Advia Chemistry Calibrator For in vitro diagnostic use in the calibration of ADVIA¿ Chemistry systems for the following methods: Albumin (ALB, ALB_c), Direct Bilirubin (DBIL_2), Total Bilirubin (TBIL_2), Calcium (CA, CA_c, CA_2, CA_2c), Cholesterol (CHOL), Creatinine (CREA, CREA_2, CRE_2c, ECRE_2), Glucose Hexakinase (GLUH), Glucose Oxidase (GLUO), Iron (IRON_2), Lithium (LITH), Magnesium (MG, MG_c), Phosphorus - inorganic (IP, IP_c), Total Protein (TP, TP_c), Triglycerides (TRIG), Urea Nitrog
Product Description
Advia Chemistry Calibrator For in vitro diagnostic use in the calibration of ADVIA¿ Chemistry systems for the following methods: Albumin (ALB, ALB_c), Direct Bilirubin (DBIL_2), Total Bilirubin (TBIL_2), Calcium (CA, CA_c, CA_2, CA_2c), Cholesterol (CHOL), Creatinine (CREA, CREA_2, CRE_2c, ECRE_2), Glucose Hexakinase (GLUH), Glucose Oxidase (GLUO), Iron (IRON_2), Lithium (LITH), Magnesium (MG, MG_c), Phosphorus - inorganic (IP, IP_c), Total Protein (TP, TP_c), Triglycerides (TRIG), Urea Nitrogen (UN, UN_c), Uric Acid (UA, UA_c)
Reason for Recall
Siemens has confirmed that the DBIL_2 Assigned Calibrator Values (ACV) currently assigned to the affected lots may cause, on average, a positive proportional bias of approximately 41% on patient samples compared to the reference method for Direct Bilirubin measurement.
Details
- Recalling Firm
- Siemens Healthcare Diagnostics, Inc.
- Units Affected
- 10,614 units
- Distribution
- Worldwide Distribution - US (Nationwide) Internationally to Canada, Mexico, Algeria, Australia, Austria, Bangladesh, Belgium, Brazil, Chile, Cyprus, Denmark, Egypt, Ecuador, Finland, France, Germany, Great Britain, Hong Kong, Hungary, Italy, Japan, Lithuania, Malaysia, Mexico, Netherlands, New Zealand, Norway, Pakistan, Paraguay, Peru, Poland, Portugal, Russian Fed. Slovakia, Slovenia, Spain, Sweden, Switzerland, Thailand, Turkey, Vatican City and Vietnam
- Location
- Tarrytown, NY
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11