Severity
Moderate
FDA Devices recall · Reported February 9, 2022
The firm has received six (6) complaint reports from customers indicating difficulty removing the balloon protector sleeve from the catheter (e.g., excessive force required to rem…
Covidien Llc recalled Chameleon PTA Balloon Dilation Catheter 8 mm x 40 mm, Chameleon PTA Balloon Catheter with… - a moderate-severity action.
Chameleon PTA Balloon Dilation Catheter 8 mm x 40 mm, Chameleon PTA Balloon Catheter with… was recalled by Covidien Llc in February 9, 2022. Reason: The firm has received six (6) complaint reports from customers indicating difficulty removing the balloon pro…. Check the official notice for the remedy. Verify recall #Z-0557-2022 with the FDA Devices before acting.
The recall
Covidien Llc issued this moderate-severity FDA Devices recall-The firm has received six (6) complaint reports from customers indicating difficulty removing the balloon pro….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0557-2022 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0557-2022) was formally reported on February 9, 2022, with the manufacturer initiating the action on December 17, 2021. It is classified under Moderate severity (Class II), with a current status of Ongoing. Covidien Llc is listed as the recalling firm, operating out of Mansfield, MA. Federal records list the affected scope as 1,994 devices.
The documented reason for this recall is: The firm has received six (6) complaint reports from customers indicating difficulty removing the balloon protector sleeve from the catheter (e.g., excessive force required to remove the protector sleeve). If the balloo… Distribution data in the federal record shows the product reached: Worldwide distribution - US Nationwide and the countries of Germany, Israel, Italy, Portugal, Spain, Chile, Singapore, and China.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
1,994 devices
Related Recalls
6
3 from same agency
Chameleon PTA Balloon Dilation Catheter 8 mm x 40 mm, Chameleon PTA Balloon Catheter with Injection Port 8 mm x 40 mm; Cardiovascular. Product Numbers CH08-40-75GB and CH08-40-75US
The firm has received six (6) complaint reports from customers indicating difficulty removing the balloon protector sleeve from the catheter (e.g., excessive force required to remove the protector sleeve). If the balloon is damaged while removing the protector sleeve and the catheter is subsequently used in a patient procedure, patient harm may potentially occur, including perforation of vessels, foreign body in patient, embolism, risks associated with surgical intervention, or delay to treatment/therapy.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-0557-2022 |
| Date reported | February 9, 2022 |
| Date initiated | December 17, 2021 |
| Recalling firm | Covidien Llc |
| Firm location | Mansfield, MA |
| Affected scope | 1,994 devices |
| Distribution | Worldwide distribution - US Nationwide and the countries of Germany, Israel, Italy, Portugal, Spain, Chile, Singapore, and China. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported February 9, 2022. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.