Baxter MiniCap Extended Life PD Transfer Set with Twist Clamp, REF R5C4482E

Reported: February 16, 2022 Initiated: January 7, 2022 #Z-0576-2022

Product Description

Baxter MiniCap Extended Life PD Transfer Set with Twist Clamp, REF R5C4482E

Reason for Recall

Cleaning products, containing, but not limited to, hydrogen peroxide, bleach, alcohol, or antiseptic agents and solvent, such as those containing acetone, toluene, xylene, or cyclohexanone may cause damage, such as leaking or cracking, to the MiniCap Extended Life PD transfer sets.

Details

Units Affected
62,712 units
Distribution
Worldwide
Location
Deerfield, IL

Frequently Asked Questions

What product was recalled?
Baxter MiniCap Extended Life PD Transfer Set with Twist Clamp, REF R5C4482E. Recalled by Baxter Healthcare Corporation. Units affected: 62,712 units.
Why was this product recalled?
Cleaning products, containing, but not limited to, hydrogen peroxide, bleach, alcohol, or antiseptic agents and solvent, such as those containing acetone, toluene, xylene, or cyclohexanone may cause damage, such as leaking or cracking, to the MiniCap Extended Life PD transfer sets.
Which agency issued this recall?
This recall was issued by the FDA Devices on February 16, 2022. Severity: Moderate. Recall number: Z-0576-2022.