PlainRecalls
ModerateClass IITerminated

FDA Devices recall · Reported December 26, 2012

Medtronic, DBS Extension Kit for Deep Brain Stimulation, Catalog # 7482, 37085, 37086. Rx Only, Manufactured in: Medtronic, Inc. Villalba, Puerto Rico. Medtronic¿ DBSTM Therapy for Parkinson's Disease - Bilateral stimulation of the internal globus pallidus (GPi) or the subthalamic nucleus (STN) using Medtronic DBS Therapy for Parkinson's Disease is indicated for adjunctive therapy in reducing some of the symptoms of advanced, levodopa-responsive Parkinson's disease that are not adequately co

Medtronic has identified 140 Restore DBS Extension Kits (model numbers 7482, 37085, and 37086), which may contain a damaged winged connector boot. The connector boot may not have …

Recall #
Z-0567-2013
Affected scope
140 units
Initiated
December 3, 2012
Compiled from official public sources by the editorial team.
Verify with FDA Devices →
View my saved recalls

Medtronic Neuromodulation recalled Medtronic, DBS Extension Kit for Deep Brain Stimulation, Catalog # 7482, 37085, 37086. Rx… - a moderate-severity action.

Medtronic, DBS Extension Kit for Deep Brain Stimulation, Catalog # 7482, 37085, 37086. Rx… was recalled by Medtronic Neuromodulation in December 26, 2012. Reason: Medtronic has identified 140 Restore DBS Extension Kits (model numbers 7482, 37085, and 37086), which may con…. Check the official notice for the remedy. Verify recall #Z-0567-2013 with the FDA Devices before acting.

The recall

Medtronic Neuromodulation issued this moderate-severity FDA Devices recall-Medtronic has identified 140 Restore DBS Extension Kits (model numbers 7482, 37085, and 37086), which may con….

Moderate
severity level
140 units
affected scope
Class II
classification
December 26, 2012
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-0567-2013 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This FDA Devices action (record #Z-0567-2013) was formally reported on December 26, 2012, with the manufacturer initiating the action on December 3, 2012. It is classified under Moderate severity (Class II), with a current status of Terminated. Medtronic Neuromodulation is listed as the recalling firm, operating out of Minneapolis, MN. Federal records list the affected scope as 140 units.

The documented reason for this recall is: Medtronic has identified 140 Restore DBS Extension Kits (model numbers 7482, 37085, and 37086), which may contain a damaged winged connector boot. The connector boot may not have been molded properly and may be cracked.… Distribution data in the federal record shows the product reached: Worldwide Distribution -- USA (nationwide) including the states of CT, MI, NH, NY, OH, PA and VT., and the countries of AUSTRALIA, JAPAN, BELGIUM, FRANCE, GERMANY, ITALY, NETHERLANDS, POLAND, SPAIN, SWITZERLAND, TURK…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 40,322 medical devices recalls on record

-1,00001,0002,0003,0004,000 20052008201120142017202020232026 1,248
Severity2487872426High severity (most serious)Moderate severityLow severity
Where this recall sits in the database

Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Severity

Moderate

Affected scope

140 units

Related Recalls

6

3 from same agency

Product description

Medtronic, DBS Extension Kit for Deep Brain Stimulation, Catalog # 7482, 37085, 37086. Rx Only, Manufactured in: Medtronic, Inc. Villalba, Puerto Rico. Medtronic¿ DBSTM Therapy for Parkinson's Disease - Bilateral stimulation of the internal globus pallidus (GPi) or the subthalamic nucleus (STN) using Medtronic DBS Therapy for Parkinson's Disease is indicated for adjunctive therapy in reducing some of the symptoms of advanced, levodopa-responsive Parkinson's disease that are not adequately controlled with medication. STN/GPi target system components: ¿ Kinetra¿ Model 7428 or Soletra¿ Model 7426 Neurostimulator ¿ Activa¿ PC Model 37601, Activa RC Model 37612, Activa SC Model 37602, or Activa SC Model 37603 Neurostimulator ¿ Model 7482, Model 7482A, Model 7483, Model 37085, or Model 37086 Extension ¿ Model 3387 or Model 3389 Lead Medtronic DBS Therapy for Tremor - Unilateral thalamic stimulation of the ventral intermediate nucleus (VIM) using Medtronic DBS Therapy for Tremor is indicated for the suppression of tremor in the upper extremity. The system is intended for patients who are diagnosed with essential tremor or parkinsonian tremor not adequately controlled by medications and where the tremor constitutes a significant functional disability. VIM target system components: ¿ Soletra Model 7426 Neurostimulator ¿ Activa PC Model 37601, Activa RC Model 37612, Activa SC Model 37602, or Activa SC Model 37603 Neurostimulator ¿ Model 7482, Model 7482A, Model 7483, Model 37085, or Model 37086 Extension ¿ Model 3387 or Model 3389 Lead.

