Infusomat utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog numbers: 1. UNIV.15DROP PUMP SET W/3 SAFELINE LL; 15 drops/mL, priming volume: 22mL, length: 119in, SafeLine; Catalog Number: 362032. 2. UNIV. 15 DR PUMP SET, 1.2 FIL, 2 CRSITE; UNIV. 15 DR PUMP SET, 1.2 FIL, 2 CRSITE; Catalog Number: 362033. 3. 60DROP METRISET PU

Reported: December 3, 2025 Initiated: October 29, 2025 #Z-0640-2026 49,330,370 units

B Braun Medical Inc issued this FDA Devices recall on December 3, 2025. Classified as Moderate severity (Class II). Approximately 49,330,370 units are affected. The recall was issued because: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inabil…. This recall notice is sourced from official FDA Devices enforcement records. Below you will find the complete product description, hazard information, remedy instructions, and related recalls from the same manufacturer or product category.

Recall Insight

This FDA Devices action (record #Z-0640-2026) was formally reported on December 3, 2025, with the manufacturer initiating the action on October 29, 2025. It is classified under Moderate severity (Class II), with a current status of Ongoing. B Braun Medical Inc is listed as the recalling firm, operating out of Bethlehem, PA. Federal records indicate 49,330,370 units are affected, placing this recall in the million-unit bracket that typically triggers nationwide consumer alerts and retailer sweeps.

The documented reason for this recall is: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). Distribution data in the federal record shows the product reached: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

To put this record in context, PlainRecalls indexes 83,949 recalls across the FDA, CPSC, NHTSA and USDA FSIS going back to 1995. Within the same product category, the database holds 6 closely related recalls, of which 6 were also issued by FDA Devices. That clustering is a signal — repeated actions in a narrow category often indicate a systemic quality-control issue, a supplier-wide contamination, or a design defect that has propagated across product lines. This recall is roughly 1 year old; older recalls can remain relevant because many units enter resale, rental, and secondary-market channels where the original warning never reaches the end user. Always cross-check the recall number against the official agency page before relying on any summary.

Recall Distribution by Severity Class

Severity1Class I (Critical)Class II (Moderate)Class III (Low)
Recall Distribution by Severity Class

Severity

Moderate

Units Affected

49,330,370

Related Recalls

6

6 from same agency

Product Description

