Computed Tomography X-Ray System Brilliance Workspace Portal, Model #728269 Philips Medical Systems, Highland Heights, OH. The Brilliance Workspace Portal is a component of the Computed Tomography X-Ray System that is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. The portal (together with the scanner) provides image processing and display through software applications that process, analyze, di
Philips was notified that when performing post processing with the Brilliance Workspace Portal platform with software versions 2.6.1.4, the coronary arteries in the rendered image…
Philips Medical Systems (Cleveland) Inc recalled Computed Tomography X-Ray System Brilliance Workspace Portal, Model #728269 Philips Medic… - a moderate-severity action.
Computed Tomography X-Ray System Brilliance Workspace Portal, Model #728269 Philips Medic… was recalled by Philips Medical Systems (Cleveland) Inc in January 23, 2013. Reason: Philips was notified that when performing post processing with the Brilliance Workspace Portal platform with …. Check the official notice for the remedy. Verify recall #Z-0675-2013 with the FDA Devices before acting.
The recall
Philips Medical Systems (Cleveland) Inc issued this moderate-severity FDA Devices recall-Philips was notified that when performing post processing with the Brilliance Workspace Portal platform with ….
- Moderate
- severity level
- 108 units
- affected scope
- Class II
- classification
- January 23, 2013
- reported
Sourced from official FDA Devices enforcement records. Verify recall #Z-0675-2013 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
Recall Insight
This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Medical Devices recalls over time
Where this recall sits in its category - 41,021 medical devices recalls on record
Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Product description
Computed Tomography X-Ray System Brilliance Workspace Portal, Model #728269 Philips Medical Systems, Highland Heights, OH. The Brilliance Workspace Portal is a component of the Computed Tomography X-Ray System that is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. The portal (together with the scanner) provides image processing and display through software applications that process, analyze, display, quantify and interpret medical images/data.
Reason for recall
Philips was notified that when performing post processing with the Brilliance Workspace Portal platform with software versions 2.6.1.4, the coronary arteries in the rendered image are segmented automatically. However, the center line is not extracted or highlighted upon hovering and making it impossible to activate the vessel and choose the appropriate name (label) from the drop-down menu.
Recall Profile
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0675-2013 |
| Date reported | January 23, 2013 |
| Date initiated | September 11, 2012 |
| Recalling firm | Philips Medical Systems (Cleveland) Inc |
| Firm location | Cleveland, OH |
| Affected scope | 108 units |
| Distribution | Worldwide Distribution-USA including the states of AZ, CA, CO, IL, IL, IN, MA, ME, MI, MO, NC, NJ, NY, OR, TN, TX, VA and WI and the countries of Austria, Canada, China, France, India, Ireland, Italy, Japan, Netherlands, Norway, Spain, Sw… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
- Check the recall number (Z-0675-2013) and product description against the item you own. Search the archive
- Confirm the current status with FDA Devices before acting, recall details can be updated.
- Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Where was the recalled product distributed? ▼
Should I stop using a recalled medication or medical device? ▼
Nationwide recalls with similar severity or category timing
Drug and device notices outside this firm and state, matched by severity timing and product category.
Similar severity · nearest report date
Same archive severity grade (Moderate), closest by report date among other firms and states.
Similar Medical Devices · nearest report date
Same product category, closest by report date among other firms and states (deduped against the severity set).
Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls
Methodology · Corrections · Archive as of August 2026