A correction was issued for the SIGMA Spectrum Infusion Pump with Master Drug Library and Rear Case Assembly in order to replace speakers installed in pumps or included in spare p…
Baxter Healthcare Corp. recalled V6 Rear Case Assembly, Product Code 35701, a replacement part for the SIGMA Spectrum Infu… - a moderate-severity action.
V6 Rear Case Assembly, Product Code 35701, a replacement part for the SIGMA Spectrum Infu… was recalled by Baxter Healthcare Corp. in December 7, 2016. Reason: A correction was issued for the SIGMA Spectrum Infusion Pump with Master Drug Library and Rear Case Assembly …. Check the official notice for the remedy. Verify recall #Z-0672-2017 with the FDA Devices before acting.
The recall
Baxter Healthcare Corp. issued this moderate-severity FDA Devices recall-A correction was issued for the SIGMA Spectrum Infusion Pump with Master Drug Library and Rear Case Assembly ….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0672-2017 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0672-2017) was formally reported on December 7, 2016, with the manufacturer initiating the action on September 28, 2016. It is classified under Moderate severity (Class II), with a current status of Terminated. Baxter Healthcare Corp. is listed as the recalling firm, operating out of Deerfield, IL. Federal records list the affected scope as 439 units.
The documented reason for this recall is: A correction was issued for the SIGMA Spectrum Infusion Pump with Master Drug Library and Rear Case Assembly in order to replace speakers installed in pumps or included in spare part rear case assemblies that have an in… Distribution data in the federal record shows the product reached: Nationwide Distribution, Puerto Rico and Canada. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
439 units
Related Recalls
6
3 from same agency
V6 Rear Case Assembly, Product Code 35701, a replacement part for the SIGMA Spectrum Infusion Pump -- SIGMA LLC Medina, NY 14103 --- NOTE: The speakers are a part of the rear case assembly located in the back of the infusion pump. The speaker is intended to generate audible tones during each keystroke, and during alarm conditions. However, some speakers have a variation in their internal wiring that increases the potential for the speaker to fail prematurely. Spectrum Infusion Pumps with non-functional speakers will continue to display visual alarm indications, and will otherwise continue to operate as intended. Product Code 35701 is a replacement assembly that is shipped to customers for self-repair of the Spectrum Infusion Pump. It is not a registered finished good.
A correction was issued for the SIGMA Spectrum Infusion Pump with Master Drug Library and Rear Case Assembly in order to replace speakers installed in pumps or included in spare part rear case assemblies that have an increased potential to fail prematurely.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0672-2017 |
| Date reported | December 7, 2016 |
| Date initiated | September 28, 2016 |
| Recalling firm | Baxter Healthcare Corp. |
| Firm location | Deerfield, IL |
| Affected scope | 439 units |
| Distribution | Nationwide Distribution, Puerto Rico and Canada |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported December 7, 2016. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.