FDA Devices recall · Reported December 11, 2024
BD identified through potency testing as part of a stability test request to monitor Ampicillin AM-2 due to decrease in potency results of 65% at 18 months, may result in falsely …
Becton Dickinson & Co. recalled BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) - In-Vitro BD BBL Sensi Disc Ampicillin 2 ¿g (AM… - a moderate-severity action.
BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) - In-Vitro BD BBL Sensi Disc Ampicillin 2 ¿g (AM… was recalled by Becton Dickinson & Co. in December 11, 2024. Reason: BD identified through potency testing as part of a stability test request to monitor Ampicillin AM-2 due to d…. Check the official notice for the remedy. Verify recall #Z-0621-2025 with the FDA Devices before acting.
The recall
Becton Dickinson & Co. issued this moderate-severity FDA Devices recall-BD identified through potency testing as part of a stability test request to monitor Ampicillin AM-2 due to d….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0621-2025 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0621-2025) was formally reported on December 11, 2024, with the manufacturer initiating the action on November 6, 2024. It is classified under Moderate severity (Class II), with a current status of Ongoing. Becton Dickinson & Co. is listed as the recalling firm, operating out of Sparks, MD. Federal records list the affected scope as 13,190 units.
The documented reason for this recall is: BD identified through potency testing as part of a stability test request to monitor Ampicillin AM-2 due to decrease in potency results of 65% at 18 months, may result in falsely resistant result for ampicillin suscepti… Distribution data in the federal record shows the product reached: Worldwide - US Nationwide distribution in the states of CA, GA, IL, KY, MA, MD, NE, PA and the countries of Canada, Belgium, Singapore, Malaysia.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
13,190 units
Related Recalls
6
3 from same agency
BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) - In-Vitro BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) are used for semi-quantitative in vitro susceptibility testing by the agar disc diffusion test procedure of common, rapidly growing and certain fastidious bacterial pathogens. Catalog Number: 231263
BD identified through potency testing as part of a stability test request to monitor Ampicillin AM-2 due to decrease in potency results of 65% at 18 months, may result in falsely resistant result for ampicillin susceptibility
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-0621-2025 |
| Date reported | December 11, 2024 |
| Date initiated | November 6, 2024 |
| Recalling firm | Becton Dickinson & Co. |
| Firm location | Sparks, MD |
| Affected scope | 13,190 units |
| Distribution | Worldwide - US Nationwide distribution in the states of CA, GA, IL, KY, MA, MD, NE, PA and the countries of Canada, Belgium, Singapore, Malaysia. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported December 11, 2024. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.