NobelReplace Conical Connection RP 4.3x11.5 mm REF 36707 STERILE R

Reported: March 2, 2022 Initiated: December 17, 2021 #Z-0658-2022

Product Description

NobelReplace Conical Connection RP 4.3x11.5 mm REF 36707 STERILE R

Reason for Recall

Due to 16 complaints related to drivers not engaging in implants. According to the firm's investigations, the incorrect offset gauge was potentially used during production of the implants. As a result, this could potentially lead to aborted or a delay in surgical procedure.

Details

Recalling Firm
Nobel Biocare Usa Llc
Units Affected
702 pieces
Distribution
Worldwide distribution - U.S. Nationwide distribution in the states of AL, AR, AZ, CA, CO, FL, GA, HI, ID, IL, IN, KS, KY, MA, ME, MI, MN, MS, NC, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, VA, VT, WA and WI. The countries of Canada, Mexico and Turkey.