease assay results, depending on the specific assay and ADVIA Centaur system used.
Details
- Recalling Firm
- Siemens Healthcare Diagnostics, Inc
- Units Affected
- 50 test=792 and 250 test=525
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to the following countries: Argentina Australia Austria Azerbaijan Belgium Bosnia Herzeg. Brazil Canada Chile COLOMBIA Croatia Czech Republic Denmark Egypt Estonia Finland France Germany Greece Hong Kong Hungary Iceland India Indonesia Iran Ireland Italy Japan Kazakhstan Latvia Lithuania Malaysia Mexico Netherlands New Caledonia Norway Oman P.R. China Philippines Poland Portugal Qatar Rep. of Yemen Republic Korea Reunion Romania Russian Fed. Saudi Arabia Serbia Slovakia Slovenia South Africa Spain Sweden Switzerland Taiwan Thailand Turkey U.A.E. United Kingdom Uruguay Vietnam
- Location
- East Walpole, MA
Frequently Asked Questions
What product was recalled? ▼
Siemens Healthcare ADVIA Centaur Digitoxin (DGTN) 250 Test 09046419 50 Test 08861968 Intended Use For in vitro diagnostic use in the quantitative determination of digitoxin in serum or plasma using the ADVIA Centaur and ADVIA Centaur XP systems.. Recalled by Siemens Healthcare Diagnostics, Inc. Units affected: 50 test=792 and 250 test=525.
Why was this product recalled? ▼
Operating range of the ADVIA Centaur systems is 18¿C to 30¿C (64.4¿F to 86.0¿F), a change in room temperature may increase or decrease assay results, depending on the specific assay and ADVIA Centaur system used.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on December 19, 2012. Severity: Moderate. Recall number: Z-0510-2013.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11