Severity
Moderate
FDA Devices recall · Reported January 2, 2019
If a Setup field (CT, kV, MV or MVCT) is created by COPYING a treatment field or another setup field, then the copied field will have the same isocenter as the parent and that is…
Elekta, Inc. recalled MOSAIQ Oncology Information System Version 2.64. MOSAIQ is an oncology information s… - a moderate-severity action.
MOSAIQ Oncology Information System Version 2.64. MOSAIQ is an oncology information s… was recalled by Elekta, Inc. in January 2, 2019. Reason: If a Setup field (CT, kV, MV or MVCT) is created by COPYING a treatment field or another setup field, then th…. Check the official notice for the remedy. Verify recall #Z-0686-2019 with the FDA Devices before acting.
The recall
Elekta, Inc. issued this moderate-severity FDA Devices recall-If a Setup field (CT, kV, MV or MVCT) is created by COPYING a treatment field or another setup field, then th….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0686-2019 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0686-2019) was formally reported on January 2, 2019, with the manufacturer initiating the action on December 4, 2018. It is classified under Moderate severity (Class II), with a current status of Terminated. Elekta, Inc. is listed as the recalling firm, operating out of Atlanta, GA. Federal records list the affected scope as 551 units.
The documented reason for this recall is: If a Setup field (CT, kV, MV or MVCT) is created by COPYING a treatment field or another setup field, then the copied field will have the same isocenter as the parent and that isocenter may not be the correct isocenter… Distribution data in the federal record shows the product reached: Distributed in the United States and outside of the Unites States: Algeria; Antigua and Barbuda; Argentina; Australia; Belgium; Botswana; Brazil; Canada; China; Colombia; Denmark; France; Germany; India; Indonesia; Iran…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
551 units
Related Recalls
6
3 from same agency
MOSAIQ Oncology Information System Version 2.64. MOSAIQ is an oncology information system used to manage workflows for treatment planning and delivery
If a Setup field (CT, kV, MV or MVCT) is created by COPYING a treatment field or another setup field, then the copied field will have the same isocenter as the parent and that isocenter may not be the correct isocenter for imaging when it is sent to the machine.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0686-2019 |
| Date reported | January 2, 2019 |
| Date initiated | December 4, 2018 |
| Recalling firm | Elekta, Inc. |
| Firm location | Atlanta, GA |
| Affected scope | 551 units |
| Distribution | Distributed in the United States and outside of the Unites States: Algeria; Antigua and Barbuda; Argentina; Australia; Belgium; Botswana; Brazil; Canada; China; Colombia; Denmark; France; Germany; India; Indonesia; Iran; Ireland; Israel; I… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported January 2, 2019. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.