HA Pins (DFS HA Cort Screw), Hoffman/Colles Pins (Colles CSER Sterile w/ Pin Kit) Item Nos: HAA60-09030 HAA60-09040 HAA60-10030 HAA60-10040 HAA60-11030 HAA60-11040 HAA60-11050 HAA60-12040 HAA60-12060 HAA60-13030 HAA60-13040 HAA60-13050 HAA60-13060 HAA60-13070 HAA60-14050 HAA60-14080 HAA60-14090 HAA60-15020 HAA60-15030 HAA60-15040 HAA60-15050 HAA60-15060 HAA60-16030 HAA60-16040 HAA60-16050 HAA60-16070 HAA60-16090 HAA60-17040 HAA60-17060 HAA60-17080 HAA60-18010 HA

Reported: December 4, 2019 Initiated: August 21, 2019 #Z-0528-2020 178886 units

Zimmer Biomet, Inc. issued this FDA Devices recall on December 4, 2019. Classified as Moderate severity (Class II). Approximately 178886 units are affected. The recall was issued because: Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue.. This recall notice is sourced from official FDA Devices enforcement records. Below you will find the complete product description, hazard information, remedy instructions, and related recalls from the same manufacturer or product category.

Recall Insight

This FDA Devices action (record #Z-0528-2020) was formally reported on December 4, 2019, with the manufacturer initiating the action on August 21, 2019. It is classified under Moderate severity (Class II), with a current status of Terminated. Zimmer Biomet, Inc. is listed as the recalling firm, operating out of Warsaw, IN. Federal records indicate 178886 units are affected, a scale large enough to require multi-state distribution tracking.

The documented reason for this recall is: Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue. Distribution data in the federal record shows the product reached: US Nationwide distribution and countries of Argentina, Australia, Brazil, Canada, Chile, Colombia, Costa Rica, Ecuador, El Salvador, Italy, Netherlands, India, Japan, Korea, Malaysia, Mexico, New Zealand, Panama, Singap…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

To put this record in context, PlainRecalls indexes 83,949 recalls across the FDA, CPSC, NHTSA and USDA FSIS going back to 1995. Within the same product category, the database holds 6 closely related recalls, of which 6 were also issued by FDA Devices. That clustering is a signal — repeated actions in a narrow category often indicate a systemic quality-control issue, a supplier-wide contamination, or a design defect that has propagated across product lines. This recall is roughly 7 years old; older recalls can remain relevant because many units enter resale, rental, and secondary-market channels where the original warning never reaches the end user. Always cross-check the recall number against the official agency page before relying on any summary.

Recall Distribution by Severity Class

Severity1Class I (Critical)Class II (Moderate)Class III (Low)
Recall Distribution by Severity Class

Severity

Moderate

Units Affected

178886

Related Recalls

6

6 from same agency

Recall Progress (industry avg ~60%) 60.0%

Product Description

HA Pins (DFS HA Cort Screw), Hoffman/Colles Pins (Colles CSER Sterile w/ Pin Kit) Item Nos: HAA60-09030 HAA60-09040 HAA60-10030 HAA60-10040 HAA60-11030 HAA60-11040 HAA60-11050 HAA60-12040 HAA60-12060 HAA60-13030 HAA60-13040 HAA60-13050 HAA60-13060 HAA60-13070 HAA60-14050 HAA60-14080 HAA60-14090 HAA60-15020 HAA60-15030 HAA60-15040 HAA60-15050 HAA60-15060 HAA60-16030 HAA60-16040 HAA60-16050 HAA60-16070 HAA60-16090 HAA60-17040 HAA60-17060 HAA60-17080 HAA60-18010 HAA60-18030 HAA60-18040 HAA60-18050 HAA60-18060 HAA60-18090 HAA60-20030 HAA60-20040 HAA60-20050 HAA60-20060 HAA60-20080 HAA60-20090 HAA60-22050 HAA60-22060 HAA60-25030 HAA60-25040 HAA60-25050 HAA60-25060 HAA60-30060 HAB60-09030 HAB60-10030 HAB60-10040 HAB60-11040 HAB60-11050 HAB60-12040 HAB60-12060 HAB60-13040 HAB60-13050 HAB60-13060 HAB60-14050 HAB60-15060 HAB60-16070 HAB60-16090 HAB60-17080 HAB60-18010 HAB60-18090 HAB60-20080 HAB60-20090 808200000 Product Usage: These devices are intended for osteosynthesis, external fixation of fractures or bone lengthening procedures.

Reason for Recall

Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue.