Infusomat utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog numbers: 1. UNIV.15DROP PUMP SET W/3 SAFELINE LL; 15 drops/mL, priming volume: 22mL, length: 119in, SafeLine; Catalog Number: 362032. 2. UNIV. 15 DR PUMP SET, 1.2 FIL, 2 CRSITE; UNIV. 15 DR PUMP SET, 1.2 FIL, 2 CRSITE; Catalog Number: 362033. 3. 60DROP METRISET PUMP SET, 3 SAFELINE; Priming volume: 19mL, length: 120in, SafeLine; Catalog Number: 362034. 4. Y TYPE BLOOD PUMP SET, SAFELINE INJ; Priming volume: 39mL, length: 120in, SafeLine¿; Catalog Number: 362043. 5. UNIV. 15 DROP PUMP SET W/ 0.2 FILTER; 15 drops/mL, priming volume: 26mL, length: 124in, SafeLine; Catalog Number: 362050. 6. SPACE PUMP SET 15 DROP W/2 SAFEDAY; SPACE PUMP SET 15 DROP W/2 SAFEDAY; Catalog Number: 362420. 7. UNIV. 15 DROP PUMP SET W/ 3 ULTRASITE LL; 15 drops/mL, priming volume: 21mL, length: 120in, ULTRASITE; Catalog Number: 362431. 8. SPACE PUMP IV SET, 120 IN.; 15 drops/mL, priming volume: 21mL, length: 118in, ULTRASITE; Catalog Number: 362432. 9. 60 DROP METRISET PUMP SET,3 SFLINE ASV; Priming volume: 21mL, length: 121in, SafeLine; Catalog Number: 363010. 10. Y-TYPE BLOOD PUMP SET, SAFELINE, ASV; Priming volume: 40mL, length: 123in, SafeLine; Catalog Number: 363019. 11. UNIV. 15 DROP PUMP SET W/3 SAFELINE, ASV; 15 drops/mL, priming volume: 22mL, length: 119in, SafeLine; Catalog Number: 363030. 12. UNIV. 15 DROP PUMP SET, 3 ULTRASITE, ASV; 15 drops/mL, priming volume: 22mL, length: 123in, ULTRASITE; Catalog Number: 363230. 13. Y-TYPE BLOOD PUMP SET, CARESITE, ASV; Priming volume: 41mL, length: 120in, CARESITE; Catalog Number: 363419. 14. SPACE PUMP IV SET, 2 CARESITES, ASV; 15 drops/mL, priming volume: 24mL, length: 120in, CARESITE; Catalog Number: 363420. 15. UNIV. 15 DROP PUMP SET, 2 CARESITE, ASV; 15 drops/mL, priming volume: 22mL, length: 120in, CARESITE; Catalog Number: 363421. 16. 60DROP METRISET PUMP, 3 CARESITES, ASV; Priming volume: 21mL, length: 130in, CARESITE; Catalog Number: 363422. 17. UNIV. 60 DROP PUMP SET, 2 CARESITES, ASV; 60 drops/mL, priming volume: 21mL, length: 118in, CARESITE; Catalog Number: 363423. 18. UNIV. 15 DR PUMP SET, 1.2FIL, 2CRSTE ASV; UNIV. 15 DR PUMP SET, 1.2FIL, 2CRSTE ASV; Catalog Number: 363424. 19. UNIV. 15 DROP PUMP SET, 3 CARESITE, ASV; 15 drops/mL, priming volume: 23mL, length: 123in, CARESITE; Catalog Number: 363430. 20. UNIV. 15 DROP PUMP SET, 3 CARESITE, ASV; UNIV. 15 DROP PUMP SET, 3 CARESITE, ASV; Catalog Number: 363433. 21. SPACE PUMP IV SET 15D, 110 IN.; "SPACE PUMP IV SET 15D, 110 IN. CLOSE"; Catalog Number: 480254. 22. SPACE PUMP IV SET, 127 IN.; SPACE PUMP IV SET, 127 IN.; Catalog Number: 480255. 23. SPACE PUMP SET 15D, CARESITE LAD 122 IN. ; 15 drops/mL, priming volume: 23mL, length: 122in, CARESITE; Catalog Number: 480256. 24. 15 DROP PUMP SET 1 SAFEDAY 0.2 FILTER; 15 DROP PUMP SET 1 SAFEDAY 0.2 FILTER; Catalog Number: 480258. 25. SPACE PUMP IV SET 15D, 90 IN.; SPACE PUMP IV SET 15D, 90 IN.; Catalog Number: 480263. 26. UNIV. 15 DROP PUMP SET W 3 CARESITE L.L.; 15 drops/mL, priming volume: 22mL, length: 120in, CARESITE; Catalog Number: 490100. 27. 60 DROP METRISET PUMP SET, 3 CARESITES; Priming volume: 19mL, length: 120in, CARESITE; Catalog Number: 490101.