Reason for recall

Medtronic has identified 140 Restore DBS Extension Kits (model numbers 7482, 37085, and 37086), which may contain a damaged winged connector boot. The connector boot may not have been molded properly and may be cracked. Approximately 14% of the 140 affected extension kits may contain a damaged radiopaque winged connector boot.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Terminated
Recall number Z-0567-2013
Date reported December 26, 2012
Date initiated December 3, 2012
Recalling firm Medtronic Neuromodulation
Firm location Minneapolis, MN
Affected scope 140 units
Distribution Worldwide Distribution -- USA (nationwide) including the states of CT, MI, NH, NY, OH, PA and VT., and the countries of AUSTRALIA, JAPAN, BELGIUM, FRANCE, GERMANY, ITALY, NETHERLANDS, POLAND, SPAIN, SWITZERLAND, TURKEY and UNITED KINGD…

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

140 units units affected - limited or regional distribution scale.

Regional (<10K units) ✓ This recall
Multi-state (10K – 100K units) -
Large-scale (100K – 1M units) -
Massive (≥1M units) -

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-0567-2013) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
Medtronic, DBS Extension Kit for Deep Brain Stimulation, Catalog # 7482, 37085, 37086. Rx Only, Manufactured in: Medtronic, Inc. Villalba, Puerto Rico. Medtronic¿ DBSTM Therapy for Parkinson's Disease - Bilateral stimulation of the internal globus pallidus (GPi) or the subthalamic nucleus (STN) using Medtronic DBS Therapy for Parkinson's Disease is indicated for adjunctive therapy in reducing some of the symptoms of advanced, levodopa-responsive Parkinson's disease that are not adequately controlled with medication. STN/GPi target system components: ¿ Kinetra¿ Model 7428 or Soletra¿ Model 7426 Neurostimulator ¿ Activa¿ PC Model 37601, Activa RC Model 37612, Activa SC Model 37602, or Activa SC Model 37603 Neurostimulator ¿ Model 7482, Model 7482A, Model 7483, Model 37085, or Model 37086 Extension ¿ Model 3387 or Model 3389 Lead Medtronic DBS Therapy for Tremor - Unilateral thalamic stimulation of the ventral intermediate nucleus (VIM) using Medtronic DBS Therapy for Tremor is indicated for the suppression of tremor in the upper extremity. The system is intended for patients who are diagnosed with essential tremor or parkinsonian tremor not adequately controlled by medications and where the tremor constitutes a significant functional disability. VIM target system components: ¿ Soletra Model 7426 Neurostimulator ¿ Activa PC Model 37601, Activa RC Model 37612, Activa SC Model 37602, or Activa SC Model 37603 Neurostimulator ¿ Model 7482, Model 7482A, Model 7483, Model 37085, or Model 37086 Extension ¿ Model 3387 or Model 3389 Lead.. Recalled by Medtronic Neuromodulation. Units affected: 140 units.
Why was this product recalled?
Medtronic has identified 140 Restore DBS Extension Kits (model numbers 7482, 37085, and 37086), which may contain a damaged winged connector boot. The connector boot may not have been molded properly and may be cracked. Approximately 14% of the 140 affected extension kits may contain a damaged radiopaque winged connector boot.
Which agency issued this recall?
This recall was issued by the FDA Devices on December 26, 2012. Severity: Moderate. Recall number: Z-0567-2013.
Where was the recalled product distributed?
Distribution: Worldwide Distribution -- USA (nationwide) including the states of CT, MI, NH, NY, OH, PA and VT., and the countries of AUSTRALIA, JAPAN, BELGIUM, FRANCE, GERMANY, ITALY, NETHERLANDS, POLAND, SPAIN, SWITZERLAND, TURKEY and UNITED KINGDOM..
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-0567-2013) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nearby Recalls in This Category

Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →

Browse all Medical Devices recalls →

Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported December 26, 2012. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.