FDA Devices recall · Reported December 19, 2012
It has been determined that cobas b 123 Fluid Pack COOX cuvettes for lot numbers 21426121 and 21426171, exhibit a significant increase in the out-of-box failure rate caused by a c…
Roche Diagnostics Operations, Inc. recalled Cobas b 123 POC system cobas b 123 POC system is fully automated POC system for whole … - a moderate-severity action.
Cobas b 123 POC system cobas b 123 POC system is fully automated POC system for whole … was recalled by Roche Diagnostics Operations, Inc. in December 19, 2012. Reason: It has been determined that cobas b 123 Fluid Pack COOX cuvettes for lot numbers 21426121 and 21426171, exhib…. Check the official notice for the remedy. Verify recall #Z-0522-2013 with the FDA Devices before acting.
The recall
Roche Diagnostics Operations, Inc. issued this moderate-severity FDA Devices recall-It has been determined that cobas b 123 Fluid Pack COOX cuvettes for lot numbers 21426121 and 21426171, exhib….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0522-2013 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0522-2013) was formally reported on December 19, 2012, with the manufacturer initiating the action on November 14, 2012. It is classified under Moderate severity (Class II), with a current status of Terminated. Roche Diagnostics Operations, Inc. is listed as the recalling firm, operating out of Indianapolis, IN. Federal records list the affected scope as 23 units of Fluid Pack COOX, 200 were distributed.
The documented reason for this recall is: It has been determined that cobas b 123 Fluid Pack COOX cuvettes for lot numbers 21426121 and 21426171, exhibit a significant increase in the out-of-box failure rate caused by a change in the thickness in the cuvette la… Distribution data in the federal record shows the product reached: Nationwide Distribution including NE and AZ.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
23 units of Fluid Pack COOX, 200 were distributed
Related Recalls
6
3 from same agency
Cobas b 123 POC system cobas b 123 POC system is fully automated POC system for whole blood in vitro measurement of pH, blood gases (BG), electrolytes Na+ K+, iCa2+ (ISE), hematocrit (Hct), metabolites (Gluc, Lac), total hemoglobin (tHb), hemoglobin derivatives (O2Hb, HHB, COHB, MetHb), and oxygen saturation (SO2). In addition, the cobas b 123 POC system calculates derived parameters. It is dedicated for use in a Point-of-Care environment and laboratory.
It has been determined that cobas b 123 Fluid Pack COOX cuvettes for lot numbers 21426121 and 21426171, exhibit a significant increase in the out-of-box failure rate caused by a change in the thickness in the cuvette layer. Internal investigations found the altered thickness of the cuvette was due to production equipment settings. Cuvettes were produced within specifications, but within the upper
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0522-2013 |
| Date reported | December 19, 2012 |
| Date initiated | November 14, 2012 |
| Recalling firm | Roche Diagnostics Operations, Inc. |
| Firm location | Indianapolis, IN |
| Affected scope | 23 units of Fluid Pack COOX, 200 were distributed |
| Distribution | Nationwide Distribution including NE and AZ. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported December 19, 2012. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.