PlainRecalls
ModerateClass IITerminated

FDA Devices recall · Reported February 28, 2018

Protura Software which utilizes Elekta's iCOM interface: MT6XSM1.4.0, MT6XSM1.4.0-1, MT6XSM1.4.0-2, MT6SXM1.4.0-3, MT6XSM1.5.0-2, MT6XSM1.6.0-1, MT6XSM1.7.2, MT6XSM1.7.2-1, MT6XSM1.7.2-3, MT6XSM1.7.3, MT6XSM1.7.3-1, MT6XSM1.7.3-3. The Protura Software is intended to interface between record and verify systems, linear accelerator (Linac) software systems, and/or image guidance systems and the Protura Couch. The system is intended to support a treatment table and add pitch and roll to normal

When an error message remains displayed and is not cleared in the Protura software with the Elekta pedestal coordinates, moving the pedestal could cause the Protura software to no…

Recall #
Z-0671-2018
Affected scope
5 installations
Initiated
November 8, 2017
Verify with FDA Devices →
View my saved recalls

Med Tec Inc recalled Protura Software which utilizes Elekta's iCOM interface: MT6XSM1.4.0, MT6XSM1.4.0-1, MT6X… - a moderate-severity action.

Protura Software which utilizes Elekta's iCOM interface: MT6XSM1.4.0, MT6XSM1.4.0-1, MT6X… was recalled by Med Tec Inc in February 28, 2018. Reason: When an error message remains displayed and is not cleared in the Protura software with the Elekta pedestal c…. Check the official notice for the remedy. Verify recall #Z-0671-2018 with the FDA Devices before acting.

The recall

Med Tec Inc issued this moderate-severity FDA Devices recall-When an error message remains displayed and is not cleared in the Protura software with the Elekta pedestal c….

Moderate
severity level
Class II
classification
February 28, 2018
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-0671-2018 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 41,021 medical devices recalls on record

01,0002,0003,0004,000 20052008201120142017202020232026 1,947
Severity1906864652High severity (most serious)Moderate severityLow severity
Where this agency-classified recall sits among FDA Class I/II/III notices

Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Product description

Protura Software which utilizes Elekta's iCOM interface: MT6XSM1.4.0, MT6XSM1.4.0-1, MT6XSM1.4.0-2, MT6SXM1.4.0-3, MT6XSM1.5.0-2, MT6XSM1.6.0-1, MT6XSM1.7.2, MT6XSM1.7.2-1, MT6XSM1.7.2-3, MT6XSM1.7.3, MT6XSM1.7.3-1, MT6XSM1.7.3-3. The Protura Software is intended to interface between record and verify systems, linear accelerator (Linac) software systems, and/or image guidance systems and the Protura Couch. The system is intended to support a treatment table and add pitch and roll to normal X, Y, Z and yaw motions to provide 6-degree axis movement. The product includes an interface between the treatment planning system and the treatment table to direct the table motion based on coordinates supplied. These products connect or support a treatment table by use of various mounting plates.

Reason for recall

When an error message remains displayed and is not cleared in the Protura software with the Elekta pedestal coordinates, moving the pedestal could cause the Protura software to not update with the pedestal location and buffers the pedestal movement history.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Terminated
Recall number Z-0671-2018
Date reported February 28, 2018
Date initiated November 8, 2017
Recalling firm Med Tec Inc
Firm location Orange City, IA
Affected scope 5 installations
Distribution US Distribution to Iowa and South Dakota. Shipped internationally to the Netherlands.

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-0671-2018) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
Protura Software which utilizes Elekta's iCOM interface: MT6XSM1.4.0, MT6XSM1.4.0-1, MT6XSM1.4.0-2, MT6SXM1.4.0-3, MT6XSM1.5.0-2, MT6XSM1.6.0-1, MT6XSM1.7.2, MT6XSM1.7.2-1, MT6XSM1.7.2-3, MT6XSM1.7.3, MT6XSM1.7.3-1, MT6XSM1.7.3-3. The Protura Software is intended to interface between record and verify systems, linear accelerator (Linac) software systems, and/or image guidance systems and the Protura Couch. The system is intended to support a treatment table and add pitch and roll to normal X, Y, Z and yaw motions to provide 6-degree axis movement. The product includes an interface between the treatment planning system and the treatment table to direct the table motion based on coordinates supplied. These products connect or support a treatment table by use of various mounting plates.. Recalled by Med Tec Inc. Units affected: 5 installations.
Why was this product recalled?
When an error message remains displayed and is not cleared in the Protura software with the Elekta pedestal coordinates, moving the pedestal could cause the Protura software to not update with the pedestal location and buffers the pedestal movement history.
Which agency issued this recall?
This recall was issued by the FDA Devices on February 28, 2018. Severity: Moderate. Recall number: Z-0671-2018.
Where was the recalled product distributed?
Distribution: US Distribution to Iowa and South Dakota. Shipped internationally to the Netherlands..
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nationwide recalls with similar severity or category timing

Drug and device notices outside this firm and state, matched by severity timing and product category.

Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls

This recall: FDA Devices, reported February 28, 2018. Severity classes follow each agency's own taxonomy; related-recall context is computed across the full archive.

Methodology · Corrections · Archive as of August 2026