Endurant, Endurant II and Endurant IIs Stent Graft System: Medtronic Endurant Stent Graft System Model/REF ENBF2313C120EE; ENBF2313C120EE; ENBF2313C145EE; ENBF2313C170EE; ENBF2316C120EE; ENBF2316C145EE; ENBF2316C170EE; ENBF2513C120EE; ENBF2513C145EE; ENBF2513C170EE; ENBF2516C120EE; ENBF2516C145EE; ENBF2516C170EE; ENBF2813C145EE; ENBF2813C170EE; ENBF2816C120EE; ENBF2816C145EE; ENBF2816C170EE; ENBF3216C120EE; ENBF3216C145EE; ENBF3216C170EE; ENBF3616C145EE; ENBF3616C170EE;
Product Description
Endurant, Endurant II and Endurant IIs Stent Graft System: Medtronic Endurant Stent Graft System Model/REF ENBF2313C120EE; ENBF2313C120EE; ENBF2313C145EE; ENBF2313C170EE; ENBF2316C120EE; ENBF2316C145EE; ENBF2316C170EE; ENBF2513C120EE; ENBF2513C145EE; ENBF2513C170EE; ENBF2516C120EE; ENBF2516C145EE; ENBF2516C170EE; ENBF2813C145EE; ENBF2813C170EE; ENBF2816C120EE; ENBF2816C145EE; ENBF2816C170EE; ENBF3216C120EE; ENBF3216C145EE; ENBF3216C170EE; ENBF3616C145EE; ENBF3616C170EE; ENCF2323C45EE; ENCF2525C45EE; ENCF2828C45EE; ENCF3232C45EE; ENCF3636C45EE; ENTF2323C70EE; ENTF2525C70EE; ENTF2828C70EE; Medtronic Endurant II Stent Graft System Model/REF ETBF2313C124E; ETBF2313C124EE; ETBF2313C145EE; ETBF2313C145EJ; ETBF2313C166EE; ETBF2313C166EJ; ETBF2316C124E; ETBF2316C124EE; ETBF2316C124EJ; ETBF2316C145E; ETBF2316C145EE; ETBF2316C145EJ; ETBF2316C166E; ETBF2316C166EE; ETBF2316C166EJ; ETBF2513C124E; ETBF2513C124EE; ETBF2513C124EJ; ETBF2513C145EE; ETBF2513C145EJ; ETBF2513C166E; ETBF2513C166EE; ETBF2513C166EJ; ETBF2516C124EE; ETBF2516C145E; ETBF2516C145EE; ETBF2516C145EJ; ETBF2516C166E; ETBF2516C166EE; ETBF2516C166EJ; ETBF2813C124EE; ETBF2813C124EJ; ETBF2813C145E; ETBF2813C145EE; ETBF2813C145EJ; ETBF2813C166E; ETBF2813C166EE; ETBF2813C166EJ; ETBF2816C124EE; ETBF2816C124EJ; ETBF2816C145E; ETBF2816C145EE; ETBF2816C145EJ; ETBF2816C166E; ETBF2816C166EE; ETBF2816C166EJ; ETBF2820C124EE; ETBF2820C145EE; ETBF2820C145EJ; ETBF2820C166E; ETBF2820C166EE; ETBF2820C166EJ; ETBF3216C124E; ETBF3216C124EE; ETBF3216C124EJ; ETBF3216C145E; ETBF3216C145EE; ETBF3216C145EJ; ETBF3216C166EE; ETBF3216C166EJ; ETBF3220C124EE; ETBF3220C145EE; ETBF3220C166E; ETBF3220C166EE; ETBF3220C166EJ; ETBF3616C145E; ETBF3616C145EE; ETBF3616C145EJ; ETBF3616C166E; ETBF3616C166EE; ETBF3616C166EJ; ETBF3620C145EE; ETBF3620C166EE; ETBF3620C166EJ; ETCF2323C49E; ETCF2323C49EE; ETCF2525C49E; ETCF2525C49EE; ETCF2525C49EJ; ETCF2828C49E; ETCF2828C49EE; ETCF2828C49EJ; ETCF3232C49E; ETCF3232C49EE; ETCF3232C49EJ; ETCF3636C49E; ETCF3636C49EE; ETCF3636C49EJ; ETTF2323C70E; ETTF2323C70EE; ETTF2323C70EJ; ETTF2525C70E; ETTF2525C70EE; ETTF2525C70EJ; ETTF2828C70E; ETTF2828C70EE; ETTF2828C70EJ; ETTF3232C70E; ETTF3232C70EE; ETTF3232C70EJ; ETTF3636C70E; ETTF3636C70EE; ETTF3636C70EJ; ETUF2314C102E; ETUF2314C102EE; ETUF2314C102EJ; ETUF2514C102E; ETUF2514C102EE; ETUF2514C102EJ; ETUF2814C102E; ETUF2814C102EE; ETUF2814C102EJ; ETUF3214C102E; ETUF3214C102EE; ETUF3614C102E; ETUF3614C102EE; Medtronic Endurant IIs Stent Graft System Model/REF: ESBF2314C103E; ESBF2314C103EE; ESBF2314C103EJ; ESBF2514C103E; ESBF2514C103EE; ESBF2514C103EJ; ESBF2814C103E; ESBF2814C103EE; ESBF2814C103EJ; ESBF3214C103E; ESBF3214C103EE; ESBF3214C103EJ; ESBF3614C103E; ESBF3614C103EE; ESBF3614C103EJ; Used to treat infrarenal abdominal aortic or aortoiliac aneurysms using an endovascular approach.
Reason for Recall
Stent Graft Systems built with specific batches of taper tip assemblies have the potential for the taper tip to detach from the delivery system. Detachment of the taper tip during the implant procedure may lead to a secondary intervention to attempt to remove the taper tip, either by endovascular retrieval or surgical conversion. There is also potential that the taper tip is intentionally left behind in the patient, at the physician's discretion, if not easily removable.
Details
- Recalling Firm
- Medtronic Vascular, Inc.
- Units Affected
- 14,629 devices
- Distribution
- Worldwide distribution - U.S. Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS,KY, LA, MA, MD, ME, MI, MN, MOMS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV. The countries of Argentina, Australia, Austria, Belgium, Bosnia And Herzegovina, Brazil, Bulgaria, Canada, Canary Islands, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic Of, Kosovo, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Malta, Mexico, Morocco, Netherlands, New Zealand, North Macedonia, Norway, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay and Viet Nam.
- Location
- Santa Rosa, CA
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11