Severity
Moderate
FDA Devices recall · Reported December 16, 2020
Potential for Transcatheter Aortic Valve (TAV) leaflet damage when performing a post-implant balloon dilatation (PID).
Medtronic Heart Valves Division recalled Evolut (TM) R System. With the following model numbers: a. EVOLUTR-23, Labeled as V… - a moderate-severity action.
Evolut (TM) R System. With the following model numbers: a. EVOLUTR-23, Labeled as V… was recalled by Medtronic Heart Valves Division in December 16, 2020. Reason: Potential for Transcatheter Aortic Valve (TAV) leaflet damage when performing a post-implant balloon dilatati…. Check the official notice for the remedy. Verify recall #Z-0559-2021 with the FDA Devices before acting.
The recall
Medtronic Heart Valves Division issued this moderate-severity FDA Devices recall-Potential for Transcatheter Aortic Valve (TAV) leaflet damage when performing a post-implant balloon dilatati….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0559-2021 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0559-2021) was formally reported on December 16, 2020, with the manufacturer initiating the action on October 23, 2020. It is classified under Moderate severity (Class II), with a current status of Terminated. Medtronic Heart Valves Division is listed as the recalling firm, operating out of Santa Ana, CA. Federal records list the affected scope as 172,851 units, a scale large enough to require multi-state distribution tracking.
The documented reason for this recall is: Potential for Transcatheter Aortic Valve (TAV) leaflet damage when performing a post-implant balloon dilatation (PID). Distribution data in the federal record shows the product reached: Worldwide distribution - U.S. Nationwide distribution and the countries of Panama, Denmark, Austria, Germany, Finland, Sweden, Colombia, India, Mexico, Brazil, Spain, United Kingdom, Australia, Ireland, Italy, Argentina…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,422 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
172,851 units
Related Recalls
6
0 from same agency
Evolut (TM) R System. With the following model numbers: a. EVOLUTR-23, Labeled as VLV EVOLUTR-23 VT IR 6L COMM OUS, VLV EVOLUTR-23 TAV EVOLUT R CE 29L MX, VLV EVOLUTR-23 TAV CE AU 2Y MX, VLV EVOLUTR-23 TAV JPN MX, VLV EVOLUTR-23 GREY 29L COMM OUS, VLV EVOLUTR-23 TAV EVOLUT R CE 7L MX, VLV EVOLUTR-23 TAV EVOLUT R CE 26L MX, VLV EVOLUTR-23 TAV CA 2Y MX, VLV EVOLUTR-23 GREY 6L COMM OUS, VLV EVOLUTR-23-US TAV 23 US COMM MX; b. EVOLUTR-23-C, Labeled as VLV EVOLUTR-23-C TAV EVOLUT R US IDE MEX; c. EVOLUTR-23-US, Labeled as VLV EVOLUTR-23-US GREY COMM US, VLV EVOLUTR-23-US TAV 23 US COMM MX; d. EVOLUTR-26, Labeled as VLV EVOLUTR-26 TAV EVOLUT R CE 29L MX VLV EVOLUTR-26 VT IR 6L COMM OUS, VLV EVOLUTR-26 TAV EVOLUT R CE 7L MX, LV EVOLUTR-26 TAV CE AU 2Y MX, VLV EVOLUTR-26 TAV JPN MX, VLV EVOLUTR-26 GREY 29L COMM OUS, VLV EVOLUTR-26-US TAV 26 US COMM MX, VLV EVOLUTR-26 TAV CLN OUS MX, VLV EVOLUTR-26 TAV CA 2Y MX, VLV EVOLUTR-26 TAV CE CA 2Y MX, VLV EVOLUTR-26 GREY 6L COMM OUS; e. EVOLUTR-26-C, Labeled as VLV EVOLUTR-26-C TAV EVOLUT R US IDE MEX, VLV EVOLUTR-26-C TAV CLN US MX; f. EVOLUTR-26-US, Labeled