FDA Devices recall · Reported December 25, 2024

Becker External Drainage and Monitoring System: 24146 EDMS 24146 BECKER NO Y-SITE 25120 EDMS II 25120 BECKER W/1-WAY VALVE 26040 KIT 26040 BECKER EDMS II 27609 BECKER 27609 W/NEEDLESLESS INJ. SITE 27670 BECKER 27670 STOP BELOW DC NDL-LS IN 27672 BECKER 27672 ONEWAY VLV Y-SITE PT LN 27702 EDMS 27702 EURO BECKER II GR PT LN 27761 BECKER 27761 STOP BCNDLESS PATLN 24 27767 BECKER 27767 NO STPCOCK ON PTLN 63 27779 EDMS 27779 BECKER BC NDLESS BOND CONN 27811 EDMS 27811 BECKER GRN PL VCATH

Due to complaints received reporting cracks and/or leaks of the stopcocks associated with the external drainage and monitoring system (EDMS).

Recall #
Z-0661-2025
Affected scope
65,810 units
Initiated
November 1, 2024
Compiled from official public sources by the editorial team.
Verify with FDA Devices →
View my saved recalls

Medtronic Neurosurgery recalled Becker External Drainage and Monitoring System: 24146 EDMS 24146 BECKER NO Y-SITE 25120… — a critical-severity action.

Becker External Drainage and Monitoring System: 24146 EDMS 24146 BECKER NO Y-SITE 25120… was recalled by Medtronic Neurosurgery in December 25, 2024. Reason: Due to complaints received reporting cracks and/or leaks of the stopcocks associated with the external draina…. Check the official notice for the remedy. Verify recall #Z-0661-2025 with the FDA Devices before acting.

The recall

Medtronic Neurosurgery issued this critical-severity FDA Devices recall — Due to complaints received reporting cracks and/or leaks of the stopcocks associated with the external draina….

Critical
severity level
66K units
affected scope
Class I
classification
December 25, 2024
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-0661-2025 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This FDA Devices action (record #Z-0661-2025) was formally reported on December 25, 2024, with the manufacturer initiating the action on November 1, 2024. It is classified under Critical severity (Class I), with a current status of Ongoing. Medtronic Neurosurgery is listed as the recalling firm, operating out of Fort Worth, TX. Federal records list the affected scope as 65,810 units.

The documented reason for this recall is: Due to complaints received reporting cracks and/or leaks of the stopcocks associated with the external drainage and monitoring system (EDMS). Distribution data in the federal record shows the product reached: Worldwide distribution. US nationwide including Puerto Rico; Algeria, Argentina, Australia, Austria, Bahrain, Barbados, Bosnia And Herzegovina, Brunei Darussalam, Cambodia, Canada, Chile, China, Colombia, Costa Rica, C…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices — clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category — 40,409 medical devices recalls on record

-1,00001,0002,0003,0004,0005,000 20052008201120142017202020232026 1,243

Where this recall sits in the database

Severity2366872097High severity (most serious)Moderate severityLow severity
Where this recall sits in the database

Of 100,165 recalls in the database, 23,668 are high severity, 72,097 moderate, and 4,400 low. This recall is classified high severity.

Counts reflect market size and reporting activity, not inherent danger — we do not rank products by risk from raw recall volume.

Severity

Critical

Affected scope

65,810 units

Related Recalls

6

6 from same agency

Product description

Becker External Drainage and Monitoring System: 24146 EDMS 24146 BECKER NO Y-SITE 25120 EDMS II 25120 BECKER W/1-WAY VALVE 26040 KIT 26040 BECKER EDMS II 27609 BECKER 27609 W/NEEDLESLESS INJ. SITE 27670 BECKER 27670 STOP BELOW DC NDL-LS IN 27672 BECKER 27672 ONEWAY VLV Y-SITE PT LN 27702 EDMS 27702 EURO BECKER II GR PT LN 27761 BECKER 27761 STOP BCNDLESS PATLN 24 27767 BECKER 27767 NO STPCOCK ON PTLN 63 27779 EDMS 27779 BECKER BC NDLESS BOND CONN 27811 EDMS 27811 BECKER GRN PL VCATH 27931 EDMS 27931 BECKER GS TUBING SMARTSITES 46128 EDMS 46128 BECKER II BLUE PT LN 46129 EDMS 46129 EURO BECKER II BLUE PT LN The Becker External Drainage and Monitoring System (Becker EDMS) is a closed system for: 1. Draining cerebrospinal fluid (CSF) from the lateral ventricles of the brain or the lumbar subarachnoid space. 2. Monitoring CSF pressure and flow rate from the lateral ventricles of the brain and the lumbar subarachnoid space.