Reason for Recall

Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

Details

Recalling Firm
B Braun Medical Inc
Units Affected
49,330,370
Distribution
Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.
Location
Bethlehem, PA

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Ongoing
Recall number Z-0640-2026
Date reported December 3, 2025
Date initiated October 29, 2025
Recalling firm B Braun Medical Inc
Units affected 49,330,370
Distribution Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

49,330,370 units affected — million-unit bracket.

Regional (<10K units)
Multi-state (10K – 100K units)
Large-scale (100K – 1M units)
Massive (≥1M units) ✓ This recall

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

Frequently Asked Questions

What product was recalled?
Infusomat utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog numbers: 1. UNIV.15DROP PUMP SET W/3 SAFELINE LL; 15 drops/mL, priming volume: 22mL, length: 119in, SafeLine; Catalog Number: 362032. 2. UNIV. 15 DR PUMP SET, 1.2 FIL, 2 CRSITE; UNIV. 15 DR PUMP SET, 1.2 FIL, 2 CRSITE; Catalog Number: 362033. 3. 60DROP METRISET PUMP SET, 3 SAFELINE; Priming volume: 19mL, length: 120in, SafeLine; Catalog Number: 362034. 4. Y TYPE BLOOD PUMP SET, SAFELINE INJ; Priming volume: 39mL, length: 120in, SafeLine¿; Catalog Number: 362043. 5. UNIV. 15 DROP PUMP SET W/ 0.2 FILTER; 15 drops/mL, priming volume: 26mL, length: 124in, SafeLine; Catalog Number: 362050. 6. SPACE PUMP SET 15 DROP W/2 SAFEDAY; SPACE PUMP SET 15 DROP W/2 SAFEDAY; Catalog Number: 362420. 7. UNIV. 15 DROP PUMP SET W/ 3 ULTRASITE LL; 15 drops/mL, priming volume: 21mL, length: 120in, ULTRASITE; Catalog Number: 362431. 8. SPACE PUMP IV SET, 120 IN.; 15 drops/mL, priming volume: 21mL, length: 118in, ULTRASITE; Catalog Number: 362432. 9. 60 DROP METRISET PUMP SET,3 SFLINE ASV; Priming volume: 21mL, length: 121in, SafeLine; Catalog Number: 363010. 10. Y-TYPE BLOOD PUMP SET, SAFELINE, ASV; Priming volume: 40mL, length: 123in, SafeLine; Catalog Number: 363019. 11. UNIV. 15 DROP PUMP SET W/3 SAFELINE, ASV; 15 drops/mL, priming volume: 22mL, length: 119in, SafeLine; Catalog Number: 363030. 12. UNIV. 15 DROP PUMP SET, 3 ULTRASITE, ASV; 15 drops/mL, priming volume: 22mL, length: 123in, ULTRASITE; Catalog Number: 363230. 13. Y-TYPE BLOOD PUMP SET, CARESITE, ASV; Priming volume: 41mL, length: 120in, CARESITE; Catalog Number: 363419. 14. SPACE PUMP IV SET, 2 CARESITES, ASV; 15 drops/mL, priming volume: 24mL, length: 120in, CARESITE; Catalog Number: 363420. 15. UNIV. 15 DROP PUMP SET, 2 CARESITE, ASV; 15 drops/mL, priming volume: 22mL, length: 120in, CARESITE; Catalog Number: 363421. 16. 60DROP METRISET PUMP, 3 CARESITES, ASV; Priming volume: 21mL, length: 130in, CARESITE; Catalog Number: 363422. 17. UNIV. 60 DROP PUMP SET, 2 CARESITES, ASV; 60 drops/mL, priming volume: 21mL, length: 118in, CARESITE; Catalog Number: 363423. 18. UNIV. 15 DR PUMP SET, 1.2FIL, 2CRSTE ASV; UNIV. 15 DR PUMP SET, 1.2FIL, 2CRSTE ASV; Catalog Number: 363424. 19. UNIV. 15 DROP PUMP SET, 3 CARESITE, ASV; 15 drops/mL, priming volume: 23mL, length: 123in, CARESITE; Catalog Number: 363430. 20. UNIV. 15 DROP PUMP SET, 3 CARESITE, ASV; UNIV. 15 DROP PUMP SET, 3 CARESITE, ASV; Catalog Number: 363433. 21. SPACE PUMP IV SET 15D, 110 IN.; "SPACE PUMP IV SET 15D, 110 IN. CLOSE"; Catalog Number: 480254. 22. SPACE PUMP IV SET, 127 IN.; SPACE PUMP IV SET, 127 IN.; Catalog Number: 480255. 23. SPACE PUMP SET 15D, CARESITE LAD 122 IN. ; 15 drops/mL, priming volume: 23mL, length: 122in, CARESITE; Catalog Number: 480256. 24. 15 DROP PUMP SET 1 SAFEDAY 0.2 FILTER; 15 DROP PUMP SET 1 SAFEDAY 0.2 FILTER; Catalog Number: 480258. 25. SPACE PUMP IV SET 15D, 90 IN.; SPACE PUMP IV SET 15D, 90 IN.; Catalog Number: 480263. 26. UNIV. 15 DROP PUMP SET W 3 CARESITE L.L.; 15 drops/mL, priming volume: 22mL, length: 120in, CARESITE; Catalog Number: 490100. 27. 60 DROP METRISET PUMP SET, 3 CARESITES; Priming volume: 19mL, length: 120in, CARESITE; Catalog Number: 490101.. Recalled by B Braun Medical Inc. Units affected: 49,330,370.
Why was this product recalled?
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Which agency issued this recall?
This recall was issued by the FDA Devices on December 3, 2025. Severity: Moderate. Recall number: Z-0640-2026.
Where was the recalled product distributed?
Distribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore..
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-0640-2026) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Recall Context

Product recalls are issued when a manufacturer, distributor, or federal agency determines that a product poses a safety risk to consumers. This recall is classified as moderate severity, indicating the product may cause temporary or medically reversible health consequences. Across PlainRecalls, we track 83,000+ recalls from FDA, CPSC, and NHTSA to help consumers stay informed and act quickly when safety issues arise.

Nearby Recalls in This Category

Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Endo-Model Replacement Plateau; Item Number: 15-0027/11; →

Data Sources

Data as of 2025. Source: FDA, CPSC, NHTSA, USDA FSIS federal recall databases.

  • Source: FDA — Food and Drug Administration, openFDA Enforcement API (food, drug, and medical device recalls)
  • Source: CPSC — Consumer Product Safety Commission Recalls API (consumer product recalls and hazards)
  • Source: NHTSA — National Highway Traffic Safety Administration Recalls API (vehicle safety recalls)
  • Source: USDA FSIS — Food Safety and Inspection Service (meat, poultry, and egg product recalls)

Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.

All federal data sources used on this page

Source: Federal recall agencies (FDA, CPSC, NHTSA, USDA FSIS) Aggregated multi-agency recall feeds · 2024 Recall data normalized across federal agency feeds; severity classifications follow each agency's own taxonomy (FDA Class I/II/III; CPSC, NHTSA, USDA FSIS).