PlainRecalls
ModerateClass IITerminated

FDA Devices recall · Reported January 8, 2014

Integra Camino Advanced Monitor with Waveform Display Model CAM01 Integra Neurosciences, 311 Enterprise Dr Plainsboro, NJ 08536 USA The Camino Advanced Monitor is a compact, portable device for use with Camino 110-4 series of Pressure / Temperature and Pressure Transducer-Tipped Catheters. Pressure and/or Temperature are measured at the Catheter tip, eliminating the need for external transducers, fluid, pressure tubing, and flush devices. The Camino displays intracranial pressure (ICP), intra

Integra became aware through a complaint from a single non-US customer that if the optional SYS/DIAS mode is used on CAM01 monitor that contains Mod 4 Firmware, the displayed valu…

Recall #
Z-0602-2014
Affected scope
275 CAM01 Monitors
Initiated
November 26, 2013
Compiled from official public sources by the editorial team.
Verify with FDA Devices →
View my saved recalls

Integra LifeSciences Corp. recalled Integra Camino Advanced Monitor with Waveform Display Model CAM01 Integra Neurosciences, … - a moderate-severity action.

Integra Camino Advanced Monitor with Waveform Display Model CAM01 Integra Neurosciences, … was recalled by Integra LifeSciences Corp. in January 8, 2014. Reason: Integra became aware through a complaint from a single non-US customer that if the optional SYS/DIAS mode is …. Check the official notice for the remedy. Verify recall #Z-0602-2014 with the FDA Devices before acting.

The recall

Integra LifeSciences Corp. issued this moderate-severity FDA Devices recall-Integra became aware through a complaint from a single non-US customer that if the optional SYS/DIAS mode is ….

Moderate
severity level
Class II
classification
January 8, 2014
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-0602-2014 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This FDA Devices action (record #Z-0602-2014) was formally reported on January 8, 2014, with the manufacturer initiating the action on November 26, 2013. It is classified under Moderate severity (Class II), with a current status of Terminated. Integra LifeSciences Corp. is listed as the recalling firm, operating out of Plainsboro, NJ. Federal records list the affected scope as 275 CAM01 Monitors.

The documented reason for this recall is: Integra became aware through a complaint from a single non-US customer that if the optional SYS/DIAS mode is used on CAM01 monitor that contains Mod 4 Firmware, the displayed values for the systolic and diastolic intrac… Distribution data in the federal record shows the product reached: Worldwide Distribution - USA ( Nationwide ) and the countries of : UK, Saudi Arabia, South Africa, Brunei, China, France, Poland, Turkey, Greece, Sweden, United Arab Emirates, Russia, Spain, Pakistan, Finland, Israel, I…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 40,322 medical devices recalls on record

-1,00001,0002,0003,0004,000 20052008201120142017202020232026 1,248
Severity2487872426High severity (most serious)Moderate severityLow severity
Where this recall sits in the database

Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Severity

Moderate

Affected scope

275 CAM01 Monitors

Related Recalls

6

3 from same agency

Product description

Integra Camino Advanced Monitor with Waveform Display Model CAM01 Integra Neurosciences, 311 Enterprise Dr Plainsboro, NJ 08536 USA The Camino Advanced Monitor is a compact, portable device for use with Camino 110-4 series of Pressure / Temperature and Pressure Transducer-Tipped Catheters. Pressure and/or Temperature are measured at the Catheter tip, eliminating the need for external transducers, fluid, pressure tubing, and flush devices. The Camino displays intracranial pressure (ICP), intracranial temperature (ICT), and calculates cerebral perfusion pressure (CPP) (CPP = mean arterial pressure - mean intracranial pressure). The Camino provides a continuous display of the ICP waveform, as well as mean ICP, CPP, ICT, or systolic and diastolic waveform values. A continuous record of mean pressure and CPP values over the most recent 24-hour period is stored in memory, and can be displayed on command as a TREND either as the most recent 12 or 24 hour period. Outputs for ICP and ICT are available for use with patient bedside monitors. An isolated analogue output provides a continuous ICP waveform for hard copy documentation or data acquisition. An input to receive mean arterial pressure from a compatible patient bedside monitor is available for use when it is desired to have a CPP reading. The mean arterial pressure cannot be acquired from a non-invasive blood pressure cuff. Although the Camino is intended to be a stand-alone system, it also conveniently connects to hospital bedside monitoring systems.