as VLV EVOLUTR-26 TAV EVOLUT R CE 29L MX, VLV EVOLUTR-26-US GREY COMM US, VLV EVOLUTR-26-US TAV 26 US COMM MX; g. EVOLUTR-29, Labeled as VLV EVOLUTR-29 TAV EVOLUT R CE 29L MX, VLV EVOLUTR-29 GREY 6L COMM OUS, VLV EVOLUTR-29 VT IR 6L COMM OUS, VLV EVOLUTR-29 TAV CE AU 2Y MX, VLV EVOLUTR-29 TAV JPN MX, VLV EVOLUTR-29 GREY 29L COMM OUS, VLV EVOLUTR-29 TAV EVOLUT R CE 7L MX, VLV EVOLUTR-29 TAV CLN OUS MX, VLV EVOLUTR-29 TAV CA 2Y MX, VLV EVOLUTR-29 TAV CE CA 2Y MX, VLV EVOLUTR-29-US TAV 29 US COMM MX; h. EVOLUTR-29-C, Labeled as VLV EVOLUTR-29-C TAV EVOLUT R US IDE MEX, LV EVOLUTR-29-C TAV CLN US MX, VLV EVOLUTR-29-US TAV 29 US COMM MX; i. EVOLUTR-29-US, Labeled as VLV EVOLUTR-29-US GREY COMM US, VLV EVOLUTR-29-US TAV 29 US COMM MX, VLV EVOLUTR-29 TAV EVOLUT R CE 29L MX; j. EVOLUTR-34, Labeled as VLV EVOLUTR-34 BLUE 29L COMM OUS, VLV EVOLUTR-34 VT IR 6L COMM OUS, VLV EVOLUTR-34 BLUE 6L COMM OUS, VLV EVOLUTR-34 TAV CE AU 2Y MX, VLV EVOLUTR-34 TAV JPN MX, VLV EVOLUTR-34 TAV CLN OUS MX, VLV EVOLUTR-34 TAV CA 2Y MX, VLV EVOLUTR-34 TAV CE CA 2Y MX, VLV EVOLUTR-34 TAV EVOLUT R CE 29L 1Y MX, VLV EVOLUTR-34 TAV EVO R CE 29L 2Y MX SA, DRAPE 9732722 TUBE STERILE O-ARM 20PK; k. EVOLUTR-34-C, Labeled as VLV EVOLUTR-34-C TAV EVOLUT R US IDE MX, VLV EVOLUTR-34-C TAV CLN US MX; l. EVOLUTR-34-US, Labeled as VLV EVOLUTR-34-US BLUE COMM US, VLV EVOLUTR-34-US TAV 34 US COMM MX, VLV EVOLUTR-34 TAV EVOLUT R CE 29L 1Y MX, VLV EVOLUTR-34-US TAV 34 US COMM MX SA.
Potential for Transcatheter Aortic Valve (TAV) leaflet damage when performing a post-implant balloon dilatation (PID).
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0559-2021 |
| Date reported | December 16, 2020 |
| Date initiated | October 23, 2020 |
| Recalling firm | Medtronic Heart Valves Division |
| Firm location | Santa Ana, CA |
| Affected scope | 172,851 units |
| Distribution | Worldwide distribution - U.S. Nationwide distribution and the countries of Panama, Denmark, Austria, Germany, Finland, Sweden, Colombia, India, Mexico, Brazil, Spain, United Kingdom, Australia, Ireland, Italy, Argentina, Japan, Poland, Rus… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Keep tracking product safety across the federal recall archive.
Recall Checker
Search the full archive by product name, brand, or recall number across every agency.
Check a product →
FDA Devices recalls
Every recall issued by FDA Devices, newest first.
Browse the feed →
RecallRadar
Live feed of the latest recalls across the FDA, CPSC and NHTSA, filter by agency and severity.
View the live feed →
Rankings
The largest recalls by units affected and the most-recalled product categories.
See the rankings →
Browse by category
Find recalls by product type to spot recurring defect patterns.
All categories →
What to do next
A step-by-step guide to refunds, repairs, and returns after a recall.
Read the guide →
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-07-02
· 2026-07-02
· 2026-06-25
· 2026-06-25
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Source: FDA, CPSC, and NHTSA federal recall databases. This recall: FDA Devices, reported December 16, 2020.
Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.