Severity
Moderate
FDA Devices recall · Reported December 25, 2013
Siemens has confirmed that the TACR method may demonstrate reduced on-board stability which may result in imprecise and inaccurate QC and patient results.
Siemens Healthcare Diagnostics, Inc. recalled Dimension(R) TACR Flex(R) reagent cartridge (DF107) The TACR method is an in vitro dia… - a moderate-severity action.
Dimension(R) TACR Flex(R) reagent cartridge (DF107) The TACR method is an in vitro dia… was recalled by Siemens Healthcare Diagnostics, Inc. in December 25, 2013. Reason: Siemens has confirmed that the TACR method may demonstrate reduced on-board stability which may result in imp…. Check the official notice for the remedy. Verify recall #Z-0517-2014 with the FDA Devices before acting.
The recall
Siemens Healthcare Diagnostics, Inc. issued this moderate-severity FDA Devices recall-Siemens has confirmed that the TACR method may demonstrate reduced on-board stability which may result in imp….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0517-2014 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0517-2014) was formally reported on December 25, 2013, with the manufacturer initiating the action on April 25, 2013. It is classified under Moderate severity (Class II), with a current status of Terminated. Siemens Healthcare Diagnostics, Inc. is listed as the recalling firm, operating out of Newark, DE. Federal records list the affected scope as 13,751.
The documented reason for this recall is: Siemens has confirmed that the TACR method may demonstrate reduced on-board stability which may result in imprecise and inaccurate QC and patient results. Distribution data in the federal record shows the product reached: Nationwide Distribution including AK, AL, AR, AZ, CA, DC, FL, GA, IA, IL, LA, MD, MI, MO, MS, NC, NH, NJ, NM, OH, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, and WI.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
13,751
Related Recalls
6
3 from same agency
Dimension(R) TACR Flex(R) reagent cartridge (DF107) The TACR method is an in vitro diagnostic test intended to quantitatively measure Tacrolimus in human whole blood on the Dimension(R) clinical chemistry system. Measurements of Tacrolimus are used as an aid in the management of Tacrolimus therapy in kidney and liver transplant patients
Siemens has confirmed that the TACR method may demonstrate reduced on-board stability which may result in imprecise and inaccurate QC and patient results.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0517-2014 |
| Date reported | December 25, 2013 |
| Date initiated | April 25, 2013 |
| Recalling firm | Siemens Healthcare Diagnostics, Inc. |
| Firm location | Newark, DE |
| Affected scope | 13,751 |
| Distribution | Nationwide Distribution including AK, AL, AR, AZ, CA, DC, FL, GA, IA, IL, LA, MD, MI, MO, MS, NC, NH, NJ, NM, OH, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, and WI. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported December 25, 2013. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.