PlainRecalls
FDA Devices Moderate Class II Ongoing

BD Pyxis MedStation ES 7 Drawer Auxiliary Tower, REF: 324

Reported: December 18, 2024 Initiated: November 14, 2024 #Z-0671-2025

Product Description

BD Pyxis MedStation ES 7 Drawer Auxiliary Tower, REF: 324

Reason for Recall

Due to a software issue, automated dispensing cabinet devices may open the wrong cubie pocket or position, which can lead to: 1) Unintended and incorrect medication removal, or 2) Potential for a different medication (medication pulled in error) to be refilled/loaded, which may lead to may lead to low inventory or a stockout.

Details

Recalling Firm
CareFusion 303, Inc.
Units Affected
40,763
Distribution
Worldwide - US Nationwide distribution including in the states of OK, LA, FL, AR, VA, KY, PA, AL, NC, CA, MN, AK, TX, MI, ME, OH, OR, NY, IL, RI, GA, WI, NH, KS, CO, WA, MO, MD, AZ, SC, NV, WV, NM, IN, NJ, TN, DE, MT, MA, VT, DC, NE, CT, ID, WY, MS, UT, HI, IA, SD, ND, GU, MP, PR and the countries of CA, BR, JP, ES, NZ, AU, FR, AR, MX, IT, CL, TR, SA, MY, CH, IE, TH, SG, KW, BE, DE, PT, QA, MA, FI, AE, EG, BS, GB, BM, GR, BH, KR, MC, PL, LB, HK, IN, IL, MO, IQ, TN, NO, PH, CO, TW, CN, JO, OM.
Location
San Diego, CA

Frequently Asked Questions

What product was recalled?
BD Pyxis MedStation ES 7 Drawer Auxiliary Tower, REF: 324. Recalled by CareFusion 303, Inc.. Units affected: 40,763.
Why was this product recalled?
Due to a software issue, automated dispensing cabinet devices may open the wrong cubie pocket or position, which can lead to: 1) Unintended and incorrect medication removal, or 2) Potential for a different medication (medication pulled in error) to be refilled/loaded, which may lead to may lead to low inventory or a stockout.
Which agency issued this recall?
This recall was issued by the FDA Devices on December 18, 2024. Severity: Moderate. Recall number: Z-0671-2025.