Severity
Moderate
FDA Devices recall · Reported December 27, 2023
Firm received complaints of breaks in the insulation, shielding, or wires of the venous probe connection cable. Fracture of the venous probe cable may influence the management of …
Maquet Medical Systems USA recalled Venous probe connection cable (REF 701069333 and 701048804) used with the Cardiohelp-i Sy… - a moderate-severity action.
Venous probe connection cable (REF 701069333 and 701048804) used with the Cardiohelp-i Sy… was recalled by Maquet Medical Systems USA in December 27, 2023. Reason: Firm received complaints of breaks in the insulation, shielding, or wires of the venous probe connection cabl…. Check the official notice for the remedy. Verify recall #Z-0605-2024 with the FDA Devices before acting.
The recall
Maquet Medical Systems USA issued this moderate-severity FDA Devices recall-Firm received complaints of breaks in the insulation, shielding, or wires of the venous probe connection cabl….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0605-2024 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0605-2024) was formally reported on December 27, 2023, with the manufacturer initiating the action on November 13, 2023. It is classified under Moderate severity (Class II), with a current status of Ongoing. Maquet Medical Systems USA is listed as the recalling firm, operating out of Wayne, NJ. Federal records list the affected scope as 1213 US, 2589 OUS.
The documented reason for this recall is: Firm received complaints of breaks in the insulation, shielding, or wires of the venous probe connection cable. Fracture of the venous probe cable may influence the management of extracorporeal support. Patient may be e… Distribution data in the federal record shows the product reached: Domestic distribution Nationwide. Foreign distribution to Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Canada, Cayman Islands, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Estonia, Finland,…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
1213 US, 2589 OUS
Related Recalls
6
3 from same agency
Venous probe connection cable (REF 701069333 and 701048804) used with the Cardiohelp-i System, part numbers 701072780 and 701041802
Firm received complaints of breaks in the insulation, shielding, or wires of the venous probe connection cable. Fracture of the venous probe cable may influence the management of extracorporeal support. Patient may be exposed to inappropriate blood flow, blood temperature, or hematocrit/hemoglobin levels as a result of the CARDIOHELP-i not receiving information from the venous probe.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-0605-2024 |
| Date reported | December 27, 2023 |
| Date initiated | November 13, 2023 |
| Recalling firm | Maquet Medical Systems USA |
| Firm location | Wayne, NJ |
| Affected scope | 1213 US, 2589 OUS |
| Distribution | Domestic distribution Nationwide. Foreign distribution to Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Canada, Cayman Islands, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Estonia, Finland, France, French Poly… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported December 27, 2023. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.