FDA Devices recall · Reported January 8, 2014
Integra Camino Advanced Monitor with Waveform Display Model CAM01 Integra Neurosciences, 311 Enterprise Dr Plainsboro, NJ 08536 USA The Camino Advanced Monitor is a compact, portable device for use with Camino 110-4 series of Pressure / Temperature and Pressure Transducer-Tipped Catheters. Pressure and/or Temperature are measured at the Catheter tip, eliminating the need for external transducers, fluid, pressure tubing, and flush devices. The Camino displays intracranial pressure (ICP), intra
Integra became aware through a complaint from a single non-US customer that if the optional SYS/DIAS mode is used on CAM01 monitor that contains Mod 4 Firmware, the displayed valu…
Integra LifeSciences Corp. recalled Integra Camino Advanced Monitor with Waveform Display Model CAM01 Integra Neurosciences, … - a moderate-severity action.
Integra Camino Advanced Monitor with Waveform Display Model CAM01 Integra Neurosciences, … was recalled by Integra LifeSciences Corp. in January 8, 2014. Reason: Integra became aware through a complaint from a single non-US customer that if the optional SYS/DIAS mode is …. Check the official notice for the remedy. Verify recall #Z-0602-2014 with the FDA Devices before acting.
The recall
Integra LifeSciences Corp. issued this moderate-severity FDA Devices recall-Integra became aware through a complaint from a single non-US customer that if the optional SYS/DIAS mode is ….
- Moderate
- severity level
- Class II
- classification
- January 8, 2014
- reported
Sourced from official FDA Devices enforcement records. Verify recall #Z-0602-2014 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
Recall Insight
This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Medical Devices recalls over time
Where this recall sits in its category - 41,021 medical devices recalls on record
Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Product description
Integra Camino Advanced Monitor with Waveform Display Model CAM01 Integra Neurosciences, 311 Enterprise Dr Plainsboro, NJ 08536 USA The Camino Advanced Monitor is a compact, portable device for use with Camino 110-4 series of Pressure / Temperature and Pressure Transducer-Tipped Catheters. Pressure and/or Temperature are measured at the Catheter tip, eliminating the need for external transducers, fluid, pressure tubing, and flush devices. The Camino displays intracranial pressure (ICP), intracranial temperature (ICT), and calculates cerebral perfusion pressure (CPP) (CPP = mean arterial pressure - mean intracranial pressure). The Camino provides a continuous display of the ICP waveform, as well as mean ICP, CPP, ICT, or systolic and diastolic waveform values. A continuous record of mean pressure and CPP values over the most recent 24-hour period is stored in memory, and can be displayed on command as a TREND either as the most recent 12 or 24 hour period. Outputs for ICP and ICT are available for use with patient bedside monitors. An isolated analogue output provides a continuous ICP waveform for hard copy documentation or data acquisition. An input to receive mean arterial pressure from a compatible patient bedside monitor is available for use when it is desired to have a CPP reading. The mean arterial pressure cannot be acquired from a non-invasive blood pressure cuff. Although the Camino is intended to be a stand-alone system, it also conveniently connects to hospital bedside monitoring systems.
Reason for recall
Integra became aware through a complaint from a single non-US customer that if the optional SYS/DIAS mode is used on CAM01 monitor that contains Mod 4 Firmware, the displayed values for the systolic and diastolic intracranial pressure (ICP) are reversed. But, if the CAM01 is used with an auxiliary bedside monitor, the bedside monitor displays these values correctly.
Recall Profile
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0602-2014 |
| Date reported | January 8, 2014 |
| Date initiated | November 26, 2013 |
| Recalling firm | Integra LifeSciences Corp. |
| Firm location | Plainsboro, NJ |
| Affected scope | 275 CAM01 Monitors |
| Distribution | Worldwide Distribution - USA ( Nationwide ) and the countries of : UK, Saudi Arabia, South Africa, Brunei, China, France, Poland, Turkey, Greece, Sweden, United Arab Emirates, Russia, Spain, Pakistan, Finland, Israel, Italy, Slovenia, Japa… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
- Check the recall number (Z-0602-2014) and product description against the item you own. Search the archive
- Confirm the current status with FDA Devices before acting, recall details can be updated.
- Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Where was the recalled product distributed? ▼
Should I stop using a recalled medication or medical device? ▼
Nationwide recalls with similar severity or category timing
Drug and device notices outside this firm and state, matched by severity timing and product category.
Similar severity · nearest report date
Same archive severity grade (Moderate), closest by report date among other firms and states.
Similar Medical Devices · nearest report date
Same product category, closest by report date among other firms and states (deduped against the severity set).
Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls
Methodology · Corrections · Archive as of August 2026