Details

Recalling Firm
Zimmer Biomet, Inc.
Units Affected
178886
Distribution
US Nationwide distribution and countries of Argentina, Australia, Brazil, Canada, Chile, Colombia, Costa Rica, Ecuador, El Salvador, Italy, Netherlands, India, Japan, Korea, Malaysia, Mexico, New Zealand, Panama, Singapore, Taiwan, Thailand, Trinidad and Tobago, and Venezuela.
Location
Warsaw, IN

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Terminated
Recall number Z-0528-2020
Date reported December 4, 2019
Date initiated August 21, 2019
Recalling firm Zimmer Biomet, Inc.
Units affected 178886
Distribution US Nationwide distribution and countries of Argentina, Australia, Brazil, Canada, Chile, Colombia, Costa Rica, Ecuador, El Salvador, Italy, Netherlands, India, Japan, Korea, Malaysia, Mexico, New Zealand, Panama, Singapore, Taiwan, Thailan…

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

178886 units affected — multi-state distribution scale.

Regional (<10K units)
Multi-state (10K – 100K units)
Large-scale (100K – 1M units) ✓ This recall
Massive (≥1M units)

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

Frequently Asked Questions

What product was recalled?
HA Pins (DFS HA Cort Screw), Hoffman/Colles Pins (Colles CSER Sterile w/ Pin Kit) Item Nos: HAA60-09030 HAA60-09040 HAA60-10030 HAA60-10040 HAA60-11030 HAA60-11040 HAA60-11050 HAA60-12040 HAA60-12060 HAA60-13030 HAA60-13040 HAA60-13050 HAA60-13060 HAA60-13070 HAA60-14050 HAA60-14080 HAA60-14090 HAA60-15020 HAA60-15030 HAA60-15040 HAA60-15050 HAA60-15060 HAA60-16030 HAA60-16040 HAA60-16050 HAA60-16070 HAA60-16090 HAA60-17040 HAA60-17060 HAA60-17080 HAA60-18010 HAA60-18030 HAA60-18040 HAA60-18050 HAA60-18060 HAA60-18090 HAA60-20030 HAA60-20040 HAA60-20050 HAA60-20060 HAA60-20080 HAA60-20090 HAA60-22050 HAA60-22060 HAA60-25030 HAA60-25040 HAA60-25050 HAA60-25060 HAA60-30060 HAB60-09030 HAB60-10030 HAB60-10040 HAB60-11040 HAB60-11050 HAB60-12040 HAB60-12060 HAB60-13040 HAB60-13050 HAB60-13060 HAB60-14050 HAB60-15060 HAB60-16070 HAB60-16090 HAB60-17080 HAB60-18010 HAB60-18090 HAB60-20080 HAB60-20090 808200000 Product Usage: These devices are intended for osteosynthesis, external fixation of fractures or bone lengthening procedures.. Recalled by Zimmer Biomet, Inc.. Units affected: 178886.
Why was this product recalled?
Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue.
Which agency issued this recall?
This recall was issued by the FDA Devices on December 4, 2019. Severity: Moderate. Recall number: Z-0528-2020.
Where was the recalled product distributed?
Distribution: US Nationwide distribution and countries of Argentina, Australia, Brazil, Canada, Chile, Colombia, Costa Rica, Ecuador, El Salvador, Italy, Netherlands, India, Japan, Korea, Malaysia, Mexico, New Zealand, Panama, Singapore, Taiwan, Thailand, Trinidad and Tobago, and Venezuela..
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-0528-2020) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Recall Context

Product recalls are issued when a manufacturer, distributor, or federal agency determines that a product poses a safety risk to consumers. This recall is classified as moderate severity, indicating the product may cause temporary or medically reversible health consequences. Across PlainRecalls, we track 83,000+ recalls from FDA, CPSC, and NHTSA to help consumers stay informed and act quickly when safety issues arise.

Nearby Recalls in This Category

Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Endo-Model Replacement Plateau; Item Number: 15-0027/11; →

Data Sources

Data as of 2025. Source: FDA, CPSC, NHTSA, USDA FSIS federal recall databases.

  • Source: FDA — Food and Drug Administration, openFDA Enforcement API (food, drug, and medical device recalls)
  • Source: CPSC — Consumer Product Safety Commission Recalls API (consumer product recalls and hazards)
  • Source: NHTSA — National Highway Traffic Safety Administration Recalls API (vehicle safety recalls)
  • Source: USDA FSIS — Food Safety and Inspection Service (meat, poultry, and egg product recalls)

Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.

All federal data sources used on this page

Source: Federal recall agencies (FDA, CPSC, NHTSA, USDA FSIS) Aggregated multi-agency recall feeds · 2024 Recall data normalized across federal agency feeds; severity classifications follow each agency's own taxonomy (FDA Class I/II/III; CPSC, NHTSA, USDA FSIS).