Severity
Moderate
FDA Devices recall · Reported February 16, 2022
Cleaning products, containing, but not limited to, hydrogen peroxide, bleach, alcohol, or antiseptic agents and solvent, such as those containing acetone, toluene, xylene, or cycl…
Baxter Healthcare Corporation recalled Baxter MiniCap Extended Life PD Transfer Set with Twist Clamp, REF R5C4482E - a moderate-severity action.
Baxter MiniCap Extended Life PD Transfer Set with Twist Clamp, REF R5C4482E was recalled by Baxter Healthcare Corporation in February 16, 2022. Reason: Cleaning products, containing, but not limited to, hydrogen peroxide, bleach, alcohol, or antiseptic agents a…. Check the official notice for the remedy. Verify recall #Z-0576-2022 with the FDA Devices before acting.
The recall
Baxter Healthcare Corporation issued this moderate-severity FDA Devices recall-Cleaning products, containing, but not limited to, hydrogen peroxide, bleach, alcohol, or antiseptic agents a….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0576-2022 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0576-2022) was formally reported on February 16, 2022, with the manufacturer initiating the action on January 7, 2022. It is classified under Moderate severity (Class II), with a current status of Ongoing. Baxter Healthcare Corporation is listed as the recalling firm, operating out of Deerfield, IL. Federal records list the affected scope as 62,712 units.
The documented reason for this recall is: Cleaning products, containing, but not limited to, hydrogen peroxide, bleach, alcohol, or antiseptic agents and solvent, such as those containing acetone, toluene, xylene, or cyclohexanone may cause damage, such as leak… Distribution data in the federal record shows the product reached: Worldwide. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
62,712 units
Related Recalls
6
3 from same agency
Baxter MiniCap Extended Life PD Transfer Set with Twist Clamp, REF R5C4482E
Cleaning products, containing, but not limited to, hydrogen peroxide, bleach, alcohol, or antiseptic agents and solvent, such as those containing acetone, toluene, xylene, or cyclohexanone may cause damage, such as leaking or cracking, to the MiniCap Extended Life PD transfer sets.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-0576-2022 |
| Date reported | February 16, 2022 |
| Date initiated | January 7, 2022 |
| Recalling firm | Baxter Healthcare Corporation |
| Firm location | Deerfield, IL |
| Affected scope | 62,712 units |
| Distribution | Worldwide |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported February 16, 2022. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.