FDA Devices Moderate Class II Terminated

Protura Software which utilizes Elekta's iCOM interface: MT6XSM1.4.0, MT6XSM1.4.0-1, MT6XSM1.4.0-2, MT6SXM1.4.0-3, MT6XSM1.5.0-2, MT6XSM1.6.0-1, MT6XSM1.7.2, MT6XSM1.7.2-1, MT6XSM1.7.2-3, MT6XSM1.7.3, MT6XSM1.7.3-1, MT6XSM1.7.3-3. The Protura Software is intended to interface between record and verify systems, linear accelerator (Linac) software systems, and/or image guidance systems and the Protura Couch. The system is intended to support a treatment table and add pitch and roll to normal

Reported: February 28, 2018 Initiated: November 8, 2017 #Z-0671-2018

Product Description

Protura Software which utilizes Elekta's iCOM interface: MT6XSM1.4.0, MT6XSM1.4.0-1, MT6XSM1.4.0-2, MT6SXM1.4.0-3, MT6XSM1.5.0-2, MT6XSM1.6.0-1, MT6XSM1.7.2, MT6XSM1.7.2-1, MT6XSM1.7.2-3, MT6XSM1.7.3, MT6XSM1.7.3-1, MT6XSM1.7.3-3. The Protura Software is intended to interface between record and verify systems, linear accelerator (Linac) software systems, and/or image guidance systems and the Protura Couch. The system is intended to support a treatment table and add pitch and roll to normal X, Y, Z and yaw motions to provide 6-degree axis movement. The product includes an interface between the treatment planning system and the treatment table to direct the table motion based on coordinates supplied. These products connect or support a treatment table by use of various mounting plates.

Reason for Recall

When an error message remains displayed and is not cleared in the Protura software with the Elekta pedestal coordinates, moving the pedestal could cause the Protura software to not update with the pedestal location and buffers the pedestal movement history.

Details

Recalling Firm
Med Tec Inc
Units Affected
5 installations
Distribution
US Distribution to Iowa and South Dakota. Shipped internationally to the Netherlands.
Location
Orange City, IA

Frequently Asked Questions

What product was recalled?
Protura Software which utilizes Elekta's iCOM interface: MT6XSM1.4.0, MT6XSM1.4.0-1, MT6XSM1.4.0-2, MT6SXM1.4.0-3, MT6XSM1.5.0-2, MT6XSM1.6.0-1, MT6XSM1.7.2, MT6XSM1.7.2-1, MT6XSM1.7.2-3, MT6XSM1.7.3, MT6XSM1.7.3-1, MT6XSM1.7.3-3. The Protura Software is intended to interface between record and verify systems, linear accelerator (Linac) software systems, and/or image guidance systems and the Protura Couch. The system is intended to support a treatment table and add pitch and roll to normal X, Y, Z and yaw motions to provide 6-degree axis movement. The product includes an interface between the treatment planning system and the treatment table to direct the table motion based on coordinates supplied. These products connect or support a treatment table by use of various mounting plates.. Recalled by Med Tec Inc. Units affected: 5 installations.
Why was this product recalled?
When an error message remains displayed and is not cleared in the Protura software with the Elekta pedestal coordinates, moving the pedestal could cause the Protura software to not update with the pedestal location and buffers the pedestal movement history.
Which agency issued this recall?
This recall was issued by the FDA Devices on February 28, 2018. Severity: Moderate. Recall number: Z-0671-2018.