Severity
Moderate
FDA Devices recall · Reported January 27, 2016
Recall being initiated in response to three reports of "Burst Watchdog Timeout" events occurring with the Model 106 AspireSR Generator, resulting in a device reset condition where…
Cyberonics, Inc recalled VNS Therapy AspireSR Generator Model 106. Indicated for use as an adjunctive therapy i… - a moderate-severity action.
VNS Therapy AspireSR Generator Model 106. Indicated for use as an adjunctive therapy i… was recalled by Cyberonics, Inc in January 27, 2016. Reason: Recall being initiated in response to three reports of "Burst Watchdog Timeout" events occurring with the Mod…. Check the official notice for the remedy. Verify recall #Z-0659-2016 with the FDA Devices before acting.
The recall
Cyberonics, Inc issued this moderate-severity FDA Devices recall-Recall being initiated in response to three reports of "Burst Watchdog Timeout" events occurring with the Mod….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0659-2016 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0659-2016) was formally reported on January 27, 2016, with the manufacturer initiating the action on December 18, 2015. It is classified under Moderate severity (Class II), with a current status of Terminated. Cyberonics, Inc is listed as the recalling firm, operating out of Houston, TX. Federal records list the affected scope as 4,935 units.
The documented reason for this recall is: Recall being initiated in response to three reports of "Burst Watchdog Timeout" events occurring with the Model 106 AspireSR Generator, resulting in a device reset condition where stimulation output is disabled. Distribution data in the federal record shows the product reached: Worldwide Distribution -- United States, Austria, Belgium, Croatia, Czech Republic, Finland, France, Germany, Iceland, Italy, Netherlands, Norway, Poland, Slovakia, Spain, Sweden, Switzerland, United Kingdom, Cyprus, Is…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
4,935 units
Related Recalls
6
3 from same agency
VNS Therapy AspireSR Generator Model 106. Indicated for use as an adjunctive therapy in reducing the frequency of seizures.
Recall being initiated in response to three reports of "Burst Watchdog Timeout" events occurring with the Model 106 AspireSR Generator, resulting in a device reset condition where stimulation output is disabled.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0659-2016 |
| Date reported | January 27, 2016 |
| Date initiated | December 18, 2015 |
| Recalling firm | Cyberonics, Inc |
| Firm location | Houston, TX |
| Affected scope | 4,935 units |
| Distribution | Worldwide Distribution -- United States, Austria, Belgium, Croatia, Czech Republic, Finland, France, Germany, Iceland, Italy, Netherlands, Norway, Poland, Slovakia, Spain, Sweden, Switzerland, United Kingdom, Cyprus, Israel, Kuwait, Lebano… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported January 27, 2016. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.