Severity
Moderate
FDA Devices recall · Reported December 19, 2018
When your system was installed, the Main disconnect switch, EPO button on the A1 panel, and Remote EPO button(s) (outside of the A1 panel) may not have been wired correctly which …
Ge Healthcare, LLC recalled Discovery NM/CT 670 ES, model 5376204-70-57 - a moderate-severity action.
Discovery NM/CT 670 ES, model 5376204-70-57 was recalled by Ge Healthcare, LLC in December 19, 2018. Reason: When your system was installed, the Main disconnect switch, EPO button on the A1 panel, and Remote EPO button…. Check the official notice for the remedy. Verify recall #Z-0607-2019 with the FDA Devices before acting.
The recall
Ge Healthcare, LLC issued this moderate-severity FDA Devices recall-When your system was installed, the Main disconnect switch, EPO button on the A1 panel, and Remote EPO button….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0607-2019 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0607-2019) was formally reported on December 19, 2018, with the manufacturer initiating the action on May 24, 2018. It is classified under Moderate severity (Class II), with a current status of Completed. Ge Healthcare, LLC is listed as the recalling firm, operating out of Waukesha, WI. Federal records list the affected scope as 2 units.
The documented reason for this recall is: When your system was installed, the Main disconnect switch, EPO button on the A1 panel, and Remote EPO button(s) (outside of the A1 panel) may not have been wired correctly which could result in these not shutting off a… Distribution data in the federal record shows the product reached: CA, DC, OH, UT, WI, Canada. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,422 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
2 units
Related Recalls
6
0 from same agency
Discovery NM/CT 670 ES, model 5376204-70-57
When your system was installed, the Main disconnect switch, EPO button on the A1 panel, and Remote EPO button(s) (outside of the A1 panel) may not have been wired correctly which could result in these not shutting off all power to the entire system as intended, resulting in a potential electric hazard.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Completed |
| Recall number | Z-0607-2019 |
| Date reported | December 19, 2018 |
| Date initiated | May 24, 2018 |
| Recalling firm | Ge Healthcare, LLC |
| Firm location | Waukesha, WI |
| Affected scope | 2 units |
| Distribution | CA, DC, OH, UT, WI, Canada |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-07-02
· 2026-07-02
· 2026-06-25
· 2026-06-25
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported December 19, 2018. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.