FDA Devices recall · Reported December 26, 2012
Medtronic, DBS Extension Kit for Deep Brain Stimulation, Catalog # 7482, 37085, 37086. Rx Only, Manufactured in: Medtronic, Inc. Villalba, Puerto Rico. Medtronic¿ DBSTM Therapy for Parkinson's Disease - Bilateral stimulation of the internal globus pallidus (GPi) or the subthalamic nucleus (STN) using Medtronic DBS Therapy for Parkinson's Disease is indicated for adjunctive therapy in reducing some of the symptoms of advanced, levodopa-responsive Parkinson's disease that are not adequately co
Medtronic has identified 140 Restore DBS Extension Kits (model numbers 7482, 37085, and 37086), which may contain a damaged winged connector boot. The connector boot may not have …
Medtronic Neuromodulation recalled Medtronic, DBS Extension Kit for Deep Brain Stimulation, Catalog # 7482, 37085, 37086. Rx… - a moderate-severity action.
Medtronic, DBS Extension Kit for Deep Brain Stimulation, Catalog # 7482, 37085, 37086. Rx… was recalled by Medtronic Neuromodulation in December 26, 2012. Reason: Medtronic has identified 140 Restore DBS Extension Kits (model numbers 7482, 37085, and 37086), which may con…. Check the official notice for the remedy. Verify recall #Z-0567-2013 with the FDA Devices before acting.
The recall
Medtronic Neuromodulation issued this moderate-severity FDA Devices recall-Medtronic has identified 140 Restore DBS Extension Kits (model numbers 7482, 37085, and 37086), which may con….
- Moderate
- severity level
- 140 units
- affected scope
- Class II
- classification
- December 26, 2012
- reported
Sourced from official FDA Devices enforcement records. Verify recall #Z-0567-2013 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
Recall Insight
This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Medical Devices recalls over time
Where this recall sits in its category - 41,021 medical devices recalls on record
Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Product description
Medtronic, DBS Extension Kit for Deep Brain Stimulation, Catalog # 7482, 37085, 37086. Rx Only, Manufactured in: Medtronic, Inc. Villalba, Puerto Rico. Medtronic¿ DBSTM Therapy for Parkinson's Disease - Bilateral stimulation of the internal globus pallidus (GPi) or the subthalamic nucleus (STN) using Medtronic DBS Therapy for Parkinson's Disease is indicated for adjunctive therapy in reducing some of the symptoms of advanced, levodopa-responsive Parkinson's disease that are not adequately controlled with medication. STN/GPi target system components: ¿ Kinetra¿ Model 7428 or Soletra¿ Model 7426 Neurostimulator ¿ Activa¿ PC Model 37601, Activa RC Model 37612, Activa SC Model 37602, or Activa SC Model 37603 Neurostimulator ¿ Model 7482, Model 7482A, Model 7483, Model 37085, or Model 37086 Extension ¿ Model 3387 or Model 3389 Lead Medtronic DBS Therapy for Tremor - Unilateral thalamic stimulation of the ventral intermediate nucleus (VIM) using Medtronic DBS Therapy for Tremor is indicated for the suppression of tremor in the upper extremity. The system is intended for patients who are diagnosed with essential tremor or parkinsonian tremor not adequately controlled by medications and where the tremor constitutes a significant functional disability. VIM target system components: ¿ Soletra Model 7426 Neurostimulator ¿ Activa PC Model 37601, Activa RC Model 37612, Activa SC Model 37602, or Activa SC Model 37603 Neurostimulator ¿ Model 7482, Model 7482A, Model 7483, Model 37085, or Model 37086 Extension ¿ Model 3387 or Model 3389 Lead.
Reason for recall
Medtronic has identified 140 Restore DBS Extension Kits (model numbers 7482, 37085, and 37086), which may contain a damaged winged connector boot. The connector boot may not have been molded properly and may be cracked. Approximately 14% of the 140 affected extension kits may contain a damaged radiopaque winged connector boot.
Recall Profile
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0567-2013 |
| Date reported | December 26, 2012 |
| Date initiated | December 3, 2012 |
| Recalling firm | Medtronic Neuromodulation |
| Firm location | Minneapolis, MN |
| Affected scope | 140 units |
| Distribution | Worldwide Distribution -- USA (nationwide) including the states of CT, MI, NH, NY, OH, PA and VT., and the countries of AUSTRALIA, JAPAN, BELGIUM, FRANCE, GERMANY, ITALY, NETHERLANDS, POLAND, SPAIN, SWITZERLAND, TURKEY and UNITED KINGD… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
- Check the recall number (Z-0567-2013) and product description against the item you own. Search the archive
- Confirm the current status with FDA Devices before acting, recall details can be updated.
- Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Where was the recalled product distributed? ▼
Should I stop using a recalled medication or medical device? ▼
Nationwide recalls with similar severity or category timing
Drug and device notices outside this firm and state, matched by severity timing and product category.
Similar severity · nearest report date
Same archive severity grade (Moderate), closest by report date among other firms and states.
Similar Medical Devices · nearest report date
Same product category, closest by report date among other firms and states (deduped against the severity set).
Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls
Methodology · Corrections · Archive as of August 2026