Evolut (TM) PRO System. With the following model numbers: a. EVOLUTPRO-23, Labeled as: VLV EVOLUTPRO-23 PRO COMM 29L OUS, VLV EVOLUTPRO-23 BLUE 29L COMM OUS, VLV EVOLUTPRO-23 BLUE COMM JPN, VLV EVOLUTPRO-23 BLUE 6L COMM OUS, VLV EVOLUTPRO-23 PRO COMM 29L CAN, VLV EVOLUTPRO-23 PRO COMM 29L OUS SA, VLV EVOLUTPRO-23 PRO COMM 29L AUS; b. EVOLUTPRO-23-US, Labeled as: VLV EVOLUTPRO-23-US BLUE COMM US, VLV EVOLUTPRO-23-US PRO BLNK US; c. EVOLUTPRO-26, Labeled as: VLV EVOLUTPRO-26 PRO CO

Reported: December 16, 2020 Initiated: October 23, 2020 #Z-0560-2021

Product Description

Evolut (TM) PRO System. With the following model numbers: a. EVOLUTPRO-23, Labeled as: VLV EVOLUTPRO-23 PRO COMM 29L OUS, VLV EVOLUTPRO-23 BLUE 29L COMM OUS, VLV EVOLUTPRO-23 BLUE COMM JPN, VLV EVOLUTPRO-23 BLUE 6L COMM OUS, VLV EVOLUTPRO-23 PRO COMM 29L CAN, VLV EVOLUTPRO-23 PRO COMM 29L OUS SA, VLV EVOLUTPRO-23 PRO COMM 29L AUS; b. EVOLUTPRO-23-US, Labeled as: VLV EVOLUTPRO-23-US BLUE COMM US, VLV EVOLUTPRO-23-US PRO BLNK US; c. EVOLUTPRO-26, Labeled as: VLV EVOLUTPRO-26 PRO COMM 29L OUS, VLV EVOLUTPRO-26 BLUE 6L COMM OUS, VLV EVOLUTPRO-26 PRO COMM 29L AUS, VLV EVOLUTPRO-26 BLUE COMM JPN, VLV EVOLUTPRO-26 BLUE 29L COMM OUS, VLV EVOLUTPRO-26 PRO COMM 29L CAN, VLV EVOLUTPRO-26 PRO COMM 29L OUS SA; d. EVOLUTPRO-26-US, Labeled as: VLV EVOLUTPRO-26-US BLUE COMM US, VLV EVOLUTPRO-26-US PRO BLNK US; e. EVOLUTPRO-29, Labeled as: VLV EVOLUTPRO-29 PRO COMM 29L OUS, VLV EVOLUTPRO-29 BLUE 29L COMM OUS, VLV EVOLUTPRO-29 BLUE 6L COMM OUS, VLV EVOLUTPRO-29 PRO COMM 29L AUS, VLV EVOLUTPRO-29 BLUE COMM JPN, VLV EVOLUTPRO-29 PRO COMM 29L CAN, VLV EVOLUTPRO-29 PRO COMM 29L OUS SA; f. EVOLUTPRO-29-US, Labeled as: VLV EVOLUTPRO-29-US BLUE COMM US, VLV EVOLUTPRO-29-US PRO BLNK US g. , Labeled as:

Reason for Recall

Potential for Transcatheter Aortic Valve (TAV) leaflet damage when performing a post-implant balloon dilatation (PID).

Details

Units Affected
71,030 units
Distribution
Worldwide distribution - U.S. Nationwide distribution and the countries of Panama, Denmark, Austria, Germany, Finland, Sweden, Colombia, India, Mexico, Brazil, Spain, United Kingdom, Australia, Ireland, Italy, Argentina, Japan, Poland, Russian Federation, Hungary, Kazakhstan, Luxembourg, Portugal, Norway, Egypt, Saudi Arabia.
Location
Santa Ana, CA

Frequently Asked Questions

What product was recalled?
Evolut (TM) PRO System. With the following model numbers: a. EVOLUTPRO-23, Labeled as: VLV EVOLUTPRO-23 PRO COMM 29L OUS, VLV EVOLUTPRO-23 BLUE 29L COMM OUS, VLV EVOLUTPRO-23 BLUE COMM JPN, VLV EVOLUTPRO-23 BLUE 6L COMM OUS, VLV EVOLUTPRO-23 PRO COMM 29L CAN, VLV EVOLUTPRO-23 PRO COMM 29L OUS SA, VLV EVOLUTPRO-23 PRO COMM 29L AUS; b. EVOLUTPRO-23-US, Labeled as: VLV EVOLUTPRO-23-US BLUE COMM US, VLV EVOLUTPRO-23-US PRO BLNK US; c. EVOLUTPRO-26, Labeled as: VLV EVOLUTPRO-26 PRO COMM 29L OUS, VLV EVOLUTPRO-26 BLUE 6L COMM OUS, VLV EVOLUTPRO-26 PRO COMM 29L AUS, VLV EVOLUTPRO-26 BLUE COMM JPN, VLV EVOLUTPRO-26 BLUE 29L COMM OUS, VLV EVOLUTPRO-26 PRO COMM 29L CAN, VLV EVOLUTPRO-26 PRO COMM 29L OUS SA; d. EVOLUTPRO-26-US, Labeled as: VLV EVOLUTPRO-26-US BLUE COMM US, VLV EVOLUTPRO-26-US PRO BLNK US; e. EVOLUTPRO-29, Labeled as: VLV EVOLUTPRO-29 PRO COMM 29L OUS, VLV EVOLUTPRO-29 BLUE 29L COMM OUS, VLV EVOLUTPRO-29 BLUE 6L COMM OUS, VLV EVOLUTPRO-29 PRO COMM 29L AUS, VLV EVOLUTPRO-29 BLUE COMM JPN, VLV EVOLUTPRO-29 PRO COMM 29L CAN, VLV EVOLUTPRO-29 PRO COMM 29L OUS SA; f. EVOLUTPRO-29-US, Labeled as: VLV EVOLUTPRO-29-US BLUE COMM US, VLV EVOLUTPRO-29-US PRO BLNK US g. , Labeled as:. Recalled by Medtronic Heart Valves Division. Units affected: 71,030 units.
Why was this product recalled?
Potential for Transcatheter Aortic Valve (TAV) leaflet damage when performing a post-implant balloon dilatation (PID).
Which agency issued this recall?
This recall was issued by the FDA Devices on December 16, 2020. Severity: Moderate. Recall number: Z-0560-2021.