PlainRecalls
ModerateClass IIOngoing

FDA Devices recall · Reported December 3, 2025

Catalyft PL & PL40 Expandable Interbody System labeled as follows with corresponding model numbers: 1. SPACER 6068073 CATALYFT PL SHORT 7MM. Model Number: 6068073. 2. SPACER 6068076 CATALYFT PL LONG 7MM, Model Number: 6068076. 3. SPACER 6068093 CATALYFT PL SHORT 9MM, Model Number: 6068093. 4. SPACER 6068096 CATALYFT PL LONG 9MM, Model Number: 6068096. 5. SPACER 6068113 CATALYFT PL SHORT 11MM, Model Number: 6068113. 6. SPACER 6068116 CATALYFT PL LONG 11MM, Model Number: 6068

Expandable interbody system cage may loose height or collapse (loss of lordosis), which could result in migration, subsidence and/or neurological injury. Therefore, surgical techn…

Recall #
Z-0580-2026
Affected scope
275,260 units
Initiated
September 16, 2025
Verify with FDA Devices →
View my saved recalls

Medtronic Sofamor Danek USA Inc recalled Catalyft PL & PL40 Expandable Interbody System labeled as follows with corresponding mode… — a moderate-severity action.

Catalyft PL & PL40 Expandable Interbody System labeled as follows with corresponding mode… was recalled by Medtronic Sofamor Danek USA Inc in December 3, 2025. Reason: Expandable interbody system cage may loose height or collapse (loss of lordosis), which could result in migra…. Check the official notice for the remedy. Verify recall #Z-0580-2026 with the FDA Devices before acting.

The recall

Medtronic Sofamor Danek USA Inc issued this moderate-severity FDA Devices recall — Expandable interbody system cage may loose height or collapse (loss of lordosis), which could result in migra….

Moderate
severity level
275K units
affected scope
Class II
classification
December 3, 2025
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-0580-2026 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This FDA Devices action (record #Z-0580-2026) was formally reported on December 3, 2025, with the manufacturer initiating the action on September 16, 2025. It is classified under Moderate severity (Class II), with a current status of Ongoing. Medtronic Sofamor Danek USA Inc is listed as the recalling firm, operating out of Memphis, TN. Federal records list the affected scope as 275,260 units, a scale large enough to require multi-state distribution tracking.

The documented reason for this recall is: Expandable interbody system cage may loose height or collapse (loss of lordosis), which could result in migration, subsidence and/or neurological injury. Therefore, surgical technique and instructions for use will be up… Distribution data in the federal record shows the product reached: Worldwide distribution - US Nationwide and Puerto Rico. The countries of Australia, Austria, Belgium, Canada, Croatia, Czech Republic, Denmark, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, N…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices — clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category — 40,409 medical devices recalls on record

-1,00001,0002,0003,0004,0005,000 20052008201120142017202020232026 1,243

Where this recall sits in the database

Severity2366872097High severity (most serious)Moderate severityLow severity
Where this recall sits in the database

Of 100,165 recalls in the database, 23,668 are high severity, 72,097 moderate, and 4,400 low. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger — we do not rank products by risk from raw recall volume.

Severity

Moderate

Affected scope

275,260 units

Related Recalls

6

6 from same agency

Product description

Catalyft PL & PL40 Expandable Interbody System labeled as follows with corresponding model numbers: 1. SPACER 6068073 CATALYFT PL SHORT 7MM. Model Number: 6068073. 2. SPACER 6068076 CATALYFT PL LONG 7MM, Model Number: 6068076. 3. SPACER 6068093 CATALYFT PL SHORT 9MM, Model Number: 6068093. 4. SPACER 6068096 CATALYFT PL LONG 9MM, Model Number: 6068096. 5. SPACER 6068113 CATALYFT PL SHORT 11MM, Model Number: 6068113. 6. SPACER 6068116 CATALYFT PL LONG 11MM, Model Number: 6068116. 7. SPACER 6069073 CATALYFT PL40 SHORT 7MM, Model Number: 6069073. 8. SPACER 6069076 CATALYFT PL40 LONG 7MM, Model Number: 6069076. 9. SPACER 6069093 CATALYFT PL40 SHORT 9MM, Model Number: 6069093. 10. SPACER 6069096 CATALYFT PL40 LONG 9MM, Model Number: 6069096. 11. SPACER 6069113 CATALYFT PL40 SHORT 11MM, Model Number: 6069113. 12. SPACER 6069116 CATALYFT PL40 LONG 11MM, Model Number: 6069116. 13. CATALYFT PL IMPLANT SET, Model Number: SPS03143. 14. CATALYFT PL 40 IMPLANT SET, Model Number: SPS03144. 15. KIT CATALYFT PL IMPLANTS, Model Number: SPS04012. 16. KIT CATALYFT PL 40 IMPLANTS, Model Number: SPS04013. 17. CATALYFT PL 40 IMP KIT, Model Number: SPSJEP40. 18. CATALYFT PL 40 HIGH IMP KIT, Model Number: SPSJEP4H. 19. CATALYFT PL 40 SHORT IMP KIT, Model Number: SPSJEP4S. 20. CATALYFT PL 40 SHORT HIGH IMP KIT, Model Number: SPSJEP4T. 21. CATALYFT PL STRAIGHT HIGH IMP KIT, Model Number: SPSJEPSH. 22. CATALYFT PL STRAIGHT IMP KIT, Model Number: SPSJEPSI. These systems utilize the following affected Instructions for Use (IFU), Document M333023W048E, and Surgical Technique Guide, Document M333023W217.