Reason for recall

Due to complaints received reporting cracks and/or leaks of the stopcocks associated with the external drainage and monitoring system (EDMS).

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Critical (Class I)
Status Ongoing
Recall number Z-0661-2025
Date reported December 25, 2024
Date initiated November 1, 2024
Recalling firm Medtronic Neurosurgery
Firm location Fort Worth, TX
Affected scope 65,810 units
Distribution Worldwide distribution. US nationwide including Puerto Rico; Algeria, Argentina, Australia, Austria, Bahrain, Barbados, Bosnia And Herzegovina, Brunei Darussalam, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Curacao, Cyprus, Czec…

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

65,810 units units affected — limited or regional distribution scale.

Regional (<10K units)
Multi-state (10K – 100K units) ✓ This recall
Large-scale (100K – 1M units)
Massive (≥1M units)

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-0661-2025) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting — recall details can be updated.
  • This is a Class I (high-risk) recall — stop using the product immediately and follow the disposal or return instructions. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
Becker External Drainage and Monitoring System: 24146 EDMS 24146 BECKER NO Y-SITE 25120 EDMS II 25120 BECKER W/1-WAY VALVE 26040 KIT 26040 BECKER EDMS II 27609 BECKER 27609 W/NEEDLESLESS INJ. SITE 27670 BECKER 27670 STOP BELOW DC NDL-LS IN 27672 BECKER 27672 ONEWAY VLV Y-SITE PT LN 27702 EDMS 27702 EURO BECKER II GR PT LN 27761 BECKER 27761 STOP BCNDLESS PATLN 24 27767 BECKER 27767 NO STPCOCK ON PTLN 63 27779 EDMS 27779 BECKER BC NDLESS BOND CONN 27811 EDMS 27811 BECKER GRN PL VCATH 27931 EDMS 27931 BECKER GS TUBING SMARTSITES 46128 EDMS 46128 BECKER II BLUE PT LN 46129 EDMS 46129 EURO BECKER II BLUE PT LN The Becker External Drainage and Monitoring System (Becker EDMS) is a closed system for: 1. Draining cerebrospinal fluid (CSF) from the lateral ventricles of the brain or the lumbar subarachnoid space. 2. Monitoring CSF pressure and flow rate from the lateral ventricles of the brain and the lumbar subarachnoid space.. Recalled by Medtronic Neurosurgery. Units affected: 65,810 units.
Why was this product recalled?
Due to complaints received reporting cracks and/or leaks of the stopcocks associated with the external drainage and monitoring system (EDMS).
Which agency issued this recall?
This recall was issued by the FDA Devices on December 25, 2024. Severity: Critical. Recall number: Z-0661-2025.
Where was the recalled product distributed?
Distribution: Worldwide distribution. US nationwide including Puerto Rico; Algeria, Argentina, Australia, Austria, Bahrain, Barbados, Bosnia And Herzegovina, Brunei Darussalam, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Curacao, Cyprus, Czech Republic, Denmark, Dominican Republic, Egypt, Finland, French Guiana, French Polynesia, Germany, Ghana, Greece, Guadeloupe, Honduras, Hong Kong, Iceland, Indonesia, Israel, Italy, Jamaica, Jordan, Kazakhstan, Kenya, Kuwait, Lao People s Democratic Republic, Latvia, Lebanon, Libya, Malaysia, Malta, Mauritania, Morocco, Myanmar, Namibia, Netherlands, New Zealand, Nicaragua, Norway, Oman, Panama, Peru, Poland, Portugal, Qatar, Romania, Russian Federation, Rwanda, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, United Republic of Tanzania, Thailand, Trinidad And Tobago, Tunisia, Turkmenistan, United Arab Emirates, United Kingdom, Uruguay, Vietnam, and Yemen..
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-0661-2025) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nearby Recalls in This Category

Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with GE HealthCare TRADE IN CARESCAPE TELEMETRY SERVER ARK-2250L… →

Browse all Medical Devices recalls →

Data Sources

Source: FDA, CPSC, and NHTSA federal recall databases. This recall: FDA Devices, reported December 25, 2024.

  • Source: FDA — Food and Drug Administration, openFDA Enforcement API (food, drug, and medical device recalls)
  • Source: CPSC — Consumer Product Safety Commission Recalls API (consumer product recalls and hazards)
  • Source: NHTSA — National Highway Traffic Safety Administration Recalls API (vehicle safety recalls)

Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.

Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records — no number is typed in by an editor. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of June 2026.