Severity
Moderate
FDA Devices recall · Reported December 25, 2024
The warming surfaces or incubator well temperatures of the listed products may exceed the desired set temperature. If temperatures exceed the desired set value, the device will a…
CooperSurgical, Inc. recalled Brand Name: K-Systems Product Name: G73 Dry Bath Incubator Model/Catalog Number: K22065… - a moderate-severity action.
Brand Name: K-Systems Product Name: G73 Dry Bath Incubator Model/Catalog Number: K22065… was recalled by CooperSurgical, Inc. in December 25, 2024. Reason: The warming surfaces or incubator well temperatures of the listed products may exceed the desired set tempera…. Check the official notice for the remedy. Verify recall #Z-0687-2025 with the FDA Devices before acting.
The recall
CooperSurgical, Inc. issued this moderate-severity FDA Devices recall-The warming surfaces or incubator well temperatures of the listed products may exceed the desired set tempera….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0687-2025 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0687-2025) was formally reported on December 25, 2024, with the manufacturer initiating the action on November 15, 2024. It is classified under Moderate severity (Class II), with a current status of Ongoing. CooperSurgical, Inc. is listed as the recalling firm, operating out of Trumbull, CT. Federal records list the affected scope as 106 units.
The documented reason for this recall is: The warming surfaces or incubator well temperatures of the listed products may exceed the desired set temperature. If temperatures exceed the desired set value, the device will alarm appropriately. Distribution data in the federal record shows the product reached: States: NC, NJ, TX, VA, IL, NY, CA, PA Please note: The countries below, excluding Iraq, all consider these devices to be "lab equipment", not medical devices and are not subject to classification or recall guidelines.…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
106 units
Related Recalls
6
3 from same agency
Brand Name: K-Systems Product Name: G73 Dry Bath Incubator Model/Catalog Number: K22065 Product Description: An electrically-powered device intended to provide suitable warming and/or incubation conditions for reproductive biological materials (e.g., gametes and embryos), at closely regulated and monitored temperatures, during assisted reproduction examination and processing procedures. It may support a more comprehensive growth environment [e.g., temperature, carbon dioxide (CO2) and oxygen (O2) concentration]; it typically has adjustment and monitoring controls. Component: N/A
The warming surfaces or incubator well temperatures of the listed products may exceed the desired set temperature. If temperatures exceed the desired set value, the device will alarm appropriately.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-0687-2025 |
| Date reported | December 25, 2024 |
| Date initiated | November 15, 2024 |
| Recalling firm | CooperSurgical, Inc. |
| Firm location | Trumbull, CT |
| Affected scope | 106 units |
| Distribution | States: NC, NJ, TX, VA, IL, NY, CA, PA Please note: The countries below, excluding Iraq, all consider these devices to be "lab equipment", not medical devices and are not subject to classification or recall guidelines. Countries: AT, AU… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported December 25, 2024. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.