NC Emerge Monorail PTCA Dilatation Catheter. Sterilized using ethylene oxide. Packaged in the following sizes: 1. REF/Catalog No. H7493926708350, NC EMERGE MR US 3.50MM X 8MM; 2. REF/Catalog No. H7493927608350, NC EMERGE MR OUS 3.50MM X 8MM; 3. REF/Catalog No. H7493927608400, NC EMERGE MR OUS 4.00MM X 8MM; 4. REF/Catalog No. H7493927612200, NC EMERGE MR OUS 2.00MM X 12MM; 5. REF/Catalog No. H7493926712250, NC EMERGE MR US 2.50MM X 12MM; 6. REF/Catalog No. H7493927612250,
Product Description
NC Emerge Monorail PTCA Dilatation Catheter. Sterilized using ethylene oxide. Packaged in the following sizes: 1. REF/Catalog No. H7493926708350, NC EMERGE MR US 3.50MM X 8MM; 2. REF/Catalog No. H7493927608350, NC EMERGE MR OUS 3.50MM X 8MM; 3. REF/Catalog No. H7493927608400, NC EMERGE MR OUS 4.00MM X 8MM; 4. REF/Catalog No. H7493927612200, NC EMERGE MR OUS 2.00MM X 12MM; 5. REF/Catalog No. H7493926712250, NC EMERGE MR US 2.50MM X 12MM; 6. REF/Catalog No. H7493927612250, NC EMERGE MR OUS 2.50MM X 12MM; 7. REF/Catalog No. H7493926712270, NC EMERGE MR US 2.75MM X 12MM; 8. REF/Catalog No. H7493926712320, NC EMERGE MR US 3.25MM X 12MM; 9. REF/Catalog No. H7493927612350, NC EMERGE MR OUS 3.50MM X 12MM; 10. REF/Catalog No. H7493926712400, NC EMERGE MR US 4.00MM X 12MM; 11. REF/Catalog No. H7493927612500, NC EMERGE MR OUS 5.00MM X 12MM; 12. REF/Catalog No. H7493927615270, NC EMERGE MR OUS 2.75MM X 15MM; 13. REF/Catalog No. H7493926715300, NC EMERGE MR US 3.00MM X 15MM; 14. REF/Catalog No. H7493927615300, NC EMERGE MR OUS 3.00MM X 15MM; 15. REF/Catalog No. H7493927615320, NC EMERGE MR OUS 3.25MM X 15MM; 16. REF/Catalog No. H7493926715350, NC EMERGE MR US 3.50MM X 15MM.
Reason for Recall
Certain dilation and extension catheters are being recalled because these catheters may have missed the hydrophilic coating step, which creates a lubricious surface on the devices, during the manufacturing process.
Details
- Recalling Firm
- Boston Scientific Corporation
- Units Affected
- 14,145 devices total
- Distribution
- Nationwide including Washington D.C. and Puerto Rico. OUS to include Canada.
- Location
- Maple Grove, MN
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11