Reason for recall

Expandable interbody system cage may loose height or collapse (loss of lordosis), which could result in migration, subsidence and/or neurological injury. Therefore, surgical technique and instructions for use will be updated to mitigate risk of loss of lordosis to occur.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Ongoing
Recall number Z-0580-2026
Date reported December 3, 2025
Date initiated September 16, 2025
Recalling firm Medtronic Sofamor Danek USA Inc
Firm location Memphis, TN
Affected scope 275,260 units
Distribution Worldwide distribution - US Nationwide and Puerto Rico. The countries of Australia, Austria, Belgium, Canada, Croatia, Czech Republic, Denmark, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Netherlands, New Cale…

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

275,260 units units affected — multi-state distribution scale.

Regional (<10K units)
Multi-state (10K – 100K units)
Large-scale (100K – 1M units) ✓ This recall
Massive (≥1M units)

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-0580-2026) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting — recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
Catalyft PL & PL40 Expandable Interbody System labeled as follows with corresponding model numbers: 1. SPACER 6068073 CATALYFT PL SHORT 7MM. Model Number: 6068073. 2. SPACER 6068076 CATALYFT PL LONG 7MM, Model Number: 6068076. 3. SPACER 6068093 CATALYFT PL SHORT 9MM, Model Number: 6068093. 4. SPACER 6068096 CATALYFT PL LONG 9MM, Model Number: 6068096. 5. SPACER 6068113 CATALYFT PL SHORT 11MM, Model Number: 6068113. 6. SPACER 6068116 CATALYFT PL LONG 11MM, Model Number: 6068116. 7. SPACER 6069073 CATALYFT PL40 SHORT 7MM, Model Number: 6069073. 8. SPACER 6069076 CATALYFT PL40 LONG 7MM, Model Number: 6069076. 9. SPACER 6069093 CATALYFT PL40 SHORT 9MM, Model Number: 6069093. 10. SPACER 6069096 CATALYFT PL40 LONG 9MM, Model Number: 6069096. 11. SPACER 6069113 CATALYFT PL40 SHORT 11MM, Model Number: 6069113. 12. SPACER 6069116 CATALYFT PL40 LONG 11MM, Model Number: 6069116. 13. CATALYFT PL IMPLANT SET, Model Number: SPS03143. 14. CATALYFT PL 40 IMPLANT SET, Model Number: SPS03144. 15. KIT CATALYFT PL IMPLANTS, Model Number: SPS04012. 16. KIT CATALYFT PL 40 IMPLANTS, Model Number: SPS04013. 17. CATALYFT PL 40 IMP KIT, Model Number: SPSJEP40. 18. CATALYFT PL 40 HIGH IMP KIT, Model Number: SPSJEP4H. 19. CATALYFT PL 40 SHORT IMP KIT, Model Number: SPSJEP4S. 20. CATALYFT PL 40 SHORT HIGH IMP KIT, Model Number: SPSJEP4T. 21. CATALYFT PL STRAIGHT HIGH IMP KIT, Model Number: SPSJEPSH. 22. CATALYFT PL STRAIGHT IMP KIT, Model Number: SPSJEPSI. These systems utilize the following affected Instructions for Use (IFU), Document M333023W048E, and Surgical Technique Guide, Document M333023W217.. Recalled by Medtronic Sofamor Danek USA Inc. Units affected: 275,260 units.
Why was this product recalled?
Expandable interbody system cage may loose height or collapse (loss of lordosis), which could result in migration, subsidence and/or neurological injury. Therefore, surgical technique and instructions for use will be updated to mitigate risk of loss of lordosis to occur.
Which agency issued this recall?
This recall was issued by the FDA Devices on December 3, 2025. Severity: Moderate. Recall number: Z-0580-2026.
Where was the recalled product distributed?
Distribution: Worldwide distribution - US Nationwide and Puerto Rico. The countries of Australia, Austria, Belgium, Canada, Croatia, Czech Republic, Denmark, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Netherlands, New Caledonia, New Zealand, Norway, Portugal, Reunion, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom..
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-0580-2026) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nearby Recalls in This Category

Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with GE HealthCare TRADE IN CARESCAPE TELEMETRY SERVER ARK-2250L… →

Data Sources

Source: FDA, CPSC, and NHTSA federal recall databases. This recall: FDA Devices, reported December 3, 2025.

  • Source: FDA — Food and Drug Administration, openFDA Enforcement API (food, drug, and medical device recalls)
  • Source: CPSC — Consumer Product Safety Commission Recalls API (consumer product recalls and hazards)
  • Source: NHTSA — National Highway Traffic Safety Administration Recalls API (vehicle safety recalls)

Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.

Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records — no number is typed in by an editor. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of June 2026.