Tracheostomy care kits: (1) Medline Tracheostomy Clean and Care Tray with 14 Fr Suction Catheter, REF DYND40580, Sterile; (2) Medline Tracheostomy Clean and Care Tray, REF DYND4061030, Sterile; (3) Medline Basic Tracheostomy Care Kit, REF CC3T4691A, Sterile; (4) Medline Tracheostomy Care Kit with Glove, REF CC4681A, Sterile; (5) Medline Mini Tracheostomy Care Kit, REF CC3T3030A, Sterile; and (6) Medline Mini Tracheostomy Care Kit, REF DYND40622, Non-sterile.

Reported: January 3, 2024 Initiated: October 16, 2023 #Z-0561-2024 620,735 kits units

MEDLINE INDUSTRIES, LP - Northfield issued this FDA Devices recall on January 3, 2024. Classified as Critical severity (Class I). Approximately 620,735 kits units are affected. The recall was issued because: The component tracheostomy brush has a sharp edge at the tip that can potentially puncture tubing during use and/or cau…. This recall notice is sourced from official FDA Devices enforcement records. Below you will find the complete product description, hazard information, remedy instructions, and related recalls from the same manufacturer or product category.

Recall Insight

This FDA Devices action (record #Z-0561-2024) was formally reported on January 3, 2024, with the manufacturer initiating the action on October 16, 2023. It is classified under Critical severity (Class I), with a current status of Ongoing. MEDLINE INDUSTRIES, LP - Northfield is listed as the recalling firm, operating out of Northfield, IL. Federal records indicate 620,735 kits units are affected, a scale large enough to require multi-state distribution tracking.

The documented reason for this recall is: The component tracheostomy brush has a sharp edge at the tip that can potentially puncture tubing during use and/or cause user injury. Additionally, there have been incidents of the tracheostomy brush bristles detachin… Distribution data in the federal record shows the product reached: US Nationwide distribution including Puerto Rico. OUS (Foreign) distribution was made to Canada, Panama, and United Arab Emirates.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

To put this record in context, PlainRecalls indexes 83,949 recalls across the FDA, CPSC, NHTSA and USDA FSIS going back to 1995. Within the same product category, the database holds 6 closely related recalls, of which 6 were also issued by FDA Devices. That clustering is a signal — repeated actions in a narrow category often indicate a systemic quality-control issue, a supplier-wide contamination, or a design defect that has propagated across product lines. This recall is roughly 2 years old; older recalls can remain relevant because many units enter resale, rental, and secondary-market channels where the original warning never reaches the end user. Always cross-check the recall number against the official agency page before relying on any summary.

Recall Distribution by Severity Class

Severity1Class I (Critical)Class II (Moderate)Class III (Low)
Recall Distribution by Severity Class

Severity

Critical

Units Affected

620,735 kits

Related Recalls

6

6 from same agency

Product Description

Tracheostomy care kits: (1) Medline Tracheostomy Clean and Care Tray with 14 Fr Suction Catheter, REF DYND40580, Sterile; (2) Medline Tracheostomy Clean and Care Tray, REF DYND4061030, Sterile; (3) Medline Basic Tracheostomy Care Kit, REF CC3T4691A, Sterile; (4) Medline Tracheostomy Care Kit with Glove, REF CC4681A, Sterile; (5) Medline Mini Tracheostomy Care Kit, REF CC3T3030A, Sterile; and (6) Medline Mini Tracheostomy Care Kit, REF DYND40622, Non-sterile.

Reason for Recall

The component tracheostomy brush has a sharp edge at the tip that can potentially puncture tubing during use and/or cause user injury. Additionally, there have been incidents of the tracheostomy brush bristles detaching prior to use in the packaging and/or during use.

Details

Units Affected
620,735 kits
Distribution
US Nationwide distribution including Puerto Rico. OUS (Foreign) distribution was made to Canada, Panama, and United Arab Emirates.
Location
Northfield, IL

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Critical (Class I)
Status Ongoing
Recall number Z-0561-2024
Date reported January 3, 2024
Date initiated October 16, 2023
Recalling firm MEDLINE INDUSTRIES, LP - Northfield
Units affected 620,735 kits
Distribution US Nationwide distribution including Puerto Rico. OUS (Foreign) distribution was made to Canada, Panama, and United Arab Emirates.

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

620,735 kits units affected — multi-state distribution scale.

Regional (<10K units)
Multi-state (10K – 100K units)
Large-scale (100K – 1M units) ✓ This recall
Massive (≥1M units)

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

Frequently Asked Questions

What product was recalled?
Tracheostomy care kits: (1) Medline Tracheostomy Clean and Care Tray with 14 Fr Suction Catheter, REF DYND40580, Sterile; (2) Medline Tracheostomy Clean and Care Tray, REF DYND4061030, Sterile; (3) Medline Basic Tracheostomy Care Kit, REF CC3T4691A, Sterile; (4) Medline Tracheostomy Care Kit with Glove, REF CC4681A, Sterile; (5) Medline Mini Tracheostomy Care Kit, REF CC3T3030A, Sterile; and (6) Medline Mini Tracheostomy Care Kit, REF DYND40622, Non-sterile.. Recalled by MEDLINE INDUSTRIES, LP - Northfield. Units affected: 620,735 kits.
Why was this product recalled?
The component tracheostomy brush has a sharp edge at the tip that can potentially puncture tubing during use and/or cause user injury. Additionally, there have been incidents of the tracheostomy brush bristles detaching prior to use in the packaging and/or during use.
Which agency issued this recall?
This recall was issued by the FDA Devices on January 3, 2024. Severity: Critical. Recall number: Z-0561-2024.
Where was the recalled product distributed?
Distribution: US Nationwide distribution including Puerto Rico. OUS (Foreign) distribution was made to Canada, Panama, and United Arab Emirates..
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-0561-2024) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Recall Context

Product recalls are issued when a manufacturer, distributor, or federal agency determines that a product poses a safety risk to consumers. This recall is classified as critical severity, meaning the product carries a reasonable probability of serious adverse health consequences or death. Across PlainRecalls, we track 83,000+ recalls from FDA, CPSC, and NHTSA to help consumers stay informed and act quickly when safety issues arise.

Nearby Recalls in This Category

Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Endo-Model Replacement Plateau; Item Number: 15-0027/11; →

Data Sources

Data as of 2025. Source: FDA, CPSC, NHTSA, USDA FSIS federal recall databases.

  • Source: FDA — Food and Drug Administration, openFDA Enforcement API (food, drug, and medical device recalls)
  • Source: CPSC — Consumer Product Safety Commission Recalls API (consumer product recalls and hazards)
  • Source: NHTSA — National Highway Traffic Safety Administration Recalls API (vehicle safety recalls)
  • Source: USDA FSIS — Food Safety and Inspection Service (meat, poultry, and egg product recalls)

Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.

All federal data sources used on this page

Source: Federal recall agencies (FDA, CPSC, NHTSA, USDA FSIS) Aggregated multi-agency recall feeds · 2024 Recall data normalized across federal agency feeds; severity classifications follow each agency's own taxonomy (FDA Class I/II/III; CPSC, NHTSA, USDA FSIS).