FDA Devices Moderate Class II Ongoing

Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL

Reported: December 4, 2024 Initiated: October 9, 2024 #Z-0554-2025

Product Description

Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL

Reason for Recall

There have been cases of failure of the main board due to condensation water occurring in the gap between thermal insulation material on the underside of the main base dropping onto the main board The symptoms and errors differ depending on the position where the water drops adhere. To date there have been examples of reports of inter-board communication errors, HOME POSITION error or SLIP error of XYZ drive units. The failure could result in delay of patient sample testing.

Details

Units Affected
48
Distribution
Worldwide - US Nationwide distribution in the states of DE, PA and the country of DE.
Location
Kyoto

Frequently Asked Questions

What product was recalled?
Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL. Recalled by Shimadzu Corp. Analytical & Measuring Instruments Div.. Units affected: 48.
Why was this product recalled?
There have been cases of failure of the main board due to condensation water occurring in the gap between thermal insulation material on the underside of the main base dropping onto the main board The symptoms and errors differ depending on the position where the water drops adhere. To date there have been examples of reports of inter-board communication errors, HOME POSITION error or SLIP error of XYZ drive units. The failure could result in delay of patient sample testing.
Which agency issued this recall?
This recall was issued by the FDA Devices on December 4, 2024. Severity: Moderate. Recall number: Z-0554-2025.