Reason for recall

Integra became aware through a complaint from a single non-US customer that if the optional SYS/DIAS mode is used on CAM01 monitor that contains Mod 4 Firmware, the displayed values for the systolic and diastolic intracranial pressure (ICP) are reversed. But, if the CAM01 is used with an auxiliary bedside monitor, the bedside monitor displays these values correctly.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Terminated
Recall number Z-0602-2014
Date reported January 8, 2014
Date initiated November 26, 2013
Recalling firm Integra LifeSciences Corp.
Firm location Plainsboro, NJ
Affected scope 275 CAM01 Monitors
Distribution Worldwide Distribution - USA ( Nationwide ) and the countries of : UK, Saudi Arabia, South Africa, Brunei, China, France, Poland, Turkey, Greece, Sweden, United Arab Emirates, Russia, Spain, Pakistan, Finland, Israel, Italy, Slovenia, Japa…

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-0602-2014) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
Integra Camino Advanced Monitor with Waveform Display Model CAM01 Integra Neurosciences, 311 Enterprise Dr Plainsboro, NJ 08536 USA The Camino Advanced Monitor is a compact, portable device for use with Camino 110-4 series of Pressure / Temperature and Pressure Transducer-Tipped Catheters. Pressure and/or Temperature are measured at the Catheter tip, eliminating the need for external transducers, fluid, pressure tubing, and flush devices. The Camino displays intracranial pressure (ICP), intracranial temperature (ICT), and calculates cerebral perfusion pressure (CPP) (CPP = mean arterial pressure - mean intracranial pressure). The Camino provides a continuous display of the ICP waveform, as well as mean ICP, CPP, ICT, or systolic and diastolic waveform values. A continuous record of mean pressure and CPP values over the most recent 24-hour period is stored in memory, and can be displayed on command as a TREND either as the most recent 12 or 24 hour period. Outputs for ICP and ICT are available for use with patient bedside monitors. An isolated analogue output provides a continuous ICP waveform for hard copy documentation or data acquisition. An input to receive mean arterial pressure from a compatible patient bedside monitor is available for use when it is desired to have a CPP reading. The mean arterial pressure cannot be acquired from a non-invasive blood pressure cuff. Although the Camino is intended to be a stand-alone system, it also conveniently connects to hospital bedside monitoring systems.. Recalled by Integra LifeSciences Corp.. Units affected: 275 CAM01 Monitors.
Why was this product recalled?
Integra became aware through a complaint from a single non-US customer that if the optional SYS/DIAS mode is used on CAM01 monitor that contains Mod 4 Firmware, the displayed values for the systolic and diastolic intracranial pressure (ICP) are reversed. But, if the CAM01 is used with an auxiliary bedside monitor, the bedside monitor displays these values correctly.
Which agency issued this recall?
This recall was issued by the FDA Devices on January 8, 2014. Severity: Moderate. Recall number: Z-0602-2014.
Where was the recalled product distributed?
Distribution: Worldwide Distribution - USA ( Nationwide ) and the countries of : UK, Saudi Arabia, South Africa, Brunei, China, France, Poland, Turkey, Greece, Sweden, United Arab Emirates, Russia, Spain, Pakistan, Finland, Israel, Italy, Slovenia, Japan, Poland, Malaysia, Germany and Panama..
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-0602-2014) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nearby Recalls in This Category

Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →

Browse all Medical Devices recalls →

Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported January 8, 2014. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.